I.T.S. Screw System with the below descriptions and… recallZ-1599-2026 · I.T.S. GmbH · initiated Jan 23, 2026
Class II Open, Classified All recalls posted in March 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from I.T.S. GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
FDA root cause category: No Marketing Application
What the company is doing
ITS initially notified its importer/distributor on about 01/23/2026 via email. Instructions were to cease distribution, and further information will be provided. An additional notification was sent to distributors and consignees on about 02/04/2026 via email. Instructions included to ensure any IFU different to the provided one is removed from systems and/or destroyed, notify clinical staff managing patients with implants, contact I.T.S. GmbH for any questions or to discuss proper scanning conditions for implanted devices in the interim while they work on labeling and IFU updates for the devices, and arrange for the return of specific originally packaged items - base plates and screws in surgery trays. Other outlined products can continue to be used upon confirmation from ITS. Consignees were also instructed to complete and return the response form. Distributors were instructed to continue to immediately halt product distribution from the U.S. office to sales groups and hospitals, remove the outdated paper IFU from packaging and add the eIFU symbol to product labeling to implement the updated IFU provided electronically - see annex "F-7.5.5-3-US_IFU_non-sterile_Rev00_DRAFT". Paper IFUs will be made available upon request. They were also requested to assist in notification of hospitals.
Recall details
- Recall number
- Z-1599-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jan 23, 2026
- Posted by the FDA
- Mar 20, 2026
- Recalling firm
- I.T.S. GmbH
- Quantity
- 5,568 units
- Product code
- HWC
- FDA submissions
- K060156
- Distribution
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Product description
I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 31404-44, 31404-46, 31404-48, 31404-50, 31404-55, 31404-60, 31404-65, 31404-70, 31404-75, 31404-80, 31457-30, 31457-36, 31457-40, 31457-46, 31457-50, 31457-55, 31457-60, 31652-100, 31652-105, 31652-110, 31652-115, 31652-120, 31652-25, 31652-30, 31652-35, 31652-40, 31652-45, 31652-55, 31652-60, 31652-65, 31652-70, 31652-75, 31652-80, 31652-85, 31652-90, 31652-95, 31731-100, 31731-105, 31731-110, 31731-115, 31731-120, 31731-50, 31731-60, 31731-65, 31731-70, 31731-75, 31731-80, 31731-85, 31731-90, 31731-95, 31732-60, 31734-100, 31734-105, 31734-110, 31734-115, 31734-120, 31734-45, 31734-50, 31734-60, 31734-65, 31734-70, 31734-75, 31734-80, 31734-85, 31734-90, 31734-95. 2. Cortical Screw, Cannulated; Article Numbers: 31357-44, 31357-22, 31357-48, 31277-18, 31357-20, 31357-22, 31357-24, 31357-26, 31357-28, 31357-30, 31357-32, 31357-34, 31357-36, 31357-38, 31357-40, 31357-42, 31357-44, 31357-46, 31357-48, 31357-60, 31357-65, 31357-70. 3. Washer; Article Number: 36851. 4. Washer, Concave; Article Numbers: 36432, 36653, 36732. 5. Washer, Flat; Article Numbers: 36431, 36651, 36653, 36731.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
1. Article Number (Lot Numbers): 31404-16 (679/062339, 679/122213, 679/122332), 31404-18 (679/122212, 679/062340), 31404-20 (18/089221, 679/062343, 679/122241), 31404-22 (18/061535, 679/012416, 679/062350), 31404-24 (18/071536, 679/062362, 679/122246), 31404-26 (679/122245, 679/062363), 31404-28 (18/186278, 679/062364), 31404-30 (18/187299, 679/012401, 679/062365, 679/122242), 31404-32 (18/111540, 679/062366, 679/122211), 31404-34 (679/122206, 18/121544, 679/012402, 679/042414, 679/072302, 679/122206), 31404-36 (18/188300, 679/052403, 679/072305, 679/112223), 31404-38 (18/189301, 679/112213, 679/012455, 679/072309), 31404-40 (18/181043, 679/062321, 679/122201, 679/122310, 679/052404, 679/032532), 31404-42 (679/122203, 18/191303, 679/072310, 679/012418, 679/052424, 679/052505), 31404-44 (18/192304, 679/112232, 679/052332, 679/052425, 679/052508), 31404-46 (679/122240, 679/062322, 679/112210, 679/052426, 679/042505), 31404-48 (679/052427, 679/072317, 679/122248), 31404-50 (18/194306, 679/112228, 679/062301, 679/122311, 679/052406, 679/052413, 679/042506, 679/062501), 31404-55 (18/195307, 679/122258, 18/211647, 679/072326, 679/012452, 679/062467, 679/062511), 31404-60 (18/196308, 679/112240, 679/052325, 679/122312, 679/062485, 679/062512), 31404-65 (679/122259, 679/072327, 18/205310, 679/022410), 31404-70 (679/012301, 18/197309, 679/072328, 679/022413, 679/062479), 31404-75 (18/205411, 18/381989, 679/012302), 31404-80 (18/172638, 18/205232, 18/321164, 679/022416), 31457-30 (26/F50332), 31457-36 (18/198343), 31457-40 (18/199344), 31457-46 (18/947783, 18/736681), 31457-50 (18/201346, 26/F67796), 31457-55 (18/202347, 26/F67797), 31457-60 (18/203355), 31652-100 (679/122117, 679/022350, 679/102335, 679/022335, 18/924999), 31652-105 (679/022336, 679/032377), 31652-110 (679/032380, 679/102339), 31652-115 (679/102336), 31652-120 (679/092002, 679/102347), 31652-25 (679/112134, 679/032340), 31652-30 (679/032341), 31652-35 (679/082315), 31652-40 (679/082316), 31652-45 (679/032330, 679/092332), 31652-55 (679/032391, 679/092333), 31652-60 (679/022337, 679/032378, 679/092336), 31652-65 (679/022338, 679/032322, 679/092347), 31652-70 (679/032320, 679/032365, 679/042405, 679/112154), 31652-75 (26/F65330, 679/032348, 679/032386), 31652-80 (679/112181, 679/032321, 679/092350, 679/032362, 679/092348, 679/042410, 18/920995), 31652-85 (679/122145, 679/032382, 679/022349, 679/102301, 679/102305, 679/042411, 18/541045), 31652-90 (679/012201, 679/032383, 679/032383, 679/022347, 679/042406), 31652-95 (18/923998, 679/012202, 679/032316), 31731-100 (18/775943, 26/F58803, 679/102420), 31731-105 (18/776944, 26/F58804, 26/F65649, 679/102421), 31731-110 (26/F58805, 26/F65650, 679/102422), 31731-115 (26/F58806, 26/F65651, 679/102423), 31731-120 (26/F58807, 26/F65652, 679/102436), 31731-50 (09/1244, 26/F14488, 26/F21185, 26/F58793), 31731-60 (26/F58795, 26/F65655), 31731-65 (26/F65656, 26/F58796), 31731-70 (26/F58797, 26/F65657), 31731-75 (26/F58798, 26/F65658), 31731-80 (679/072040, 26/F58799, 26/F65659), 31731-85 (18/331317, 26/F65660, 679/072044), 31731-90 (09/1247, 09/1248, 26/F58801, 26/F65661, 46/612), 31731-95 (18/581247, 26/F58802, 26/F65662, 679/102437), 31732-60 (18/961265), 31734-100 (26/F65672), 31734-105 (26/F58839, 26/F65673), 31734-110 (26/F58840, 26/F65674, 679/032425), 31734-115 (26/F65675), 31734-120 (26/F58842, 26/F65676, 26/F72277), 31734-45 (18/761192), 31734-50 (26/F58828, 26/F65680), 31734-60 (26/F58830), 31734-65 (26/F58831, 26/F65683, 679/032428), 31734-70 (26/F58832, 26/F65684, 26/F72275, 679/032429), 31734-75 (26/F58833, 26/F65685), 31734-80 (26/F58834, 26/F65686, 679/032435), 31734-85 (26/F58835, 26/F65687, 26/F72276), 31734-90 (26/F65688, 679/032436), 31734-95 (26/F58837, 26/F65689, 679/042401). 2. Article Number (Lot Numbers): 31277-18 (26/F36049), 31357-20 (18/091247), 31357-22 (679/031707), 31357-24 (86/485378), 31357-26 (18/121250), 31357-28 (26/F65909), 31357-30 (18/165201, 18/341054), 31357-32 (18/166217), 31357-34 (18/167218), 31357-36 (18/168219, 26/F65910), 31357-38 (18/167220), 31357-40 (26/F65911, 18/170221), 31357-42 (26/F65912), 31357-44 (18/171222, 18/361059, 26/F65913), 31357-46 (18/861060), 31357-48 (18/172223, 26/F65914), 31357-60 (26/F65648), 31357-65 (679/012125), 31357-70 (18/902013). 3. Article Number (Lot Numbers): 36851 (26/F59918, 26/F61351, 26/F64885, 26/F64886, 26/F65908, 26/F66506, 26/F71413, 26/F71764, 26/F72364, 26/F73020, 26/F74288). 4. Article Number (Lot Numbers): 36432 (26/F65379, 26/F67799, 26/F72744, 774/300002), 36653 (26/F62590, 26/F68455), 36732 (26/F67800). 5. Article Number (Lot Numbers): 36431 (26/F62589, 26/F65378, 26/F66505, 26/F67798, 26/F70092, 26/F72743, 26/F73017, 26/F73494), 36651 (26/F61694, 26/F65380, 26/F68454, 26/F73018), 36653 (26/F73019), 36731 (26/F60987, 26/F62591, 26/F67523, 26/F70093, 26/F72363, 26/F73495).
Other devices under the same FDA submission
12 devices cleared under K060156 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Straight Plate, 2.0mm 21101-13I.T.S. GmbH · Plate, Fixation, Bone3 reports
- Straight Plate, 3.5mm 21102-8I.T.S. GmbH · Plate, Fixation, Bone2 reports
- Straight Plate, 4.5mm 21103-10I.T.S. GmbH · Plate, Fixation, Bone2 reports
- Straight Plate, 4.5mm 21103-13I.T.S. GmbH · Plate, Fixation, Bone2 reports
- Cancellous Screw, Cannulated 31404-70I.T.S. GmbH · Screw, Fixation, Bone2 reports
- Straight Compression Plate 21105-10I.T.S. GmbH · Plate, Fixation, Bone2 reports
- Straight Plate, 2.0mm 21101-11I.T.S. GmbH · Plate, Fixation, Bone1 report
- Straight Plate, 3.5mm 21102-7I.T.S. GmbH · Plate, Fixation, Bone1 report
- Straight Plate, 3.5mm 21102-9I.T.S. GmbH · Plate, Fixation, Bone1 report
- Straight Plate, 4.5mm 21103-12I.T.S. GmbH · Plate, Fixation, Bone1 report
- Cancellous Screw, Cannulated 31404-60I.T.S. GmbH · Screw, Fixation, Bone1 report
- Cancellous Screw, Cannulated 31457-65I.T.S. GmbH · Screw, Fixation, Bone1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98341).
- Z-1592-2026 · Open, ClassifiedVolar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, …
- Z-1593-2026 · Open, ClassifiedClaviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-1…
- Z-1594-2026 · Open, ClassifiedBrand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Humeral Plate. Article Numbers: 211341-10, 211341-4, 211341-7, 21134-4, 21134-5, 211…
- Z-1595-2026 · Open, ClassifiedFR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
- Z-1596-2026 · Open, ClassifiedPilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 2131…
- Z-1598-2026 · Open, ClassifiedI.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Fore…
- Z-1600-2026 · Open, ClassifiedPelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 7031…
- Z-1601-2026 · Open, ClassifiedI.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 2160…
- Z-1603-2026 · Open, ClassifiedI.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301…
- Z-1604-2026 · Open, ClassifiedI.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) …
- Z-1605-2026 · Open, ClassifiedFLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Canc…
- Z-1606-2026 · Open, ClassifiedHCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31…
- Z-1607-2026 · Open, ClassifiedHLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11,…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page