Skip to content

FLS - Foot Locking Plates System and HOL - Hallux Osteotomy… recallZ-1605-2026 · I.T.S. GmbH · initiated Jan 23, 2026

Class II Open, Classified All recalls posted in March 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from I.T.S. GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

FDA root cause category: No Marketing Application

What the company is doing

ITS initially notified its importer/distributor on about 01/23/2026 via email. Instructions were to cease distribution, and further information will be provided. An additional notification was sent to distributors and consignees on about 02/04/2026 via email. Instructions included to ensure any IFU different to the provided one is removed from systems and/or destroyed, notify clinical staff managing patients with implants, contact I.T.S. GmbH for any questions or to discuss proper scanning conditions for implanted devices in the interim while they work on labeling and IFU updates for the devices, and arrange for the return of specific originally packaged items - base plates and screws in surgery trays. Other outlined products can continue to be used upon confirmation from ITS. Consignees were also instructed to complete and return the response form. Distributors were instructed to continue to immediately halt product distribution from the U.S. office to sales groups and hospitals, remove the outdated paper IFU from packaging and add the eIFU symbol to product labeling to implement the updated IFU provided electronically - see annex "F-7.5.5-3-US_IFU_non-sterile_Rev00_DRAFT". Paper IFUs will be made available upon request. They were also requested to assist in notification of hospitals.

Recall details

Recall number
Z-1605-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jan 23, 2026
Posted by the FDA
Mar 20, 2026
Recalling firm
I.T.S. GmbH
Quantity
15,517 units
Product code
HRS
FDA submissions
K131722
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Product description

FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

1. Article Number (Lot Numbers): 37303-10 (18/151720, 18/268729, 679/032432, 679/072268, 679/102237, 679/102307), 37303-12 (18/570637, 679/032434, 679/072269, 679/102346, 679/112227), 37303-14 (18/751091, 26/F66913, 679/012510, 679/062288, 679/102359, 679/112410), 37303-16 (679/072263, 679/102245, 18/151841, 26/F66914, 18/331193, 18/543625, 679/012511, 679/042403, 679/102360), 37303-18 (18/151942, 18/210467, 18/231415, 18/341194, 18/572633, 18/771093, 679/012512, 679/022507, 679/042404, 679/102248, 679/112319), 37303-20 (18/152043, 18/211468, 18/351195, 18/573634, 18/701585, 18/781094, 679/022508, 679/042420, 679/112216, 679/112328), 37303-22 (18/171705, 18/212469, 18/361196, 18/574635, 18/791095, 679/022509, 679/042417, 679/112224, 679/112340), 37303-24 (18/575639, 18/741572, 26/F66915, 679/022510, 679/112226, 679/122313, 679/122333), 37303-26 (18/152346, 18/214471, 26/F66916, 679/012403, 679/112230, 679/122323), 37303-28 (26/F66917, 679/112231, 679/122328), 37303-30 (18/192716, 18/591743, 679/122334). 2. Article Number (Lot Numbers): 32271-10 (18/061949, 18/231167, 679/062351), 32271-12 (18/048927, 26/F66235, 679/042215, 679/062342, 679/062405, 679/062489, 679/072271), 32271-14 (18/049928, 18/231165, 26/F66140, 679/042210, 679/052340, 679/052518, 679/062471, 679/062490, 679/072267), 32271-16 (18/231164, 18/311125, 26/F66141, 679/042216, 679/052519, 679/062315, 679/062483, 679/072272, 679/072401), 32271-18 (18/062950, 26/F65556, 26/F66236, 679/032261, 679/062332, 679/072402), 32271-20 (18/063951, 26/F66237, 679/052326), 32271-22 (18/064952, 26/F65557, 679/032263, 679/062354), 32271-24 (18/065953, 26/F65558, 679/062333), 32271-26 (679/032265), 32271-28 (26/F65559, 679/062357, 679/052520), 32271-32 (26/F65561), 32271-34 (26/F65562). 3. Article Number (Lot Numbers): 37304-10 (18/635486, 26/F24799, 679/012419, 86/011176), 37304-12 (18/061352, 18/521172, 18/5211721, 18/576643, 18/630487, 679/012433, 679/092207, 679/122409), 37304-14 (18/081360, 18/531173, 18/5311731, 18/577644, 18/631488, 679/012434, 679/092208, 679/122410), 37304-16 (18/541174, 18/578645, 18/632489, 679/012444, 679/092201, 679/122411), 37304-18 (18/092020, 18/551175, 18/633490, 679/022403, 679/092202), 37304-20 (18/011320, 18/561176, 18/634491, 679/012287, 679/012420, 679/082217, 86/483331), 37304-22 (18/241273, 679/012288, 679/012421), 37304-24 (18/021321, 18/571177, 679/012289, 679/012445). 4. Article Number (Lot Numbers): 21413-6 (26/F52124). 5. Article Number (Lot Numbers): 21431-ME (679/102045). 6. Article Number (Lot Numbers): 21482 (679/092020), 21483 (26/F55902). 7. Article Number (Lot Numbers): 21421-5 (26/F65377, 679/092039). 8. Article Number (Lot Numbers): 37241-10 (18/012169, 18/610104, 18/701013), 37241-12 (18/231295, 18/767108), 37241-14 (18/511016, 18/537392, 26/F24807, 679/012280), 37241-16 (18/005340, 18/121106, 18/231387, 18/521017, 18/560517), 37241-18 (18/232328, 18/531018, 679/072408, 86/311848), 37241-20 (18/233388, 18/541020, 18/562519, 679/072419, 86/287847), 37241-22 (18/066109, 18/234389, 18/551021, 26/F24594, 26/F24811), 37241-24 (18/561022, 18/564521, 18/608110, 26/F24595), 37241-26 (18/611111, 18/711023, 26/F24596, 26/F24813), 37241-28 (18/612112, 18/721024, 18/811597, 26/F24597, 86/485508), 37241-30 (18/015172, 18/613113, 18/731019, 26/F24598). 9. Article Number (Lot Numbers): 21015-08 (00/0123-12, 00/0321-43, 00/0723-07), 21015-09 (00/0123-13).

Other devices under the same FDA submission

4 devices cleared under K131722 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98341).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page