cobas pro integrated solutions with cobas c 503 analytical… recallZ-1476-2026 · Roche Diagnostics Operations, Inc. · initiated Jan 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
FDA root cause category: Software design
What the company is doing
On 1/20/26, the firm notified affected customers via written Urgent Medical Device Correction letters. Customers were provided with workaround instructions and further instructed to consult with the physician or pathologist at their facility to determine specific clinical implications for patients. This issue has been corrected with the launch of a subsequent cobas pro software version. It is mandatory to upgrade all cobas c 503 analytical units running Spline assays to the latest software version. A Roche Diagnostics Service Representative will contact affected customers to schedule an upgrade to software version 03-03.
Recall details
- Recall number
- Z-1476-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 20, 2026
- Posted by the FDA
- Feb 27, 2026
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 1261 total
- Product code
- JJE
- FDA submissions
- K191899
- Distribution
- US Nationwide distribution.
Product description
cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL
Code information (models, lots, serial numbers)
cobas pro sample supply unit, UDI-DI 07613336158852 cobas pro SSU, UDI-DI 07613336179499 all software versions previous to 03-02
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- cobas pro Sample Supply Unit 08464502001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall0 reports
- cobas pro SSU 09205632001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall0 reports
Other devices under the same FDA submission
1 device cleared under K191899 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Glucose HK 08057800190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE19 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.