cobas pro Sample Supply Unit 08464502001Recalls & reported problems

Roche Diagnostics GmbH · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 20, 2026Open, Classified

    Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.

    Z-1476-2026 · Root cause: Software design · FDA record ↗

Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices

All ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices →