DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME recallZ-1289-2026 · Olympus Corporation of the Americas · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Devices which did not undergo thermoforming could deform and lose performance.
FDA root cause category: Process control
What the company is doing
On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at [email protected] or by phone at (647) 999-3203.
Recall details
- Recall number
- Z-1289-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- Feb 5, 2026
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 1490 units (OUS only)
- Product code
- KNS
- FDA submissions
- K950166
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and International: TBD.
Product description
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Code information (models, lots, serial numbers)
Model/Catalog Number: KD-401Q-0730; Material REF: 5857030; UDI-DI: 04953170041709; All Lots with a valid expiration date
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-401Q-0730Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98252).
- Z-1268-2026 · Open, ClassifiedBrand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320; Software Version: N/A Product Description: These instruments have been …
- Z-1269-2026 · Open, ClassifiedProduct Name: Single Use Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0720 Product Description: These instruments have been designed to be used with an Oly…
- Z-1270-2026 · Open, ClassifiedProduct Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endosco…
- Z-1271-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A Product Description: These instruments have been designed to be used w…
- Z-1272-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0720; Software Version: N/A Product Description: These instruments have been designed to be used w…
- Z-1273-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0725; Software Version: N/A Product Description: These instruments have been designed to be used w…
- Z-1274-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used w…
- Z-1275-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A Product Description: These instruments have been designed to be used w…
- Z-1276-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730; Software Version: N/A Product Description: These instruments have been designed to be used w…
- Z-1277-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been…
- Z-1278-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; Software Version: N/A Product Description: These instruments have been …
- Z-1279-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; Software Version: N/A Product Description: These instruments have been …
- Z-1280-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0320; Software Version: N/A Product Description: These instruments have been …
- Z-1281-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0330; Software Version: N/A Product Description: These instruments have been …
- Z-1282-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215; Software Version: N/A Product Description: These instruments have been …
- Z-1283-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC433Q-0720; Software Version: N/A Product Description: These instruments have been …
- Z-1284-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number:KD-VC433Q-0730; Software Version: N/A Product Description: These instruments have been d…
- Z-1286-2026 · Open, ClassifiedProduct Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: These instruments have been designed to be used with an Olympus endosco…
- Z-1287-2026 · Open, ClassifiedProduct Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These instruments have been designed to be used with an Olympus endosco…
- Z-1288-2026 · Open, ClassifiedProduct Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: These instruments have been designed to be used with an Olympus endosco…
- Z-1290-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments have been designed to be used with an Olympus endoscope…
- Z-1291-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1530 Product Description: These instruments have been designed to be used with an Olympus endoscope…
- Z-1292-2026 · Open, ClassifiedProduct Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: These instruments have been designed to be used with an Olympus endoscope…
- Z-1293-2026 · Open, ClassifiedBrand Name: Disposable Triple Lumen Sphincterotome Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0320 Product Description: These instruments ha…
- Z-1294-2026 · Open, ClassifiedProduct Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0725 Product Description: These instruments have been designed to be used with an Olympus endosco…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page