DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-401Q-0730Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170041709), received between Oct 8, 2020 and Jun 16, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jan 7, 2026Class IIOpen, Classified
Devices which did not undergo thermoforming could deform and lose performance.
Z-1289-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Material Split, Cut or Torn1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2022Mechanical Problem; Material Split, Cut or Torn; Mechanical Problem; Material Split, Cut or Torn
AS REPORTED, DURING AN ENDOSCOPIC CHOLANGIO-PANCREATOGRAPHY THERAPEUTIC PROCEDURE, AT THE TIME OF THE SPHINCTEROTOMY OF THE ODDI SPHINCTER, THE WIRE IS CUT BY THE ELECTRIC ARC OF THE KNIFE. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SPHINCTEROTOME (SIMILAR DEVICE) FROM THE SAME LOT WITHOUT ANY ISSUE. NO PATIENT CONSEQUENCES. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14710619 ↗MalfunctionMay 13, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED FROM OLYMPUS (B)(4) THAT DURING SPHINCTEROTOMY PROCEDURE USING THE SUBJECT DEVICE, WHEN THE CUTTING KD WIRE BROKE. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11822988 ↗MalfunctionOct 8, 2020Break; Break
WE RECEIVED THE FOLLOWING REPORT. DURING A BILIARY PROCEDURE AND PAPILLA CUTTING, THE SUBJECT DEVICE WAS USED. THE CUTTING WIRE BROKE. ANOTHER DEVICE WAS THEN USED WITHOUT ANY PROBLEM. THERE WAS NO PATIENT INJURY REPORTED. THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. ON SEPTEMBER 16, 2020, OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOUND THAT THE CUTTING WIRE WAS BROKEN OFF, AND THAT THE BROKEN SECTION OF THE CUTTING WIRE WAS MELTED AND BURNED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10653557 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page