Centricity recallZ-1475-2026 · GE Healthcare · initiated Jan 16, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,

FDA root cause category: Device Design

What the company is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 01/16/2026 by letter delivered via traceable means. The notice explained the issue and risk, and requested the following: Actions to be taken by Customer/User: "To determine whether you have ZFP Patient Timeline configured at your site, please contact your site administrator. " If you do not have the ZFP Patient Timeline, you can continue to use Centricity Universal Viewer Zero Footprint Client (ZFP) DICOM viewer." "Review all patient studies that were reviewed using the ZFP Patient Timeline to ensure the correct images were reviewed for each patient. GE HealthCare recommends that you discontinue the use of ZFP Patient Timeline." If you do have the ZFP Patient Timeline, pending correction by GE HealthCare, please follow the instructions below: If your site requires the use of Patient Timeline, the following steps must be followed each time a different patient is loaded to ensure the proper series is displayed for this patient: 1. Change the layout to single viewport by selecting the Full view in the Layout Selector (User Manual Section 2.5.1). 2. Manually drag and drop the series from the Navigator into the viewport (User Manual Section 2.4.1). GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-1475-2026
Status
Open, Classified
Initiated by the company
Jan 16, 2026
Posted by the FDA
Feb 27, 2026
Recalling firm
GE Healthcare
Quantity
250 units
Product code
LLZ
FDA submissions
K131977
Distribution
Worldwide distribution - US Nationwide and the countries of Belarus, Brazil, Canada, China, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Lebanon, Malaysia, Netherlands, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkiye, United Arab Emirates, United Kingdom, Vietnam.

Product description

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological

Code information (models, lots, serial numbers)

Show the full code information (17k characters)

1) 2110344-039, UDI/DI 00840682102988, System ID Numbers: 105906-2-Centricity Universal Viewer Zero Footprint / ZFP-02312614, 4491-1-Centricity Universal Viewer Zero Footprint / ZFP-22180897, 10908-1-Centricity Universal Viewer Zero Footprint / ZFP-25199248, 10908-7-Centricity Universal Viewer Zero Footprint / ZFP-22203146, 10908-7-Centricity Universal Viewer Zero Footprint / ZFP-25188736, 4263-3-Centricity Universal Viewer Zero Footprint / ZFP-00106869, 4263-3-Centricity Universal Viewer Zero Footprint / ZFP-00106870, 4263-5-Centricity Universal Viewer Zero Footprint / ZFP-01300798, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25222931, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248683, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248688, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248690, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248695, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248713, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248716, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-26608643, 10458-9-Centricity Universal Viewer Zero Footprint / ZFP-26580437, 10458-9-Centricity Universal Viewer Zero Footprint / ZFP-26694204, 10458-9-Centricity Universal Viewer Zero Footprint / ZFP-26694229, 105197-2-Centricity Universal Viewer Zero Footprint / ZFP-20813950, 4642-6-Centricity Universal Viewer Zero Footprint / ZFP-25210052, 4642-6-Centricity Universal Viewer Zero Footprint / ZFP-25210057, 4642-6-Centricity Universal Viewer Zero Footprint / ZFP-25210107, 103829-6-Centricity Universal Viewer Zero Footprint / ZFP-20900298, 103829-6-Centricity Universal Viewer Zero Footprint / ZFP-21180523, 103829-6-Centricity Universal Viewer Zero Footprint / ZFP-21180525, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-26390526, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-26694221, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-26694222, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-24463042, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-25239139, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-25239140, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-26714200, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-26714202, 3471-36-Centricity Universal Viewer Zero Footprint / ZFP-24399434, 3471-36-Centricity Universal Viewer Zero Footprint / ZFP-26295812, 1198-18-Centricity Universal Viewer Zero Footprint / ZFP-02348419, 1198-18-Centricity Universal Viewer Zero Footprint / ZFP-18990433, 1198-78-Centricity Universal Viewer Zero Footprint / ZFP-22223452, 100581-1-Centricity Universal Viewer Zero Footprint / ZFP-25168130, 8746-1-Centricity Universal Viewer Zero Footprint / ZFP-23900206, 8746-1-Centricity Universal Viewer Zero Footprint / ZFP-21318104, 1207-14-Centricity Universal Viewer Zero Footprint / ZFP-24836084, 2123-19-Centricity Universal Viewer Zero Footprint / ZFP-24836053, 102901-1-Centricity Universal Viewer Zero Footprint / ZFP-23251238, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466083, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-26132903, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-26132895, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466076, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466077, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466078, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466079, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466088, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-00098228, 103696-33-Centricity Universal Viewer Zero Footprint / ZFP-20098431, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-20036715, 4578-1-Centricity Universal Viewer Zero Footprint / ZFP-25014571, 100468-1-Centricity Universal Viewer Zero Footprint / ZFP-24480189, 2440-1-Centricity Universal Viewer Zero Footprint / ZFP-22206970, 613493-Centricity Universal Viewer Zero Footprint / ZFP-26356112, 672721-Centricity Universal Viewer Zero Footprint / ZFP-25151288, 313741-Centricity Universal Viewer Zero Footprint / ZFP-26548614, 101223-6-Centricity Universal Viewer Zero Footprint / ZFP-20569855, 101223-22-Centricity Universal Viewer Zero Footprint / ZFP-20575531, 101223-22-Centricity Universal Viewer Zero Footprint / ZFP-20575532, 101223-22-Centricity Universal Viewer Zero Footprint / ZFP-20575543, 101223-16-Centricity Universal Viewer Zero Footprint / ZFP-20575533, 101223-16-Centricity Universal Viewer Zero Footprint / ZFP-20575535, 283039-Centricity Universal Viewer Zero Footprint / ZFP-02283242, 283039-Centricity Universal Viewer Zero Footprint / ZFP-02170574, 283039-Centricity Universal Viewer Zero Footprint / ZFP-01819807, 104894-1-Centricity Universal Viewer Zero Footprint / ZFP-01225123, 104894-1-Centricity Universal Viewer Zero Footprint / ZFP-01225124, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21180575, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21200874, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265320, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265321, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265322, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21200873, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265323, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01225125, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265343, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21207211, 1539601-Centricity Universal Viewer Zero Footprint / ZFP-26379461, A41676-Centricity Universal Viewer Zero Footprint / ZFP-21214076, 246573-Centricity Universal Viewer Zero Footprint / ZFP-26617853, M40455-Centricity Universal Viewer Zero Footprint / ZFP-00044391, C41917-Centricity Universal Viewer Zero Footprint / ZFP-00117236, D41917-Centricity Universal Viewer Zero Footprint / ZFP-00117239, M42001-Centricity Universal Viewer Zero Footprint / ZFP-02394845, 16921400-Centricity Universal Viewer Zero Footprint / ZFP-25194303, 509982-Centricity Universal Viewer Zero Footprint / ZFP-26608620, 509982-Centricity Universal Viewer Zero Footprint / ZFP-19670323, HC0220-Centricity Universal Viewer Zero Footprint / ZFP-01286629, HC0220-Centricity Universal Viewer Zero Footprint / ZFP-26660354, HC2720-Centricity Universal Viewer Zero Footprint / ZFP-00163499, UC2847-Centricity Universal Viewer Zero Footprint / ZFP-00260806, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-01461578, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-26703904, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-26703931, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-26703990, 600791-Centricity Universal Viewer Zero Footprint / ZFP-24944069, 600791-Centricity Universal Viewer Zero Footprint / ZFP-26859675, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-24467504, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-02286748, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624821, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624822, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624823, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624824, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624826, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624828, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624829, 351738-Centricity Universal Viewer Zero Footprint / ZFP-26304641, 434487-Centricity Universal Viewer Zero Footprint / ZFP-22169581, 722453-Centricity Universal Viewer Zero Footprint / ZFP-25287444, 560496-Centricity Universal Viewer Zero Footprint / ZFP-26748573, HCNE01-Centricity Universal Viewer Zero Footprint / ZFP-20108177, 1257845X - 434487-Centricity Universal Viewer Zero Footprint / ZFP-22203508, 2010846X - 601860-Centricity Universal Viewer Zero Footprint / ZFP-24336091, HC0565-Centricity Universal Viewer Zero Footprint / ZFP-26608610, 4625913X - HC0565-Centricity Universal Viewer Zero Footprint / ZFP-25017402, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-00001140, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666351, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666352, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666354, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666356, 4790503-Centricity Universal Viewer Zero Footprint / ZFP-26599556, 12711614-Centricity Universal Viewer Zero Footprint / ZFP-26449902, 81077-Centricity Universal Viewer Zero Footprint / ZFP-25304950, 81077-Centricity Universal Viewer Zero Footprint / ZFP-25329396, 591292-Centricity Universal Viewer Zero Footprint / ZFP-21198479, B53686-Centricity Universal Viewer Zero Footprint / ZFP-01271448, D58105-Centricity Universal Viewer Zero Footprint / ZFP-25204625, 16872968-Centricity Universal Viewer Zero Footprint / ZFP-24855625, 16872968-Centricity Universal Viewer Zero Footprint / ZFP-26929832, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24448019, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-26640843, 16999320-Centricity Universal Viewer Zero Footprint / ZFP-24833747, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24434112, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24439350, 553884-Centricity Universal Viewer Zero Footprint / ZFP-25203487, 553884-Centricity Universal Viewer Zero Footprint / ZFP-26906289, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24838806, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24305476, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-26127359, JP-283699-Centricity Universal Viewer Zero Footprint / ZFP-20978978, JP-260056-Centricity Universal Viewer Zero Footprint / ZFP-20587825, JP-330087-Centricity Universal Viewer Zero Footprint / ZFP-24931880, 851228-Centricity Universal Viewer Zero Footprint / ZFP-26136995, KW1018-Centricity Universal Viewer Zero Footprint / ZFP-26917029, KW1018-Centricity Universal Viewer Zero Footprint / ZFP-26917028, 60020-Centricity Universal Viewer Zero Footprint / ZFP-01819295, 17026623-Centricity Universal Viewer Zero Footprint / ZFP-25248193, 4510134-Centricity Universal Viewer Zero Footprint / ZFP-24525506, 17191257-Centricity Universal Viewer Zero Footprint / ZFP-26628542, 17066964-Centricity Universal Viewer Zero Footprint / ZFP-24484202, 1886799-Centricity Universal Viewer Zero Footprint / ZFP-25169256, 16951628-Centricity Universal Viewer Zero Footprint / ZFP-26453548, SA1232-Centricity Universal Viewer Zero Footprint / ZFP-25227096, SA1233-Centricity Universal Viewer Zero Footprint / ZFP-00066573, SA1013-Centricity Universal Viewer Zero Footprint / ZFP-22879437, SA1013-Centricity Universal Viewer Zero Footprint / ZFP-24281567, 17111683-Centricity Universal Viewer Zero Footprint / ZFP-21733830, 82667-Centricity Universal Viewer Zero Footprint / ZFP-25194300, 17021465-Centricity Universal Viewer Zero Footprint / ZFP-24487930, 747887-Centricity Universal Viewer Zero Footprint / ZFP-25210479, MPX431-Centricity Universal Viewer Zero Footprint / ZFP-24465830, MEX255-Centricity Universal Viewer Zero Footprint / ZFP-26802528, GED649-Centricity Universal Viewer Zero Footprint / ZFP-19666476, 947379-Centricity Universal Viewer Zero Footprint / ZFP-24846063, O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-20618153, 228493-Centricity Universal Viewer Zero Footprint / ZFP-24821472, 407237-Centricity Universal Viewer Zero Footprint / ZFP-20997395, 730140-Centricity Universal Viewer Zero Footprint / ZFP-00038193, 0853550198-Centricity Universal Viewer Zero Footprint / ZFP-02093675, 6051-Centricity Universal Viewer Zero Footprint / ZFP-00116181, 6002-Centricity Universal Viewer Zero Footprint / ZFP-00158571, 486961-Centricity Universal Viewer Zero Footprint / ZFP-25022095, AE1564-Centricity Universal Viewer Zero Footprint / ZFP-00555894, AE1564-Centricity Universal Viewer Zero Footprint / ZFP-26560468, RWEAE-Centricity Universal Viewer Zero Footprint / ZFP-00165095, 243119-Centricity Universal Viewer Zero Footprint / ZFP-21322536, 243119-Centricity Universal Viewer Zero Footprint / ZFP-26631095, 370259-Centricity Universal Viewer Zero Footprint / ZFP-25309376, 370259-Centricity Universal Viewer Zero Footprint / ZFP-26784611, 17053632-Centricity Universal Viewer Zero Footprint / ZFP-24829942, 6345592-Centricity Universal Viewer Zero Footprint / ZFP-24825373, 17152365-Centricity Universal Viewer Zero Footprint / ZFP-26632357, 17040488-Centricity Universal Viewer Zero Footprint / ZFP-26404455. 2) 2110344-046, UDI/DI 00840682102988, System ID Numbers: 6579-11-Centricity Universal Viewer Zero Footprint / ZFP-26923415, 10458-1-Centricity Universal Viewer Zero Footprint / ZFP-26898336, 3886-4-Centricity Universal Viewer Zero Footprint / ZFP-22180910, 3886-4-Centricity Universal Viewer Zero Footprint / ZFP-22180906, 9094-1-Centricity Universal Viewer Zero Footprint / ZFP-26705438, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839308, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839310, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839311, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839312, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839313, 4326-2-Centricity Universal Viewer Zero Footprint / ZFP-26715184, 10190-9-Centricity Universal Viewer Zero Footprint / ZFP-26850722, 4084-1-Centricity Universal Viewer Zero Footprint / ZFP-26747422, 103696-30-Centricity Universal Viewer Zero Footprint / ZFP-21237479, 4204-1-Centricity Universal Viewer Zero Footprint / ZFP-26884848, 4204-1-Centricity Universal Viewer Zero Footprint / ZFP-26884853, 17094888-Centricity Universal Viewer Zero Footprint / ZFP-26918064, 268882-Centricity Universal Viewer Zero Footprint / ZFP-26747414, 268882-Centricity Universal Viewer Zero Footprint / ZFP-26769258, 268882-Centricity Universal Viewer Zero Footprint / ZFP-26769261, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388499, 104991-4-Centricity Universal Viewer Zero Footprint / ZFP-21350494, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388493, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388494, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388495, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388497, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388498, 104991-5-Centricity Universal Viewer Zero Footprint / ZFP-24388490, 104991-5-Centricity Universal Viewer Zero Footprint / ZFP-24388491, 104991-5-Centricity Universal Viewer Zero Footprint / ZFP-24388492, A91901-Centricity Universal Viewer Zero Footprint / ZFP-24484469, 145790 - HC0399-Centricity Universal Viewer Zero Footprint / ZFP-24374380, HC0014-Centricity Universal Viewer Zero Footprint / ZFP-21359919, 255473-Centricity Universal Viewer Zero Footprint / ZFP-26865306, JP-280250-Centricity Universal Viewer Zero Footprint / ZFP-26671908, JP-283699-Centricity Universal Viewer Zero Footprint / ZFP-26689509, JP-280250-Centricity Universal Viewer Zero Footprint / ZFP-02360544, JP-272375-Centricity Universal Viewer Zero Footprint / ZFP-26918260, 347891-Centricity Universal Viewer Zero Footprint / ZFP-23297397, 347891-Centricity Universal Viewer Zero Footprint / ZFP-26880666, 17195358-Centricity Universal Viewer Zero Footprint / ZFP-23986377, RU1086-Centricity Universal Viewer Zero Footprint / ZFP-26799525, RU1282-Centricity Universal Viewer Zero Footprint / ZFP-26800953, 17121345-Centricity Universal Viewer Zero Footprint / ZFP-26796880, RU8902-Centricity Universal Viewer Zero Footprint / ZFP-17775055, SA2591-Centricity Universal Viewer Zero Footprint / ZFP-26880655, 521920-Centricity Universal Viewer Zero Footprint / ZFP-26912759, 659749-Centricity Universal Viewer Zero Footprint / ZFP-26824375, ZA2239-Centricity Universal Viewer Zero Footprint / ZFP-26805562, 16959088-Centricity Universal Viewer Zero Footprint / ZFP-26777223, GEX184-Centricity Universal Viewer Zero Footprint / ZFP-26368616, 242833-Centricity Universal Viewer Zero Footprint / ZFP-26825072, 80182-Centricity Universal Viewer Zero Footprint / ZFP-26876144, 16882165-Centricity Universal Viewer Zero Footprint / ZFP-26673881, 16880927-Centricity Universal Viewer Zero Footprint / ZFP-26818090. 3) 2110344-047, UDI/DI 00840682102988, System ID Numbers: Local DB site#000330-Centricity Universal Viewer Zero Footprint / ZFP-22138067, Local DB site#000330-Centricity Universal Viewer Zero Footprint / ZFP-22138066, JP-341108-Centricity Universal Viewer Zero Footprint / ZFP-25012607, JP-220062-Centricity Universal Viewer Zero Footprint / ZFP-24475044.