EVO/EVO+VISIAN Implantable Collamer Lens recallZ-1480-2026 · Staar Surgical AG · initiated Jan 21, 2026
Class II Open, Classified All recalls posted in March 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Staar Surgical AG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
FDA root cause category: Process control
What the company is doing
On 1/21/2026, a Field Safety Notice was mailed to customers who were asked to do the following: 1. If any of the devices identified in the Field Safety Notice Return Response form have not already been implanted, remove them from your inventory and quarantine the product. 2. Complete the Field Safety Notice Return Response form indicating if each device has been implanted or is in your possession. 3. Complete and email the Field Safety Notice Return Response form to [email protected]. 5. Return affected devices to the firm. Should you have any questions, contact the firm's Representative or Customer Service at [email protected] or by phone at +41323328888 Monday through Friday, 8:00am to 5:00pm CET.
Recall details
- Recall number
- Z-1480-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jan 21, 2026
- Posted by the FDA
- Mar 2, 2026
- Recalling firm
- Staar Surgical AG
- Quantity
- 7
- Product code
- QCB
- FDA submissions
- P030016
- Distribution
- International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Product description
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Code information (models, lots, serial numbers)
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EVO/EVO+VISIAN Implantable Collamer Lens VTICM5 12.1 (-5.0/1.0)Staar Surgical Company · Phakic Toric Intraocular Lens1 recall4 reports
Product lines
Other devices under the same FDA submission
1,416 devices cleared under P030016 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall12 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-11.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-10.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall7 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-10.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall7 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-12.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page