Lifepak 15, Ref: 99577-000025, 99577-000047, 99577-001256… recallZ-2065-2026 · Physio-Control, Inc. · initiated Apr 13, 2026
Class II Open, Classified All recalls posted in May 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Physio-Control, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
FDA root cause category: Process control
What the company is doing
On 04/13/2026, recall notices were sent to customers who were asked to do the following: 1) Return the business reply form by email to [email protected] to confirm receipt of this notification. 2) Upon receipt of the completed business reply form, firm will contact you to arrange for corrective action on the device if service has not been performed. 3) Continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter "Maintaining the Equipment" subsection "General Maintenance and Testing". 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until the correction has been completed. 5) Inform firm if any of the subject devices have been distributed to other organizations. If so, provide contact details, so the firm can inform the recipients. If you have any questions or concerns, please contact the firm's Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday - Friday or by email at [email protected]
Recall details
- Recall number
- Z-2065-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 13, 2026
- Posted by the FDA
- May 4, 2026
- Recalling firm
- Physio-Control, Inc.
- Quantity
- 206
- Product code
- MKJ
- FDA submissions
- P160026
- Distribution
- US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.
Product description
LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
Device/UDI/Serial Number: LIFEPAK 15/0088387391164815011543166573, 0088387387169015033143394891, 0088387387169015050743524360, 0088387387169015092843914574, 0088387387204815093043915122, 0088387387169016010444187447, 0088387387162116012044252611, 0088387387169016012644262721, 0088387387169016012644277082, 0088387387169016012744282280, 0088387387171316060244574340, 0088387387171316080444896329, 0088387387169016083144962804, 0088387387170617013145447229, 0088387387169017041245743786, 0088387387171317050945834209, 0088387387171317062845998142, 0088387387171317082846103043, 0088387387169017101046303288, 0088387387171317103146303092, 0088387387162117122146587130, 0088387387171317122646602003, 0088387387171317122846608783, 0088387387171317122946614788, 0088387387171318022246813507, 0088387387169018032946966267, 0088387387171318061947337396, 0088387387169018070547401928, 0088387387171318092747611441, 0088387387169018112947784931, 0088387387169018112947786317, 0088387387171318121847847661, 0088387387171318121847848950, 0088387387708118121947845317, 0088387387708118121947855371, 0088387387169018122047857172, 0088387387169018122047858660, 0088387387171318122847872868, 0088387387171319031248018370, 0088387387169019061248247559, 0088387387168319062748298704, 0088387387169019100848453060, 0088387387171319101448455890, 0088387387171319121848586351, 0088387387171319121948586599, 0088387387168320012848651507, 0088387387171320030548701689, 0088387387708120042848784975, 0088387387171320071648887706, 0088387387168320081748921846, 0088387387169020092448981398, 0088387387169020102649032288, 0088387387169020102649050140, 0088387387162120110249055931, 0088387387168320122949156402, 0088387387169021011149162067, 0088387387169021011149167463, 0088387387169021013049212578, 0088387387169021020949225869, 0088387387169021020949226060, 0088387387176821021949240566, 0088387387168321033149345530, 0088387387169021042149378981, 0088387387169021042649381781, 0088387387169021042749381576, 0088387387168321081849559854, 0088387387168321081949560127, 0088387387164521091649607224, 0088387387171321102149662806, 0088387377697122070750045009, 0088387387169022090150136719, 0088387387171322090850152289, 0088387387708122091350158293, 0088387387169022101950223934, 0088387387708122112150288760, 0088387387171322120150307353, 0088387387169023062850629377, 0088387387169023072450653772, 0088387387169023072450667197, 0088387387708123072850669746, 0088387387708123082850714419, 0088387387171323082950720588, 0088387387169023112150850129, 0088387387708123122050883524, 0088387388434824052251092313, 0088387387170624121151422867, 0088387387170624122351440213, 0088387387170624122351444039, 0088387388434824052251092313, 0088387387169019100848453060, 0088387387201715090843839264, 0088387387171318121847850793/ 38523489, 39736196, 40465345, 40717931, 40765749, 41130420, 41139184, 41420681, 41446039, 41446693, 41704583, 42221131, 42920956, 42921954, 43166573, 43394891, 43524360, 43914574, 43915122, 44187447, 44252611, 44262721, 44277082, 44282280, 44574340, 44896329, 44962804, 45447229, 45743786, 45834209, 45998142, 46103043, 46303288, 46303092, 46587130, 46602003, 46608783, 46614788, 46813507, 46966267, 47337396, 47401928, 47611441, 47784931, 47786317, 47847661, 47848950, 47845317, 47855371, 47857172, 47858660, 47872868, 48018370, 48247559, 48298704, 48453060, 48455890, 48586351, 48586599, 48651507, 48701689, 48784975, 48887706, 48921846, 48981398, 49032288, 49050140, 49055931, 49156402, 49162067, 49167463, 49212578, 49225869, 49226060, 49240566, 49345530, 49378981, 49381781, 49381576, 49559854, 49560127, 49607224, 49662806, 50045009, 50136719, 50152289, 50158293, 50223934, 50288760, 50307353, 50629377, 50653772, 50667197, 50669746, 50714419, 50720588, 50850129, 50883524, 51092313, 51422867, 51440213, 51444039, 51092313, 48453060, 43839264, 47850793, 38369584. LIFEPAK 1000/0088507469887815020943235807, 0088507469895315031343346860, 0088507469887815111944068667, 0088507469887815111944068669, 0088507469895316011844239775, 0088507469887816041444539726, 0088507469887817061745966118, 0088507469887817061745966121, 0088507469887817061745966122, 0088507469887817061745966126, 0088507469887817061945966131, 0088507469887817062945994506, 0088507469887817062946004311, 0088507469887817063046012479, 0088507469887818060547291148, 0088387381546519080648366102, 0088387381545820092948987376, 0088387381545821011249174469, 0088387381545821081349544565, 0088387381545821090249586700, 0088387381546521122249752235, 0088387381545822012649794992, 0088387381545822032949882696, 0088387381545822032949882698, 0088387381545822032949882819/ 34933095, 35096556, 35184465, 36153943, 36199099, 36560529, 36560531, 37119097, 37226402, 37330295, 38074718, 38074722, 38074727, 38074737, 38074741, 38074744, 38074745, 38074747, 38074748, 38074749, 38077938, 38077939, 38077940, 38077941, 38077944, 38378802, 38513659, 38656117, 38656121, 38656122, 38957375, 38957392, 38957397, 38957399, 38957402, 38972629, 38972677, 39775405, 39775406, 39775407, 39775415, 39964908, 40323073, 40412729, 40541187, 40584085, 40594714, 40707896, 40725568, 40725569, 40725571, 40725572, 40725575, 40725584, 40725582, 40725583, 41306695, 41692754, 41863786, 42504017, 42520778, 42803174, 42921847, 43235807, 43346860, 44068667, 44068669, 44239775, 44539726, 45966118, 45966121, 45966122, 45966126, 45966131, 45994506, 46004311, 46012479, 47291148, 48366102, 48987376, 49174469, 49544565, 49586700, 49752235, 49794992, 49882696, 49882698, 49882819. LIFEPAK 35/0088387377265224042251041190, 0088387377265224041851042010, 0088387377265224042251044581, 0088387376560924071151156570, 0088387376560924071151157072, 0088387376560924071151157449, 0088387376560924080651182632, 0088387376560924091251263526, 0088387376560924121051417585, 0088387376560925020451496956/ 51041190, 51042010, 51044581, 51156570, 51157072, 51157449, 51182632, 51263526, 51417585, 51496956. LIFEPAK 20e/0088387385386320082448935344, 0088387385386320082448935408, 0088387385386321080249537075/ 48935344, 48935408, 49537075
Other devices under the same FDA submission
27 devices cleared under P160026 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 death228 reports
- LIFEPAK 1000 defibrillator 99425Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 death188 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)4 deaths173 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)4 deaths129 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)124 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)5 deaths101 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)91 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)3 deaths78 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)77 reports
- LIFEPAK 1000 defibrillator 99425Physio-Control, Inc. · Automated external defibrillators (non-wearable)76 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall3 reports
- LIFEPAK 15 monitor/defibrillator 97577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.