Centricity recallZ-1570-2026 · GE Medical Systems, LLC · initiated Jan 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.

FDA root cause category: Software design

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 1/30/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue using your device by following the instructions below: Ensure workstations have appropriate security controls per product manuals. Implement network account authentication by using Active Directory/LDAP services for user management. If network account authentication is not possible at your site, please contact GE HealthCare Service to request temporary steps to mitigate the issue. Ensure all potential users in your facility are informed of this safety notification. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-1570-2026
Status
Open, Classified
Initiated by the company
Jan 30, 2026
Posted by the FDA
Mar 16, 2026
Recalling firm
GE Medical Systems, LLC
Quantity
797 units
Product code
LLZ
FDA submissions
K150420
Distribution
Worldwide - US Nationwide distribution.

Product description

Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2100169-001, 2088026-985, 2088026-971, 2088026-614, 2088026-713, 2088026-994, 2100491-006, 2088026-723, 2088026-308, 2104867-044, 2100491-012, 2088026-962, 2088026-969, 2102676-001, 2088026-936, 2088026-960, K1220JWED, 2088026-806, 2088026-506, 2088026-026, 2088026-970, 2088026-943, 2088026-990, 2088026-940, 2088026-959, 2088026-967, 2088026-989, 2104867-045, 2088026-993, 2088026-306, 2102675-010, 2088026-948, 2088026-957, 2088026-973, 2088026-946, 2088026-715, 2088026-906

Code information (models, lots, serial numbers)

Show the full code information (33k characters)

UDI Numbers: (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP90172094097001H (01)00840682103800(10)60SP90172094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP102094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90112094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60SP90162094097001G (01)00840682103800(10)60P90110.2094097001H (01)00840682103800(10)60P90110.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP90162094097001G (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)SP32094097001C (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)SP32094097001C (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP90172094097001H (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60SP5.0.22094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.22094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60SP90142094097001G (01)00840682103800(10)60SP90142094097001G (01)00840682103800(10)60SP90142094097001G (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90152094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP102094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60SP10.12094097001G (01)00840682103800(10)60SP10.12094097001G (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90132094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)SP32094097001C (01)00840682103800(10)SP32094097001C (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)SP32094097001C (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.22094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.22094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7022094097001D (01)00840682103800(10)6.0SP7022094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)SP32094097001C (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)SP32094097001C (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)6.0SP92094097001D (01)00840682103800(10)60P10412094097001

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98428).