Flexible Cryoprobe (OD 2.4mm, L1.15mm) recallZ-1568-2026 · Erbe USA Inc · initiated Feb 12, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Erbe USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Probes may rupture/burst during activation

FDA root cause category: Process control

What the company is doing

On May 5, 2026, ERBE USA issued a "Urgent Medical Device Recall Expansion notification via E-Mail. The expansion was issued due to new lot numbers that may cause the same failure mode. The original notification was issued on Feb. 12, 2026. ERBE asked consignees to take the following actions: 1. Discontinue use of the products with the affected lot numbers. 2. Examine your inventory, remove, and quarantine any identified devices with the affected lot numbers 3. Ensure all applicable personnel within the vicinity of device activation carefully read the content of this notification. If affected product has been transferred to a different location, please ensure that location and personnel receive a copy of this notification. 4. Please contact Erbe at [email protected] or 678-483-8499, to obtain a Return Material Authorization number 5. Please complete the Response Form and return to Erbe 6. If you have further distributed this product, identify your locations, and forward this notification to them. 7. Erbe requests you to report any complaints, including any instances of ruptures/bursts, to our Technical Service Department at 770-955-4400. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program.

Recall details

Recall number
Z-1568-2026
Status
Open, Classified
Initiated by the company
Feb 12, 2026
Posted by the FDA
Mar 20, 2026
Recalling firm
Erbe USA Inc
Quantity
3603 units
Product code
GEH
FDA submissions
K190651
Distribution
US Nationwide distribution, including Puerto Rico.

Product description

Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use

Code information (models, lots, serial numbers)

UDI: 04050147021846/ Expanded Lots:WO472498 WO472499 WO472615 WO472616 WO473794 WO473795 WO474908 WO474909 WO474910 WO475357 WO475358 WO477030 WO477513 WO477514 WO478643 WO478646 WO478647 WO478648 WO478843 WO478844;Initial Lots: WO461840 WO461846 WO462097 W2462284 WO463327 WO463328 WO463388 W2463594 WO464285 WO464286 WO465063 WO465064 W4465351 W2465601 W2465689 W2465749 W2465769 WO467873 WO467874 WO467875 WO468648 WO468649 WO469364 W2459861 W2459862 W2459863

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K190651 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98429).