ERBE 10402-000Recalls & reported problems
Erbe Elektromedizin GmbH · ERBECRYO® 2 Cryosurgical unit · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K190651); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jun 26, 2026Open, Classified
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Z-2938-2026 · Root cause: Device Design · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190651. One submission often covers many models, so check the product and lot details in each record.
- Z-1568-2026 · Feb 12, 2026 · Open, ClassifiedFlexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
- Z-1567-2026 · Feb 12, 2026 · Open, ClassifiedFlexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
- Z-1566-2026 · Feb 12, 2026 · Open, ClassifiedFlexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
Other Unit, Cryosurgical, Accessories devices
- IceFORCE® 2.1 CX 90° Cryoablation Needle FPRPR3602Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls4 deaths274 reports
- IceRod® 1.5 CX 90° Cryoablation Needle FPRPR3533Boston Scientific Corporation · Unit, Cryosurgical, Accessories1 recall168 reports
- IcePearl® 2.1 CX 90° Cryoablation Needle FPRPR3601Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls78 reports
- IceFORCE™ 2.1 CX L 90° Cryoablation Needle FPRPR3618Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls6 reports
- IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls3 reports
- IcePearl™ 2.1 CX L 90° Cryoablation Needle FPRPR3617Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls1 report
- IcePearl® 2.1 CX Prostate Cryoablation Kit / Visual-ICE® System FPRPR3606Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
- IceFORCE® 2.1 CX Prostate Cryoablation Kit / Visual-ICE® System FPRPR3605Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.