AccuVu recallZ-1875-2026 · Angiodynamics, Inc. · initiated Mar 3, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Angiodynamics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On March 3, 2026 URGENT VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: IMMEDIATELY: 1. Stop using the devices subject to this recall and share the notification to all users of the deviceswithin your facility immediately. 2. Provide a copy of this recall notification to all sites or clinics to which you have distributed affectedproduct. 3. Remove any affected items/lots from your inventory (whether in labs, Central Supply, Shipping andReceiving or ANY other location). 4. Segregate this product in a secure location for return to AngioDynamics, Inc. 5. Package and Return the Recalled Product.
Recall details
- Recall number
- Z-1875-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 3, 2026
- Posted by the FDA
- Apr 17, 2026
- Recalling firm
- Angiodynamics, Inc.
- Quantity
- 105 units (21 boxes)
- Product code
- DQO
- FDA submissions
- K161596, K112452
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, China, Chile, Curacao, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, New Zealand, Panama, Slovenia, South Africa, Spain, Sweden, Thailand.
Product description
AccuVu Angiographic Catheter, Pigtail, Non-Braided, With 21 R/O Markers; ACCU-VU PT 4F X 70CM 035 NB 10SH 21 R/O 1CM; Catalog No.: 13709804; Product/UPN No.: H787137098045 (Box), H787137098040 (Pouch); Box Quantity: 5 pouches;
Code information (models, lots, serial numbers)
ACCU-VU PT 4F X 70CM 035 NB 10SH 21 R/O 1CM; Catalog No.: 13709804; Product/UPN No.: H787137098045 (Box), H787137098040 (Pouch); UDI-DI: 25051684006085(Box), 15051684006088(Pouch); Lot No.: A2825131;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AccuVu 13709804Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall2 reports
Product lines
Other devices under the same FDA submission
66 devices cleared under K161596, K112452 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Soft-Vu 10732301Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall6 reports
- Soft-Vu 10732701Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall2 reports
- Soft-Vu 10714009Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall1 report
- Soft-Vu 10734301Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall1 report
- Soft-Vu 10707103Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10710013Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10714001Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10714002Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10714011Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10714014Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10714016Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
- Soft-Vu 10714017Angiodynamics, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98463).
- Z-1856-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 100CM 035 NB 10SH; Catalog No.: 10707103; Product/UPN No.: H787107071035 (Box), H787107071030 (Pouch); Box Quan…
- Z-1857-2026 · Open, ClassifiedSoft -Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 40CM 038 NB 0SH; Catalog No.: 10710013; Product/UPN No.: H787107100135 (Box), H787107100130 (Pouch); Box Quanti…
- Z-1858-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 038 NB 6SH; Catalog No.: 10714001; Product/UPN No.: H787107140015 (Box), H787107140010 (Pouch); Box Qua…
- Z-1859-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 65CM 038 NB 10SH; Catalog No.: 10714002; Product/UPN No.: H787107140025 (Box), H787107140020 (Pouch); Box Quant…
- Z-1860-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Berenstein, Non-Braided; SOFT-VU BER 4F X 65CM 038 NB 0SH; Catalog No.: 10714009; Product/UPN No.: H787107140095 (Box), H787107140090 (Pouch); Box Qu…
- Z-1861-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 65CM 038 NB 0SH; Catalog No.: 10714011; Product/UPN No.: H787107140115 (Box), H787107140110 (Pouch); Box Qua…
- Z-1862-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014; Product/UPN No.: H787107140145 (Box), H787107140140 (Pouch); Box Quantit…
- Z-1863-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Berenstein, Non-Braided; SOFT-VU BER 4F X 100CM 038 NB 0SH; Catalog No.: 10714016; Product/UPN No.: H787107140165 (Box), H787107140160 (Pouch); Box Q…
- Z-1864-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Cobra (1), Non-Braided; SOFT-VU CB1 4F X 100CM 038 NB 0SH; Catalog No.: 10714017; Product/UPN No.: H787107140175 (Box), H787107140170 (Pouch); Box Qu…
- Z-1865-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 100CM 038 NB 0SH; Catalog No.: 10714018; Product/UPN No.: H787107140185 (Box), H787107140180 (Pouch); Box Qu…
- Z-1866-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 100CM 038 NB 0SH; Catalog No.: 10714021; Product/UPN No.: H787107140215 (Box), H787107140210 (Pouch); Box Quanti…
- Z-1867-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Sos Omni Selective (1), Non-Braided; SOFT-VU SO1 4F X 80CM 038 NB 0SH; Catalog No.: 10714024; Product/UPN No.: H787107140245 (Box), H787107140240 (Po…
- Z-1868-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Headhunter (1), Non-Braided; SOFT-VU HH1 4F X 100CM 038 NB 0SH; Catalog No.: 10714040; Product/UPN No.: H787107140405 (Box), H787107140400 (Pouch); B…
- Z-1869-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Straight Art, Non-Braided; SOFT-VU STR ART 4F X 65CM 035 NB 0SH; Catalog No.: 10731401; Product/UPN No.: H787107314015 (Box), H787107314010 (Pouch); …
- Z-1870-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 035 NB 6SH; Catalog No.: 10732301; Product/UPN No.: H787107323015 (Box), H787107323010 (Pouch); Box Qua…
- Z-1871-2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 40CM 035 NB 0SH; Catalog No.: 10734301; Product/UPN No.: H787107343015 (Box), H787107343010 (Pouch); Box Quantit…
- Z-1872-2026 · Open, ClassifiedAccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 2 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 2 R/O 2CM; Catalog No.: 13709005; Product/UPN No.: H787137090055 (Box), H…
- Z-1873-2026 · Open, ClassifiedAccuVu Angiographic Catheter, Straight Flush, Non-Braided, With 10 R/O Markers; ACCU-VU STR FL 4F X 100CM 035 NB 6SH 10 R/O 6@1CM 4@5CM; Catalog No.: 13709604; Product/UPN No.: H78…
- Z-1874-2026 · Open, ClassifiedAccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box),…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page