Skip to content

HeartSine samaritan PAD (Public Access Defibrillator) recallZ-3270-2026 · HeartSine Technologies Ltd · initiated Jul 27, 2026

Class III Open, Classified All recalls posted in September 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from HeartSine Technologies Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.

FDA root cause category: Process control

What the company is doing

On 07/28/2026, Stryker sent several versions of an Urgent: Medical device recall letter that outlines two issues: 1) An issue with the membrane tail, an internal component, could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. 2) A limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information. For the missing required maintenance insert (issue 2), Customers are instructed to: 1. Identify impacted devices by verifying if your device serial numbers are included in the list at the link https://heartsinerecall.com/HeartSineSamaritanPAD/LookUp. For this field action notice, select AED + Pad-Pak. 2. Download a copy of the missing maintenance insert when acknowledging the recall and store with your devices user manual until a replacement is received. For questions or concerns, contact Stryker Customer Service at +1 8007879537, option 2, from 8 AM to 7 PM Eastern Time, Monday to Friday, your Authorized Distributor, or HeartSine Technologies Technical Support at [email protected].

Recall details

Recall number
Z-3270-2026
Recall class
Class III
Status
Open, Classified
Initiated by the company
Jul 27, 2026
Posted by the FDA
Sep 25, 2026
Recalling firm
HeartSine Technologies Ltd
Quantity
31,834
Product code
MKJ
FDA submissions
P160008
Distribution
U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.

Product description

HeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (American Airlines) - 2 Min CPR UDI-DI code: 0506016712207 350P US Device, US English with Gateway - 2 Min CPR UDI-DI code: 0506016712804 HeartSine SAM 360P 360P US Device, US English - 2 Min CPR UDI-DI code: 0506016712068 HeartSine SAM 450P SAM 450P US Device, US English (2 min CPR) UDI-DI code: 0506016712069

Code information (models, lots, serial numbers)

Show the full code information (81k characters)

Product - 350P US Device, US English - 2 Min CPR Item code: 350-STR-US-10 UDI-DI code: 0506016712067 Serial Numbers: 24D98056051D24 24D98057467D24 24D98057468D24 24D98057469D24 24D98068091E24 24D98068097E24 24D98070022F24 24D98070023F24 24D98070024F24 24D98070025F24 24D98070026F24 24D98070027F24 24D98070028F24 24D98070029F24 24D98070030F24 24D98070031F24 24D98070032F24 24D98070033F24 24D98070034F24 24D98070035F24 24D98070036F24 24D98070037F24 24D98070038F24 24D98070040F24 24D98070041F24 24D98070042F24 24D98070043F24 24D98070046F24 24D98070047F24 24D98070048F24 24D98070049F24 24D98070050F24 24D98070051F24 24D98070052F24 24D98070053F24 24D98070054F24 24D98070055F24 24D98070056F24 24D98070057F24 24D98070058F24 24D98070060F24 24D98070061F24 24D98070062F24 24D98070063F24 24D98070064F24 24D98070065F24 24D98070066F24 24D98070068F24 24D98070069F24 24D98070071F24 24D98070072F24 24D98070074F24 24D98070075F24 24D98070076F24 24D98070077F24 24D98070078F24 24D98070079F24 24D98070080F24 24D98070081F24 24D98070082F24 24D98070083F24 24D98070084F24 24D98070085F24 24D98070087F24 24D98070088F24 24D98070089F24 24D98070091F24 24D98070092F24 24D98070093F24 24D98070094F24 24D98070095F24 24D98070096F24 24D98070097F24 24D98070100F24 24D98070102F24 24D98070103F24 24D98070104F24 24D98070107F24 24D98070108F24 24D98070109F24 24D98070110F24 24D98070111F24 24D98070114F24 24D98070115F24 24D98070117F24 24D98070118F24 24D98070119F24 24D98070120F24 24D98070121F24 24D98070122F24 24D98070123F24 24D98070126F24 24D98070127F24 24D98070128F24 24D98070129F24 24D98070130F24 24D98070132F24 24D98070133F24 24D98070134F24 24D98070135F24 24D98070136F24 24D98070138F24 24D98070139F24 24D98070140F24 24D98070141F24 24D98070143F24 24D98070144F24 24D98070145F24 24D98070148F24 24D98070149F24 24D98070150F24 24D98070151F24 24D98070157F24 24D98070160F24 24D98070161F24 24D98070162F24 24D98070163F24 24D98070166F24 24D98056785D24 24D98057279D24 24D98057648D24 24D98053900E24 24D98054669E24 24D98054543E24 24D98054034E24 24D98050678E24 24D98050536E24 24D98055040E24 24D98055328E24 24D98050945E24 24D98055733E24 24D98070147F24 24D98070039F24 24D98070101F24 24D98070105F24 24D98070146F24 24D98070045F24 24D98070067F24 24D98070116F24 24D98070099F24 24D98070044F24 24D98070098F24 24D98070059F24 24D98070073F24 24D98070090F24 24D98070142F24 24D98070124F24 24D98070113F24 24D98070131F24 24D98070112F24 24D98070125F24 24D98070412F24 24D98026634F24 24D98026619F24 24D98048208F24 24D98026934F24 24D98026936F24 24D98048576F24 24D98048559F24 24D98048502F24 24D98048594F24 24D98048571F24 24D98048600F24 24D98048566F24 24D98048609F24 24D98048588F24 24D98048578F24 24D98048554F24 24D98048499F24 24D98048503F24 24D98048504F24 24D98048574F24 24D98048582F24 24D98048491F24 24D98048556F24 24D98048597F24 24D98048598F24 24D98048570F24 24D98048712F24 24D98048560F24 24D98048500F24 24D98048595F24 24D98048575F24 24D98048709F24 24D98048603F24 24D98048495F24 24D98048617F24 24D98048492F24 24D98048497F24 24D98048558F24 24D98048605F24 24D98048710F24 24D98048565F24 24D98048586F24 24D98048604F24 24D98048561F24 24D98048606F24 24D98048589F24 24D98048555F24 24D98048564F24 24D98048614F24 24D98048599F24 24D98048590F24 24D98048591F24 24D98048583F24 24D98048601F24 24D98048580F24 24D98048573F24 24D98048618F24 24D98048577F24 24D98048567F24 24D98048501F24 24D98048610F24 24D98048607F24 24D98048711F24 24D98048557F24 24D98048616F24 24D98048585F24 24D98048572F24 24D98048608F24 24D98048592F24 24D98048593F24 24D98048569F24 24D98048619F24 24D98048579F24 24D98048602F24 24D98048498F24 24D98048496F24 24D98048581F24 24D98048587F24 24D98048562F24 24D98048596F24 24D98048611F24 24D98048568F24 24D98048615F24 24D98048584F24 24D98048563F24 24D98026922F24 24D98026921F24 24D98048625F24 24D98026935F24 24D98026838F24 24D98026836F24 24D98026834F24 24D98026928F24 24D98026868F24 24D98026869F24 24D98026933F24 24D98048388F24 24D98028430G24 24D98062685G24 24D98071182G24 24D98071082G24 24D98038091G24 24D98071057G24 24D98038146G24 24D98038133G24 24D98074601G24 24D98074572G24 24D98074570G24 24D98074606G24 24D98074541G24 24D98024371H24 24D98024818H24 24D98024808H24 24D98002300H24 24D98002281H24 24D98024650H24 24D98019809H24 24D98019376H24 24D98040257G24 24D98040243G24 24D98040260G24 24D98040262G24 24D98040258G24 24D98040256G24 24D98040263G24 24D98040261G24 24D98001851H24 24D98038364G24 24D98071566G24 24D98071559G24 24D98071562G24 24D98038340G24 24D98038333G24 24D98038334G24 24D98038326G24 24D98071497G24 24D98071498G24 24D98071502G24 24D98019747H24 24D98038331G24 24D98038335G24 24D98038330G24 24D98071567G24 24D98071568G24 24D98038319G24 24D98038318G24 24D98038315G24 24D98038311G24 24D98038304G24 24D98038338G24 24D98038322G24 24D98038306G24 24D98038302G24 24D98038309G24 24D98038303G24 24D98038308G24 24D98071563G24 24D98038352G24 24D98038339G24 24D98038321G24 24D98038314G24 24D98038313G24 24D98038320G24 24D98038323G24 24D98071501G24 24D98038387G24 24D98071499G24 24D98071573G24 24D98071571G24 24D98071581G24 24D98071558G24 24D98071560G24 24D98038295G24 24D98071561G24 24D98071565G24 24D98071564G24 24D98071500G24 24D98071495G24 24D98071496G24 24D98071570G24 24D98038139G24 24D98038307G24 24D98038324G24 24D98038317G24 24D98038316G24 24D98038305G24 24D98038312G24 24D98038293G24 24D98038300G24 24D98038301G24 24D98038297G24 24D98038144G24 24D98038145G24 24D98038337G24 24D98038344G24 24D98038343G24 24D98038345G24 24D98038327G24 24D98038342G24 24D98038296G24 24D98038289G24 24D98038294G24 24D98038290G24 24D98038299G24 24D98064983E24 24D98064976E24 24D98064990E24 24D98050963E24 24D98067351E24 24D98067347E24 24D98067346E24 24D98055977E24 24D98055973E24 24D98055974E24 24D98050964E24 24D98067337E24 24D98055964E24 24D98055976E24 24D98060043E24 24D98055967E24 24D98055972E24 24D98055969E24 24D98055978E24 24D98055966E24 24D98055981E24 24D98055968E24 24D98055971E24 24D98055979E24 24D98055982E24 24D98055970E24 24D98050987E24 24D98050978E24 24D98055940E24 24D98050966E24 24D98050967E24 24D98050956E24 24D98050981E24 24D98067336E24 24D98067335E24 24D98050982E24 24D98050977E24 24D98050988E24 24D98064993E24 24D98055950E24 24D98055949E24 24D98055951E24 24D98055942E24 24D98055943E24 24D98055952E24 24D98064996E24 24D98067354E24 24D98067349E24 24D98064998E24 24D98065000E24 24D98055965E24 24D98055941E24 24D98055947E24 24D98055948E24 24D98064999E24 24D98067353E24 24D98064997E24 24D98064995E24 24D98060037E24 24D98055975E24 24D98055963E24 24D98050972E24 24D98038087G24 24D98038102G24 24D98071042G24 24D98038090G24 24D98038100G24 24D98038053G24 24D98038098G24 24D98038095G24 24D98038085G24 24D98038104G24 24D98055781E24 24D98055765E24 24D98055766E24 24D98055754E24 24D98055764E24 24D98055767E24 24D98055763E24 24D98055753E24 24D98055757E24 24D98055762E24 24D98055755E24 24D98055756E24 24D98055882E24 24D98055884E24 24D98055885E24 24D98055990E24 24D98055989E24 24D98055984E24 24D98055985E24 24D98055983E24 24D98060040E24 24D98060038E24 24D98055997E24 24D98055996E24 24D98055998E24 24D98060039E24 24D98055956E24 24D98055958E24 24D98055991E24 24D98050892E24 24D98050973E24 24D98055992E24 24D98040242G24 24D98055955E24 24D98060041E24 24D98055959E24 24D98055953E24 24D98055954E24 24D98055993E24 24D98055995E24 24D98055961E24 24D98055994E24 24D98055960E24 24D98055957E24 24D98055962E24 24D98055986E24 24D98055988E24 24D98055987E24 24D98055891E24 24D98055874E24 24D98055890E24 24D98055877E24 24D98055881E24 24D98055880E24 24D98055875E24 24D98055900E24 24D98055887E24 24D98055889E24 24D98055886E24 24D98050889E24 24D98055888E24 24D98050888E24 24D98055883E24 24D98050877E24 24D98055876E24 24D98067338E24 24D98055878E24 24D98055879E24 24D98071065G24 24D98051941G24 24D98051959G24 24D98051965G24 24D98051963G24 24D98051972G24 24D98051954G24 24D98051976G24 24D98051945G24 24D98051958G24 24D98051973G24 24D98051962G24 24D98051957G24 24D98051964G24 24D98051944G24 24D98051943G24 24D98051956G24 24D98051953G24 24D98051975G24 24D98051960G24 24D98051961G24 24D98051952G24 24D98051955G24 24D98051942G24 24D98051974G24 24D98071039G24 24D98071060G24 24D98061798G24 24D98061800G24 24D98071063G24 24D98071062G24 24D98071052G24 24D98071054G24 24D98061799G24 24D98071050G24 24D98071055G24 24D98071058G24 24D98061801G24 24D98040236G24 24D98040241G24 24D98040235G24 24D98040238G24 24D98040240G24 24D98040239G24 24D98040234G24 24D98040265G24 24D98040226G24 24D98040212G24 24D98040213G24 24D98040228G24 24D98040224G24 24D98040264G24 24D98040225G24 24D98040209G24 24D98040254G24 24D98040249G24 24D98040255G24 24D98040253G24 24D98040251G24 24D98040227G24 24D98040218G24 24D98040215G24 24D98040231G24 24D98040252G24 24D98040223G24 24D98040216G24 24D98040233G24 24D98040217G24 24D98040882G24 24D98040222G24 24D98040219G24 24D98040232G24 24D98040250G24 24D98040214G24 24D98040230G24 24D98040220G24 24D98040221G24 24D98002445H24 24D98040813H24 24D98040809H24 24D98040816H24 24D98040879H24 24D98040878H24 24D98040810H24 24D98093349H24 24D98093145G24 24D98093165G24 24D98093367H24 24D98040266G24 24D98093175G24 24D98093176G24 24D98093178G24 24D98093203H24 24D98093169G24 24D98093206H24 24D98093157G24 24D98093194H24 24D98040267G24 24D98093167G24 24D98093369H24 24D98093204H24 24D98093333H24 24D98093155G24 24D98093357H24 24D98093347H24 24D98093168G24 24D98093351H24 24D98093166G24 24D98093348H24 24D98093170G24 24D98093171G24 24D98093164G24 24D98093336H24 24D98093208H24 24D98093159G24 24D98093163G24 24D98093173G24 24D98093172G24 24D98093334H24 24D98093177G24 24D98093350H24 24D98093331H24 24D98093366H24 24D98093332H24 24D98093205H24 24D98093162G24 24D98093160G24 24D98093188G24 24D98093200H24 24D98093192H24 24D98040821H24 24D98093143G24 24D98093198H24 24D98093193H24 24D98093343H24 24D98093196H24 24D98093210H24 24D98093207H24 24D98093202H24 24D98093346H24 24D98093197H24 24D98093158G24 24D98093146G24 24D98093144G24 24D98093199H24 24D98093345H24 24D98093201H24 24D98093190H24 24D98093195H24 24D98093191H24 24D98093187G24 24D98093209H24 24D98093161G24 24D98093174G24 24D98040823H24 24D98093156G24 24D98093189G24 24D98039964B24 24D98026855F24 24D98039958B24 24D98052243G24 24D98020465H24 24D98093992H24 24D98026895F24 24D98026879F24 24D98029982A24 24D98048955G24 24D98026897F24 24D98068161E24 24D98053692E24 24D98053766E24 24D98050976E24 Product - 350P US Device, US English (American Airlines) - 2 Min CPR Item code: 350-STR-AA-AV UDI-DI code: 0506016712207 Serial Numbers: 24D98067253D24 24D98067252D24 24D98067251D24 24D98067249D24 24D98056938D24 24D98067250D24 24D98067248D24 24D98067153D24 24D98067160D24 24D98067155D24 24D98067163D24 24D98067180D24 24D98067182D24 24D98067219D24 24D98067186D24 24D98067220D24 24D98067221D24 24D98067215D24 24D98067156D24 24D98067200D24 24D98067195D24 24D98067242D24 24D98067231D24 24D98067234D24 24D98067244D24 24D98067197D24 24D98067232D24 24D98067194D24 24D98067166D24 24D98067165D24 24D98067161D24 24D98067169D24 24D98067241D24 24D98067162D24 24D98067192D24 24D98067159D24 24D98067198D24 24D98067157D24 24D98067203D24 24D98067191D24 24D98067158D24 24D98067246D24 24D98067225D24 24D98067247D24 24D98067206D24 24D98067173D24 24D98067204D24 24D98067213D24 24D98067209D24 24D98067187D24 24D98067207D24 24D98067178D24 24D98067217D24 24D98067214D24 24D98067222D24 24D98067189D24 24D98067212D24 24D98067171D24 24D98067170D24 24D98067177D24 24D98067223D24 24D98067237D24 24D98067172D24 24D98067245D24 24D98067210D24 24D98067211D24 24D98067205D24 24D98067233D24 24D98067208D24 24D98067174D24 24D98067226D24 24D98067190D24 24D98067164D24 24D98067168D24 24D98067167D24 24D98067175D24 24D98067176D24 24D98067181D24 24D98067185D24 24D98067184D24 24D98067216D24 24D98067188D24 24D98067183D24 24D98067218D24 24D98067179D24 24D98067228D24 24D98067229D24 24D98067202D24 24D98067236D24 24D98067235D24 24D98067201D24 24D98067239D24 24D98067230D24 24D98067243D24 24D98067196D24 24D98067193D24 24D98067240D24 24D98067227D24 24D98067224D24 24D98067144D24 24D98067238D24 24D98067121D24 24D98067131D24 Product - 350P US Device, US English with Gateway - 2 Min CPR Item code: 350-STR-US-GW UDI-DI code: 0506016712804 Serial Numbers: 24D98073174E24 24D98033695E24 24D98033694E24 24D98033696E24 24D98033711E24 24D98033688E24 24D98033686E24 Product - 360P US Device, US English - 2 Min CPR Item code: 360-STR-US-10 UDI-DI code: 0506016712068 Serial Numbers: 24E90017943A24 24E90016508A24 24E90017446A24 24E90017448A24 24E90017447A24 24E90016896A24 24E90017445A24 24E90017444A24 24E90017664A24 24E90017257A24 24E90017482A24 24E90017492A24 24E90017662A24 24E90017479A24 24E90017488A24 24E90017661A24 24E90017480A24 24E90017307A24 24E90017659A24 24E90017490A24 24E90017481A24 24E90017486A24 24E90017665A24 24E90017663A24 24E90017258A24 24E90017308A24 24E90017484A24 24E90017489A24 24E90017656A24 24E90017309A24 24E90017224A24 24E90017226A24 24E90016745A24 24E90016512A24 24E90017143A24 24E90017134A24 24E90017136A24 24E90018007A24 24E90017956A24 24E90017954A24 24E90017955A24 24E90017950A24 24E90017940A24 24E90017948A24 24E90017947A24 24E90017937A24 24E90017941A24 24E90017938A24 24E90017942A24 24E90017939A24 24E90017946A24 24E90017995A24 24E90017990A24 24E90018001A24 24E90017993A24 24E90017978A24 24E90018341A24 24E90017985A24 24E90018327A24 24E90018345A24 24E90018340A24 24E90018342A24 24E90017987A24 24E90017980A24 24E90017986A24 24E90017992A24 24E90017981A24 24E90017977A24 24E90017976A24 24E90018339A24 24E90018326A24 24E90018000A24 24E90018328A24 24E90017314A24 24E90017991A24 24E90017996A24 24E90017994A24 24E90017984A24 24E90017312A24 24E90017322A24 24E90018344A24 24E90018343A24 24E90017989A24 24E90017319A24 24E90017982A24 24E90017998A24 24E90017999A24 24E90017979A24 24E90017273A24 24E90017264A24 24E90017271A24 24E90017263A24 24E90017262A24 24E90017267A24 24E90017274A24 24E90017259A24 24E90017272A24 24E90017266A24 24E90017265A24 24E90017634A24 24E90003313B24 24E90016891A24 24E90017310A24 24E90018357A24 24E90003897B24 24E90003900B24 24E90003929B24 24E90003958B24 24E90003931B24 24E90003936B24 24E90003917B24 24E90003921B24 24E90003068B24 24E90003067B24 24E90003493B24 24E90003494B24 24E90003306B24 24E90003312B24 24E90003309B24 24E90013330B24 24E90004172B24 24E90005978B24 24E90018657B24 24E90001548B24 24E90014938B24 24E90015411B24 24E90002912B24 24E90001072B24 24E90015420B24 24E90001638B24 24E90000739B24 24E90000958B24 24E90006176B24 24E90006182B24 24E90006179B24 24E90006181B24 24E90004468B24 24E90004449B24 24E90004476B24 24E90004622B24 24E90004471B24 24E90004457B24 24E90004398B24 24E90004629B24 24E90004389B24 24E90004455B24 24E90004450B24 24E90004386B24 24E90004470B24 24E90004474B24 24E90004391B24 24E90004396B24 24E90004448B24 24E90004390B24 24E90004388B24 24E90004621B24 24E90004453B24 24E90004451B24 24E90004397B24 24E90004392B24 24E90004478B24 24E90004473B24 24E90004393B24 24E90004623B24 24E90004452B24 24E90004624B24 24E90004387B24 24E90004477B24 24E90004456B24 24E90004626B24 24E90004472B24 24E90004454B24 24E90004625B24 24E90004469B24 24E90005977B24 24E90004724B24 24E90004755B24 24E90004752B24 24E90004730B24 24E90004749B24 24E90004726B24 24E90004721B24 24E90004750B24 24E90004751B24 24E90004728B24 24E90004722B24 24E90004723B24 24E90004727B24 24E90006184B24 24E90014632B24 24E90002115B24 24E90004710B24 24E90004695B24 24E90004707B24 24E90004716B24 24E90004642B24 24E90004711B24 24E90004640B24 24E90004700B24 24E90004717B24 24E90004733B24 24E90004704B24 24E90004697B24 24E90004699B24 24E90004731B24 24E90004641B24 24E90004698B24 24E90004634B24 24E90004693B24 24E90004696B24 24E90004719B24 24E90004706B24 24E90004631B24 24E90004638B24 24E90004705B24 24E90004636B24 24E90004635B24 24E90004637B24 24E90004735B24 24E90004702B24 24E90004734B24 24E90004639B24 24E90004715B24 24E90004718B24 24E90004694B24 24E90004633B24 24E90004732B24 24E90004701B24 24E90004708B24 24E90004632B24 24E90002370B24 24E90004612B24 24E90004613B24 24E90004616B24 24E90004618B24 24E90004615B24 24E90004617B24 24E90004619B24 24E90006185B24 24E90006177B24 24E90006183B24 24E90006180B24 24E90006178B24 24E90004464B24 24E90004466B24 24E90004462B24 24E90004458B24 24E90005791B24 24E90004399B24 24E90004609B24 24E90003808B24 24E90004607B24 24E90004467B24 24E90004400B24 24E90004461B24 24E90004460B24 24E90003835B24 24E90005794B24 24E90004459B24 24E90004531B24 24E90005784B24 24E90005792B24 24E90003807B24 24E90004605B24 24E90004530B24 24E90004599B24 24E90004532B24 24E90005783B24 24E90004463B24 24E90004602B24 24E90004598B24 24E90003809B24 24E90005788B24 24E90005793B24 24E90004600B24 24E90004601B24 24E90004594B24 24E90004593B24 24E90005790B24 24E90003818B24 24E90005781B24 24E90005785B24 24E90004465B24 24E90004610B24 24E90006156B24 24E90005890B24 24E90005906B24 24E90005897B24 24E90006153B24 24E90005892B24 24E90005896B24 24E90005898B24 24E90005902B24 24E90005913B24 24E90006154B24 24E90005894B24 24E90006155B24 24E90005899B24 24E90005910B24 24E90005915B24 24E90005901B24 24E90005893B24 24E90005900B24 24E90005891B24 24E90005904B24 24E90005905B24 24E90005895B24 24E90005903B24 24E90005907B24 24E90005795B24 24E90005797B24 24E90006150B24 24E90005883B24 24E90005880B24 24E90006145B24 24E90006151B24 24E90005881B24 24E90006143B24 24E90005889B24 24E90006198B24 24E90006197B24 24E90006193B24 24E90006200B24 24E90006148B24 24E90006152B24 24E90005885B24 24E90006149B24 24E90006188B24 24E90005886B24 24E90005888B24 24E90005882B24 24E90005884B24 24E90006146B24 24E90006190B24 24E90006144B24 24E90014657B24 24E90006158B24 24E90006168B24 24E90006157B24 24E90006173B24 24E90006174B24 24E90006160B24 24E90006159B24 24E90006175B24 24E90006169B24 24E90006165B24 24E90006186B24 24E90006172B24 24E90006162B24 24E90006161B24 24E90006164B24 24E90006192B24 24E90006167B24 24E90006163B24 24E90015664B24 23E90139374B24 24E90012588A24 24E90005976B24 24E90017923B24 24E90015530B24 24E90001785B24 24E90015665B24 24E90004170B24 24E90001501B24 24E90001288B24 24E90005937B24 24E90005925B24 24E90015201A24 24E90017790A24 24E90017788A24 24E90017789A24 24E90017787A24 24E90017795A24 24E90018540A24 24E90017791A24 24E90018538A24 24E90018542A24 24E90017793A24 24E90018536A24 24E90018543A24 24E90017867A24 24E90018535A24 24E90017798A24 24E90017792A24 24E90017794A24 24E90018541A24 24E90018534A24 24E90018533A24 24E90018530A24 24E90015382A24 24E90018529A24 24E90017888A24 24E90015383A24 24E90017785A24 24E90017889A24 24E90000689A24 24E90017892A24 24E90017782A24 24E90018532A24 24E90018527A24 24E90018526A24 24E90017361A24 24E90017362A24 24E90017360A24 24E90017352A24 24E90017366A24 24E90017359A24 24E90017363A24 24E90017354A24 24E90017353A24 24E90017367A24 24E90017365A24 24E90017343A24 24E90017364A24 24E90017355A24 24E90017234A24 24E90017085A24 24E90017123A24 24E90017127A24 24E90016474A24 24E90017461A24 24E90017374A24 24E90017080A24 24E90017478A24 24E90017121A24 24E90016475A24 24E90017078A24 24E90017462A24 24E90017456A24 24E90017476A24 24E90017473A24 24E90016698A24 24E90017378A24 24E90016903A24 24E90017372A24 24E90017472A24 24E90017467A24 24E90017231A24 24E90017126A24 24E90017457A24 24E90017235A24 24E90017079A24 24E90017368A24 24E90017081A24 24E90017124A24 24E90016457A24 24E90016467A24 24E90017465A24 24E90017460A24 24E90017120A24 24E90017466A24 24E90017455A24 24E90017458A24 24E90017371A24 24E90017230A24 24E90016894A24 24E90017237A24 24E90017370A24 24E90017377A24 24E90017083A24 24E90017236A24 24E90017125A24 24E90016460A24 24E90016893A24 24E90017376A24 24E90017087A24 24E90017477A24 24E90017122A24 24E90017471A24 24E90017128A24 24E90017463A24 24E90017375A24 24E90017119A24 24E90017475A24 24E90016476A24 24E90016471A24 24E90017459A24 24E90017238A24 24E90017369A24 24E90017373A24 24E90017233A24 24E90017232A24 24E90016477A24 24E90017452A24 24E90017464A24 24E90017082A24 24E90017086A24 24E90018004A24 24E90018336A24 24E90018286A24 24E90018287A24 24E90014848A24 24E90018334A24 24E90018338A24 24E90018302A24 24E90018277A24 24E90017299A24 24E90017507A24 24E90017301A24 24E90017494A24 24E90017505A24 24E90017324A24 24E90017512A24 24E90017325A24 24E90017252A24 24E90017524A24 24E90017255A24 24E90017501A24 24E90017328A24 24E90017517A24 24E90017323A24 24E90017485A24 24E90017250A24 24E90017518A24 24E90017563A24 24E90017326A24 24E90017503A24 24E90017297A24 24E90017508A24 24E90017493A24 24E90017253A24 24E90017294A24 24E90017541A24 24E90017510A24 24E90017515A24 24E90017498A24 24E90017513A24 24E90017509A24 24E90017521A24 24E90017526A24 24E90017511A24 24E90017254A24 24E90017500A24 24E90017251A24 24E90017499A24 24E90017293A24 24E90017495A24 24E90017523A24 24E90017300A24 24E90017292A24 24E90017519A24 24E90017516A24 24E90017504A24 24E90017295A24 24E90017522A24 24E90017256A24 24E90017525A24 24E90017520A24 24E90017298A24 24E90017514A24 24E90017502A24 24E90017497A24 24E90017296A24 24E90017506A24 24E90016454A24 24E90016439A24 24E90017110A24 24E90017222A24 24E90017116A24 24E90017225A24 24E90017109A24 24E90017227A24 24E90017115A24 24E90017117A24 24E90017228A24 24E90017229A24 24E90016443A24 24E90017220A24 24E90017113A24 24E90017221A24 24E90017118A24 24E90017223A24 24E90017114A24 24E90016747A24 24E90017111A24 24E90016989A24 24E90016990A24 24E90016991A24 24E90012680A24 24E90016978A24 24E90016981A24 24E90016979A24 24E90016988A24 24E90016980A24 24E90016837A24 24E90016840A24 24E90016842A24 24E90016805A24 24E90016810A24 24E90016721A24 24E90016571A24 24E90016703A24 24E90017075A24 24E90016570A24 24E90018424A24 24E90017896A24 24E90017893A24 24E90018429A24 24E90018427A24 24E90018442A24 24E90017815A24 24E90017408A24 24E90018425A24 24E90018431A24 24E90018428A24 24E90017894A24 24E90018443A24 24E90017405A24 24E90017812A24 24E90017811A24 24E90017131A24 24E90017130A24 24E90017621A24 24E90017138A24 24E90018300A24 24E90017141A24 24E90017132A24 24E90017142A24 24E90017624A24 24E90018281A24 24E90017242A24 24E90017133A24 24E90018284A24 24E90017241A24 24E90017248A24 24E90018282A24 24E90017619A24 24E90018285A24 24E90017618A24 24E90018337A24 24E90017244A24 24E90017239A24 24E90017623A24 24E90017140A24 24E90018283A24 24E90017243A24 24E90017625A24 24E90017139A24 24E90017622A24 24E90017620A24 24E90018279A24 24E90017246A24 24E90017245A24 24E90017617A24 24E90017129A24 24E90017247A24 24E90017240A24 24E90018426A24 24E90018487A24 24E90018488A24 24E90018523A24 24E90018420A24 24E90013066A24 24E90013060A24 24E90013067A24 24E90017402A24 24E90017406A24 24E90013058A24 24E90017403A24 24E90013064A24 24E90013063A24 24E90013062A24 24E90017404A24 24E90017707A24 24E90013059A24 24E90018565A24 24E90018559A24 24E90018560A24 24E90018571A24 24E90018013A24 24E90018547A24 24E90017096A24 24E90018568A24 24E90018566A24 24E90018554A24 24E90018567A24 24E90018005A24 24E90018561A24 24E90016862A24 24E90018569A24 24E90018564A24 24E90018006A24 24E90018570A24 24E90018563A24 24E90018555A24 24E90018552A24 24E90018572A24 24E90018574A24 24E90018573A24 24E90018576A24 24E90018580A24 24E90018575A24 24E90018577A24 24E90018581A24 24E90018587A24 24E90018584A24 24E90018583A24 24E90018551A24 24E90018413A24 24E90018591A24 24E90018590A24 24E90018546A24 24E90018335A24 24E90018588A24 24E90018589A24 24E90018298A24 24E90018410A24 24E90016585A24 24E90018585A24 24E90018548A24 24E90018586A24 24E90018433A24 24E90018441A24 24E90018432A24 24E90018440A24 24E90018435A24 24E90018434A24 24E90018436A24 24E90018439A24 24E90018322A24 24E90018017A24 24E90017318A24 24E90018307A24 24E90018317A24 24E90018309A24 24E90018305A24 24E90018306A24 24E90018008A24 24E90018274A24 24E90018009A24 24E90018280A24 24E90018010A24 24E90018295A24 24E90018296A24 24E90017288A24 24E90018331A24 24E90018315A24 24E90018022A24 24E90017283A24 24E90017289A24 24E90018288A24 24E90018012A24 24E90018316A24 24E90018320A24 24E90018351A24 24E90018224A24 24E90017282A24 24E90018016A24 24E90017286A24 24E90018329A24 24E90017302A24 24E90018311A24 24E90018550A24 24E90018323A24 24E90017304A24 24E90018304A24 24E90018310A24 24E90018332A24 24E90018330A24 24E90018321A24 24E90018318A24 24E90018289A24 24E90018319A24 24E90018314A24 24E90018325A24 24E90018308A24 24E90018018A24 24E90018011A24 24E90018312A24 24E90018290A24 24E90017305A24 24E90017281A24 24E90018294A24 24E90017285A24 24E90017303A24 24E90017290A24 24E90018292A24 24E90017291A24 24E90018275A24 24E90018333A24 24E90018324A24 24E90017279A24 24E90018023A24 24E90018293A24 24E90018291A24 24E90018014A24 24E90018015A24 24E90017280A24 24E90018303A24 24E90018297A24 24E90000677A24 24E90000682A24 24E90000664A24 24E90000690A24 24E90000692A24 24E90000683A24 24E90000678A24 24E90015394A24 24E90015011A24 24E90003106B24 24E90003103B24 24E90003129B24 24E90015386A24 24E90015391A24 24E90003105B24 24E90003127B24 24E90015389A24 24E90003565B24 24E90014964A24 24E90015236A24 24E90015229A24 24E90015231A24 24E90014963A24 24E90015219A24 24E90014962A24 24E90015215A24 24E90014961A24 24E90014622A24 24E90015233A24 24E90015217A24 24E90015235A24 24E90015225A24 24E90015232A24 24E90015222A24 24E90015230A24 24E90015221A24 24E90015226A24 24E90015224A24 24E90014965A24 24E90015480A24 24E90015287A24 24E90015288A24 24E90015507A24 24E90000393A24 24E90015289A24 24E90015290A24 24E90015284A24 24E90015285A24 24E90015286A24 24E90015282A24 24E90015283A24 24E90000394A24 24E90000391A24 24E90015468A24 24E90000593A24 24E90000395A24 24E90015281A24 24E90000392A24 24E90003123B24 24E90000302A24 24E90003125B24 24E90003122B24 24E90000301A24 24E90003126B24 24E90003131B24 24E90003130B24 24E90015373A24 24E90000598A24 24E90000621A24 24E90003661B24 24E90000451A24 24E90015555A24 24E90015492A24 24E90000616A24 24E90003102B24 24E90000614A24 24E90000390A24 24E90003111B24 24E90003658B24 24E90000458A24 24E90003660B24 24E90000557A24 24E90015493A24 24E90000457A24 24E90015496A24 24E90015471A24 24E90003110B24 24E90015498A24 24E90015474A24 24E90000606A24 24E90003663B24 24E90000456A24 24E90000459A24 24E90000605A24 24E90015497A24 24E90015475A24 24E90015500A24 24E90000615A24 24E90000601A24 24E90015491A24 24E90015556A24 24E90000622A24 24E90015501A24 24E90000453A24 24E90000620A24 24E90015535A24 24E90003132B24 24E90000452A24 24E90003657B24 24E90000602A24 24E90000617A24 24E90003659B24 24E90000603A24 24E90000604A24 24E90015490A24 24E90015536A24 24E90015499A24 24E90015532A24 24E90015502A24 24E90000600A24 24E90015296A24 24E90015310A24 24E90015291A24 24E90015302A24 24E90014780A24 24E90014765A24 24E90014768A24 24E90014791A24 24E90014784A24 24E90014783A24 24E90014767A24 24E90014769A24 24E90014764A24 24E90014770A24 24E90014785A24 24E90014789A24 24E90003119B24 24E90003116B24 24E90003117B24 24E90003120B24 24E90003112B24 24E90003115B24 24E90002151B24 24E90003121B24 24E90003118B24 24E90002150B24 24E90014800A24 24E90014829A24 24E90014804A24 24E90014797A24 24E90014816A24 24E90014803A24 24E90014777A24 24E90014793A24 24E90014814A24 24E90014801A24 24E90014794A24 24E90014802A24 24E90014792A24 24E90014796A24 24E90014831A24 24E90014833A24 24E90014836A24 24E90014832A24 24E90000627A24 24E90000629A24 24E90014835A24 24E90014837A24 24E90000628A24 24E90000632A24 24E90000633A24 24E90014820A24 24E90000634A24 24E90014819A24 24E90014834A24 24E90000630A24 24E90000631A24 24E90014818A24 24E90014823A24 24E90015561A24 24E90015571A24 24E90015559A24 24E90000277A24 24E90000513A24 24E90014773A24 24E90000279A24 24E90015545A24 24E90000282A24 24E90015562A24 24E90000283A24 24E90000378A24 24E90014771A24 24E90015546A24 24E90000280A24 24E90000381A24 24E90015560A24 24E90000368A24 24E90000284A24 24E90000281A24 24E90015558A24 24E90014772A24 24E90000296A24 24E90015573A24 24E90015563A24 24E90015570A24 24E90015572A24 24E90015544A24 24E90015557A24 24E90000508A24 24E90000509A24 24E90000377A24 24E90000374A24 24E90003092B24 24E90003098B24 24E90003096B24 24E90003099B24 24E90003100B24 24E90003093B24 24E90002328B24 24E90002149B24 24E90002333B24 24E90001962B24 24E90002143B24 24E90003077B24 24E90002332B24 24E90001963B24 24E90002330B24 24E90002324B24 24E90002329B24 24E90003555B24 24E90001964B24 24E90002148B24 24E90001965B24 24E90002142B24 24E90002326B24 24E90002141B24 24E90001961B24 24E90002331B24 24E90003078B24 24E90002327B24 24E90003681B24 24E90003684B24 24E90003665B24 24E90003682B24 24E90003680B24 24E90003672B24 24E90003671B24 24E90004131B24 24E90001997B24 24E90003769B24 24E90003791B24 24E90004108B24 24E90003792B24 24E90003762B24 24E90004127B24 24E90003763B24 24E90004130B24 24E90004105B24 24E90003785B24 24E90004110B24 24E90003996B24 24E90003786B24 24E90003771B24 24E90004107B24 24E90003788B24 24E90003794B24 24E90003787B24 24E90004091B24 24E90003999B24 24E90004160B24 24E90003998B24 24E90003765B24 24E90004093B24 24E90003774B24 24E90004109B24 24E90004120B24 24E90003994B24 24E90004129B24 24E90004092B24 24E90003773B24 24E90003761B24 24E90003764B24 24E90004106B24 24E90003760B24 24E90003766B24 24E90003768B24 24E90003789B24 24E90003759B24 24E90003995B24 24E90004104B24 24E90003770B24 24E90004089B24 24E90003758B24 24E90003772B24 24E90004159B24 24E90003767B24 24E90001974B24 24E90001975B24 24E90001968B24 24E90001969B24 24E90001967B24 24E90001970B24 24E90001977B24 24E90001972B24 24E90004098B24 24E90004101B24 24E90004097B24 24E90004102B24 24E90004099B24 24E90004094B24 24E90004095B24 24E90004100B24 24E90004017B24 24E90003090B24 24E90003745B24 24E90002337B24 24E90002336B24 24E90002338B24 24E90003740B24 24E90003085B24 24E90003091B24 24E90003084B24 24E90003086B24 24E90003088B24 24E90002341B24 24E90002335B24 24E90003087B24 24E90002334B24 24E90002340B24 24E90003083B24 24E90002339B24 24E90003739B24 24E90002169B24 24E90002167B24 24E90002164B24 24E90002166B24 24E90001987B24 24E90001978B24 24E90001985B24 24E90002168B24 24E90001986B24 24E90004168B24 24E90002170B24 24E90002173B24 24E90003988B24 24E90004164B24 24E90004169B24 24E90003989B24 24E90004165B24 24E90002172B24 24E90003987B24 24E90002315B24 24E90003747B24 24E90003775B24 24E90004003B24 24E90002346B24 24E90003781B24 24E90003778B24 24E90003776B24 24E90003784B24 24E90004006B24 24E90004007B24 24E90002307B24 24E90002305B24 24E90002342B24 24E90002309B24 24E90003779B24 24E90001992B24 24E90002311B24 24E90004009B24 24E90002310B24 24E90002301B24 24E90002321B24 24E90002323B24 24E90003777B24 24E90002312B24 24E90004005B24 24E90004000B24 24E90002348B24 24E90004002B24 24E90001988B24 24E90004010B24 24E90004015B24 24E90002304B24 24E90004001B24 24E90002322B24 24E90003780B24 24E90004012B24 24E90002303B24 24E90001989B24 24E90004013B24 24E90002345B24 24E90001996B24 24E90003782B24 24E90003783B24 24E90002344B24 24E90002316B24 24E90004016B24 24E90002302B24 24E90004011B24 24E90003757B24 24E90003738B24 24E90002343B24 24E90001993B24 24E90003743B24 24E90001995B24 24E90003744B24 24E90002306B24 24E90002320B24 24E90002318B24 24E90004008B24 24E90004004B24 24E90002308B24 24E90003742B24 24E90003223B24 24E90003217B24 24E90003216B24 24E90003226B24 24E90003215B24 24E90003317B24 24E90003231B24 24E90003232B24 24E90003230B24 24E90003314B24 24E90003229B24 24E90003227B24 24E90003506B24 24E90003225B24 24E90003228B24 24E90015902A24 24E90012610A24 23E90144949A24 24E90013313A24 24E90013327A24 24E90013309A24 24E90017874A24 24E90017870A24 24E90017884A24 24E90017885A24 24E90017887A24 24E90017883A24 24E90017879A24 24E90017886A24 24E90017872A24 24E90017871A24 24E90017869A24 24E90018481A24 24E90018482A24 24E90018483A24 24E90018484A24 24E90017066A24 24E90017057A24 24E90016241A24 24E90017112A24 24E90017211A24 24E90017160A24 24E90015986A24 24E90017320A24 24E90017533A24 24E90017334A24 24E90018091A24 24E90017277A24 24E90017534A24 24E90017333A24 24E90017329A24 24E90017335A24 24E90017331A24 24E90017330A24 24E90017332A24 24E90017261A24 24E90017275A24 24E90017268A24 24E90017260A24 24E90017269A24 24E90017199A24 24E90017198A24 24E90017097A24 24E90017091A24 24E90017206A24 24E90017196A24 24E90017092A24 24E90017207A24 24E90017181A24 24E90017214A24 24E90017182A24 24E90017183A24 24E90017215A24 24E90017210A24 24E90017099A24 24E90017213A24 24E90017098A24 24E90017216A24 24E90017218A24 24E90017217A24 24E90017219A24 24E90017106A24 24E90017100A24 24E90017186A24 24E90017187A24 24E90017101A24 24E90017105A24 24E90017629A24 24E90017535A24 24E90017635A24 24E90017537A24 24E90017536A24 24E90017532A24 24E90017529A24 24E90017528A24 24E90017628A24 24E90017627A24 24E90017633A24 24E90017179A24 24E90017539A24 24E90017626A24 24E90017527A24 24E90017632A24 24E90017205A24 24E90017090A24 24E90017192A24 24E90017191A24 24E90017209A24 24E90017095A24 24E90017193A24 24E90017208A24 24E90017094A24 24E90017197A24 24E90017201A24 24E90017204A24 24E90017631A24 24E90017538A24 24E90017630A24 24E90017089A24 24E90017195A24 24E90017088A24 24E90017194A24 24E90017202A24 24E90017203A24 24E90018658A24 24E90017107A24 24E90017212A24 24E90017184A24 24E90017185A24 24E90017180A24 24E90017188A24 24E90017103A24 24E90017189A24 24E90017200A24 24E90015279A24 24E90015275A24 24E90015277A24 24E90001278A24 24E90001281A24 24E90001283A24 24E90001284A24 24E90001256A24 24E90000644A24 24E90000645A24 24E90015280A24 24E90001280A24 24E90000648A24 24E90000646A24 24E90001253A24 24E90000639A24 24E90000640A24 24E90000636A24 24E90000635A24 24E90001287A24 24E90000642A24 24E90001285A24 24E90000643A24 24E90001286A24 24E90001381A24 24E90000641A24 24E90001251A24 24E90000638A24 24E90000637A24 24E90001254A24 24E90001289A24 24E90001269A24 24E90001267A24 23E90116066K23 24E90012533A24 24E90012531A24 24E90012532A24 24E90015379A24 24E90015378A24 24E90015385A24 24E90015384A24 24E90015376A24 24E90015380A24 24E90015381A24 24E90012523A24 24E90012525A24 24E90012526A24 24E90012528A24 24E90012527A24 24E90016374A24 24E90016772A24 24E90016367A24 24E90016372A24 24E90016386A24 24E90016767A24 24E90016769A24 24E90016373A24 24E90016385A24 24E90016370A24 24E90016768A24 24E90016770A24 24E90016368A24 24E90016773A24 24E90016766A24 24E90016366A24 24E90016371A24 24E90016384A24 24E90016383A24 24E90016379A24 24E90016380A24 24E90016387A24 24E90016381A24 24E90016376A24 24E90016382A24 24E90012673A24 24E90012669A24 24E90012671A24 24E90012670A24 24E90012674A24 24E90012668A24 24E90012672A24 24E90012675A24 24E90016362A24 24E90016364A24 24E90012682A24 24E90012683A24 24E90016774A24 24E90012684A24 24E90012679A24 24E90016365A24 24E90012681A24 24E90012678A24 24E90016351A24 24E90012677A24 24E90016348A24 24E90016349A24 24E90012676A24 24E90016347A24 24E90016350A24 24E90016342A24 24E90015371A24 24E90015427A24 24E90015431A24 24E90015400A24 24E90015404A24 24E90015396A24 24E90015365A24 24E90015398A24 24E90001064A24 24E90015399A24 24E90015402A24 24E90015403A24 24E90015397A24 24E90015395A24 24E90000547A24 24E90015318A24 24E90015388A24 24E90000545A24 24E90000544A24 24E90000543A24 24E90000541A24 24E90015435A24 24E90015433A24 24E90015387A24 24E90015428A24 24E90015392A24 24E90015393A24 24E90015390A24 24E90015430A24 24E90015429A24 24E90015434A24 24E90015426A24 24E90000574A24 24E90000572A24 24E90000576A24 24E90000571A24 24E90000568A24 24E90000569A24 24E90000573A24 24E90015211A24 24E90015407A24 24E90000575A24 24E90015406A24 24E90000570A24 24E90015409A24 24E90015413A24 24E90015408A24 24E90015415A24 24E90015375A24 24E90015369A24 24E90015367A24 24E90015370A24 24E90015372A24 24E90015368A24 24E90015366A24 24E90000556A24 24E90015551A24 24E90015548A24 24E90015537A24 24E90000578A24 24E90015549A24 24E90015533A24 24E90000554A24 24E90000563A24 24E90000580A24 24E90000553A24 24E90015554A24 24E90000555A24 24E90015547A24 24E90000564A24 24E90015550A24 24E90015531A24 24E90015542A24 24E90000577A24 24E90000581A24 24E90000579A24 24E90000551A24 24E90015541A24 24E90015553A24 24E90015552A24 24E90015538A24 24E90000552A24 24E90015543A24 24E90015540A24 24E90000539A24 24E90015423A24 24E90015419A24 24E90015417A24 24E90000538A24 24E90015416A24 24E90015418A24 24E90000533A24 24E90015451A24 24E90015463A24 24E90015459A24 24E90015465A24 24E90015338A24 24E90015340A24 24E90015334A24 24E90015339A24 24E90015337A24 24E90015342A24 24E90015341A24 24E90003902B24 24E90003263B24 24E90003260B24 24E90015438A24 24E90015439A24 24E90015437A24 24E90015444A24 24E90015440A24 24E90015445A24 24E90015436A24 24E90015441A24 24E90000285A24 24E90003072B24 24E90003954B24 24E90003914B24 24E90003950B24 24E90003947B24 24E90003556B24 24E90003915B24 24E90003935B24 24E90003080B24 24E90003927B24 24E90003930B24 24E90003082B24 24E90003945B24 24E90003905B24 24E90003946B24 24E90003944B24 24E90003942B24 24E90003896B24 24E90003894B24 23E90117536A24 23E90117549A24 23E90117546A24 24E90015886A24 24E90015957A24 24E90015896A24 24E90015838A24 24E90015846A24 24E90015877A24 24E90015880A24 24E90015878A24 24E90015980A24 24E90015884A24 24E90015962A24 24E90015897A24 24E90015898A24 24E90015981A24 24E90015988A24 24E90015841A24 24E90015968A24 24E90015875A24 24E90015983A24 24E90015879A24 24E90015979A24 24E90015978A24 24E90015900A24 24E90015876A24 24E90015994A24 24E90015958A24 24E90015895A24 24E90015984A24 24E90015889A24 24E90015989A24 24E90015967A24 24E90015871A24 24E90015888A24 24E90015882A24 24E90015899A24 24E90015891A24 24E90015892A24 24E90015956A24 24E90015987A24 24E90015985A24 24E90015951A24 24E90015955A24 24E90015982A24 24E90015901A24 24E90015974A24 24E90015972A24 24E90015973A24 24E90015971A24 24E90015977A24 24E90015970A24 24E90015975A24 24E90016360A24 24E90016359A24 24E90016355A24 24E90016358A24 24E90016361A24 24E90016353A24 24E90016352A24 24E90016354A24 24E90017997A24 23E90130321K23 23E90139221K23 24E90003282B24 24E90003288B24 24E90003285B24 24E90003235B24 24E90003299B24 24E90003272B24 24E90003300B24 24E90003290B24 24E90003271B24 24E90003233B24 24E90003294B24 24E90003266B24 24E90003277B24 24E90012592A24 24E90012589A24 24E90012586A24 24E90012561A24 24E90012608A24 24E90012556A24 24E90012606A24 24E90012613A24 24E90012555A24 24E90012577A24 24E90015923A24 24E90003254B24 24E90003256B24 24E90003888B24 24E90003984B24 24E90003964B24 24E90003887B24 24E90003940B24 24E90003941B24 24E90003283B24 24E90003279B24 24E90003203B24 24E90003296B24 24E90003284B24 24E90003278B24 24E90012591A24 24E90012594A24 24E90012590A24 24E90015921A24 24E90012599A24 24E90017658A24 24E90017949A24 24E90017988A24 24E90003073B24 24E90003071B24 24E90018226A24 24E90018225A24 24E90018389A24 24E90018393A24 24E90018391A24 24E90018227A24 24E90017713A24 24E90016944A24 24E90016952A24 24E90016945A24 24E90016951A24 24E90016942A24 24E90016955A24 24E90016948A24 24E90016953A24 24E90017442A24 24E90017725A24 24E90017474A24 24E90017449A24 24E90017439A24 24E90017454A24 24E90016895A24 24E90017440A24 24E90017443A24 24E90016900A24 24E90016897A24 24E90016901A24 24E90016902A24 24E90016904A24 24E90017441A24 24E90017714A24 24E90018169A24 24E90018374A24 24E90018378A24 24E90018167A24 24E90018170A24 24E90018168A24 24E90016700A24 24E90018384A24 24E90018387A24 24E90018383A24 24E90018631A24 24E90018633A24 24E90018544A24 24E90018645A24 24E90018634A24 24E90018623A24 24E90018189A24 24E90018625A24 24E90018134A24 24E90018642A24 24E90018171A24 24E90018172A24 24E90018632A24 24E90018644A24 24E90018130A24 24E90018185A24 24E90018164A24 24E90018186A24 24E90018627A24 24E90018622A24 24E90018630A24 24E90018626A24 24E90018133A24 24E90018156A24 24E90018647A24 24E90018135A24 24E90018646A24 24E90018144A24 24E90018163A24 24E90018639A24 24E90018636A24 24E90018188A24 24E90018641A24 24E90018624A24 24E90018643A24 24E90018628A24 24E90018161A24 24E90018637A24 24E90018162A24 24E90018132A24 24E90018128A24 24E90018640A24 24E90018165A24 24E90018159A24 24E90018194A24 24E90018187A24 24E90018160A24 24E90018129A24 24E90018201A24 24E90018126A24 24E90018629A24 24E90018545A24 24E90018638A24 24E90018192A24 24E90018635A24 24E90017410A24 24E90017418A24 24E90017450A24 24E90017428A24 24E90017421A24 24E90017451A24 24E90017433A24 24E90017760A24 24E90017409A24 24E90017432A24 24E90017424A24 24E90017430A24 24E90017427A24 24E90017429A24 24E90017417A24 24E90017423A24 24E90017431A24 24E90017426A24 24E90017422A24 24E90018556A24 24E90018553A24 24E90018370A24 24E90018562A24 24E90018155A24 24E90018558A24 24E90018557A24 24E90018054A24 24E90018208A24 24E90018049A24 24E90018218A24 24E90018594A24 24E90018609A24 24E90018604A24 24E90017926A24 24E90018607A24 24E90018617A24 24E90017930A24 24E90018062A24 24E90018615A24 24E90018610A24 24E90017925A24 24E90018620A24 24E90018614A24 24E90017927A24 24E90017761A24 24E90018595A24 24E90017931A24 24E90017936A24 24E90017933A24 24E90018600A24 24E90018621A24 24E90017935A24 24E90017762A24 24E90018597A24 24E90018613A24 24E90018602A24 24E90017934A24 24E90018605A24 24E90018606A24 24E90018599A24 24E90018616A24 24E90018619A24 24E90017928A24 24E90018603A24 24E90018596A24 24E90018223A24 24E90018608A24 24E90018618A24 24E90017932A24 24E90018598A24 24E90018612A24 24E90018408A24 24E90018611A24 24E90017764A24 24E90017766A24 24E90018047A24 24E90018352A24 24E90018355A24 24E90018045A24 24E90018041A24 24E90018350A24 24E90018367A24 24E90018354A24 24E90018347A24 24E90018348A24 24E90018358A24 24E90018048A24 24E90018356A24 24E90018042A24 24E90018346A24 24E90018046A24 24E90018368A24 24E90018360A24 24E90018043A24 24E90018039A24 24E90018349A24 24E90018044A24 24E90018549A24 24E90018040A24 24E90018359A24 24E90015823A24 24E90015940A24 24E90015933A24 24E90015936A24 24E90015934A24 24E90015935A24 24E90015937A24 24E90018250A24 24E90018178A24 24E90018183A24 24E90018230A24 24E90018030A24 24E90018031A24 24E90018237A24 24E90018236A24 24E90018033A24 24E90018229A24 24E90018228A24 24E90018037A24 24E90018038A24 24E90018028A24 24E90018034A24 24E90018248A24 24E90018179A24 24E90018184A24 24E90018035A24 24E90018036A24 24E90018251A24 24E90018029A24 24E90018032A24 24E90018233A24 24E90018232A24 24E90017747A24 24E90017739A24 24E90017735A24 24E90017716A24 24E90017715A24 24E90018243A24 24E90018245A24 24E90018244A24 24E90018239A24 24E90018238A24 24E90018241A24 24E90017742A24 24E90017743A24 24E90017741A24 24E90017740A24 24E90017738A24 24E90018242A24 24E90018247A24 24E90018246A24 24E90018240A24 24E90017736A24 24E90017737A24 24E90003251B24 24E90003258B24 24E90003261B24 24E90003259B24 24E90003257B24 24E90003262B24 24E90003255B24 24E90003912B24 24E90003948B24 24E90003076B24 24E90003951B24 24E90003075B24 24E90003074B24 24E90003952B24 24E90003949B24 24E90003923B24 24E90003081B24 24E90003922B24 24E90003069B24 24E90003064B24 24E90003544B24 24E90004153B24 24E90004163B24 24E90004144B24 24E90004090B24 24E90003953B24 24E90003253B24 24E90004166B24 24E90003957B24 24E90003889B24 24E90003890B24 24E90003943B24 24E90003267B24 24E90003286B24 24E90003298B24 24E90003270B24 24E90003291B24 24E90003234B24 24E90003293B24 24E90003194B24 24E90003195B24 24E90003268B24 24E90003269B24 24E90003295B24 24E90003273B24 24E90003264B24 24E90003265B24 24E90003289B24 24E90003275B24 24E90003276B24 24E90003274B24 24E90017306A24 24E90017487A24 24E90017315A24 24E90003911B24 24E90016892A24 24E90017491A24 24E90018366A24 24E90017983A24 24E90017321A24 24E90017313A24 24E90003919B24 24E90008782D24 24E90008892D24 24E90008785D24 24E90008904D24 24E90008873D24 24E90008783D24 24E90008786D24 24E90008885D24 24E90008878D24 24E90008874D24 24E90008784D24 24E90008893D24 24E90008891D24 24E90008884D24 24E90008890D24 24E90008882D24 24E90008905D24 24E90008902D24 24E90008887D24 24E90008875D24 24E90008777D24 24E90008876D24 24E90011868D24 24E90011867D24 24E90011880D24 24E90011863D24 24E90105958H24 24E90105950H24 24E90105716H24 24E90105726H24 24E90105725H24 24E90105728H24 24E90105639H24 24E90105722H24 24E90105723H24 24E90105724H24 24E90008973D24 24E90008965D24 24E90008970D24 24E90008974D24 24E90008972D24 24E90008968D24 24E90008971D24 24E90008958D24 24E90008964D24 24E90008982D24 24E90008961D24 24E90008787D24 24E90009007D24 24E90008975D24 24E90008980D24 24E90008984D24 24E90008960D24 24E90008962D24 24E90008963D24 24E90008959D24 24E90008907D24 24E90008914D24 24E90009009D24 24E90008981D24 24E90008979D24 24E90009010D24 24E90008977D24 24E90008986D24 24E90008978D24 24E90024769D24 24E90024761D24 24E90024357D24 24E90024758D24 24E90023975D24 24E90018737D24 24E90018741D24 24E90018742D24 24E90018747D24 24E90024356D24 24E90024358D24 24E90024363D24 24E90024361D24 24E90021240D24 24E90024360D24 24E90021242D24 24E90021239D24 24E90024757D24 24E90023951D24 24E90021349D24 24E90021241D24 24E90023415C24 24E90023420C24 24E90023417C24 24E90023411C24 24E90006363D24 24E90020999D24 24E90020993D24 24E90020997D24 24E90024767D24 24E90024762D24 24E90024760D24 24E90018739D24 24E90024759D24 24E90024766D24 24E90023986D24 24E90023990D24 24E90023988D24 24E90023995D24 24E90023987D24 24E90023994D24 24E90019739D24 24E90018744D24 24E90018738D24 24E90003928B24 24E90018353A24 24E90018363A24 24E90017316A24 24E90003310B24 24E90018361A24 24E90003983B24 24E90003492B24 24E90003909B24 24E90017945A24 24E90003893B24 24E90018362A24 24E90017657A24 24E90017483A24 24E90018364A24 24E90018369A24 24E90017311A24 Product - SAM 450P US Device, US English (2 min CPR) Item code: 450-STR-US-10 UDI-DI code: 0506016712069 Serial Numbers: 24G90001624D24 24G90001620D24 24G90001610D24 23G90004172G23 23G90108237L23 23G90108235L23 23G90108212L23 23G90108209L23 23G90108236L23 23G90108213L23 23G90108208L23 23G90108215L23 23G90108210L23 23G90108211L23 23G90108495L23 23G90108226L23 23G90108322L23 23G90108326L23 23G90108222L23 23G90108499L23 23G90108225L23 23G90108497L23 23G90108498L23 23G90108223L23 23G90108492L23 23G90108501L23 23G90108240L23 23G90108227L23 23G90108242L23 23G90108239L23 23G90108684L23 23G90108221L23 23G90108247L23 23G90108479L23 23G90108241L23 23G90108244L23 23G90108246L23 23G90108218L23 23G90108313L23 23G90108219L23 23G90108477L23 23G90108238L23 23G90108473L23 23G90108341L23 23G90108368L23 23G90108345L23 23G90108502L23 23G90108339L23 23G90108365L23 23G90108363L23 23G90108347L23 23G90108344L23 23G90108342L23 23G90108527L23 23G90108524L23 23G90108528L23 23G90108525L23 23G90108513L23 23G90108506L23 23G90108537L23 23G90108531L23 23G90108536L23 23G90108532L23 23G90108538L23 23G90108526L23 23G90108507L23 23G90108523L23 23G90108517L23 23G90108515L23 23G90108533L23 23G90108205L23 23G90108402L23 23G90108203L23 23G90108398L23 23G90108408L23 23G90108199L23 23G90108417L23 23G90108415L23 23G90108409L23 23G90108441L23 23G90108692L23 23G90108437L23 23G90108207L23 23G90108411L23 23G90108399L23 23G90108412L23 23G90108401L23 23G90108403L23 23G90108416L23 23G90108699L23 23G90108440L23 23G90108200L23 23G90108198L23 23G90108404L23 23G90108201L23 23G90108204L23 23G90108655L23 23G90108693L23 23G90108202L23 23G90108428L23 23G90108419L23 23G90108442L23 23G90108410L23 23G90108418L23 23G90108405L23 23G90108706L23 23G90108707L23 23G90108505L23 23G90108709L23 23G90108704L23 23G90108708L23 23G90108705L23 23G90008119L23 23G90108259L23 23G90108256L23 23G90108457L23 23G90107062J23 23G90108274L23 23G90108271L23 23G90009047L23 23G90107662J23 24G90013199H24 24G90013402H24 24G90013387H24 24G90013395H24 24G90013388H24 24G90010488H24 24G90013392H24 24G90013397H24 24G90013386H24 24G90013390H24 24G90013403H24 24G90013391H24 24G90013401H24 24G90013389H24 24G90013195H24 24G90013396H24 24G90013399H24 24G90013404H24 24G90013400H24 24G90013652H24 24G90013201H24 24G90013635H24 24G90013622H24 24G90013617H24 24G90013650H24 24G90013628H24 24G90013620H24 24G90013192H24 24G90013193H24 24G90013627H24 24G90013630H24 24G90013224H24 24G90013223H24 24G90013632H24 24G90013625H24 24G90013636H24 24G90013221H24 24G90013197H24 24G90013629H24 24G90013631H24 24G90013637H24 24G90013633H24 24G90013634H24 24G90013623H24 24G90013626H24 24G90013385H24 24G90013383H24 24G90013382H24 24G90013381H24 24G90013378H24 24G90013384H24 24G90013612H24 24G90013661H24 24G90013648H24 24G90013655H24 24G90013656H24 24G90013665H24 24G90007696D24 24G90007698D24 24G90007706D24 24G90007704D24 24G90007701D24 24G90008495D24 24G90008493D24 24G90008494D24 24G90008496D24 24G90008491D24 24G90017726C24 24G90008499D24 24G90007690C24 24G90012744D24 24G90009451D24 24G90012735D24 24G90009476D24 24G90009446D24 24G90009455D24 24G90009462D24 24G90009449D24 24G90009463D24 24G90009480D24 24G90009461D24 24G90009544D24 24G90009583D24 24G90009441D24 24G90009456D24 24G90009443D24 24G90009580D24 24G90009447D24 24G90009383D24 24G90009457D24 24G90009458D24 24G90009445D24 24G90009577D24 24G90009584D24 24G90009431D24 24G90009468D24 24G90009430D24 24G90009433D24 24G90009434D24 24G90009432D24 24G90009438D24 24G90009437D24 24G90009473D24 24G90009435D24 24G90009465D24 24G90009467D24 24G90009484D24 24G90009439D24 24G90009440D24 24G90009448D24 24G90009481D24 24G90009474D24 24G90009459D24 24G90009472D24 24G90009573D24 24G90009426D24 24G90009471D24 24G90009453D24 24G90009454D24 24G90009477D24 24G90009450D24 24G90009478D24 24G90009387D24 24G90009452D24 24G90008310D24 24G90008303D24 24G90008307D24 24G90008306D24 24G90008313D24 24G90008314D24 24G90008304D24 24G90008301D24 24G90008302D24 24G90008276D24 24G90008316D24 24G90008318D24 24G90008251D24 24G90008337D24 24G90008311D24 24G90012963H24 24G90001550D24 24G90001553D24 24G90001558D24 24G90001605D24 24G90001614D24 24G90001611D24 24G90001617D24 24G90013288H24 24G90013250H24 24G90013294H24 24G90013287H24 24G90012927H24 24G90013218H24 24G90012916H24 24G90012899H24 24G90012936H24 24G90012935H24 24G90012929H24 24G90012890H24 24G90012934H24 24G90012872H24 24G90012925H24 24G90013209H24 24G90012895H24 24G90012915H24 24G90012930H24 24G90012914H24 24G90013370H24 24G90013325H24 24G90013366H24 24G90013216H24 24G90013369H24 24G90013364H24 24G90013365H24 24G90013329H24 24G90012922H24 24G90013311H24 24G90013324H24 24G90012886H24 24G90013320H24 24G90013371H24 24G90013215H24 24G90013234H24 24G90013292H24 24G90013290H24 24G90013296H24 24G90013289H24 24G90013291H24 24G90013256H24 24G90013253H24 24G90013270H24 24G90013259H24 23G90102088J23 23G90102097J23 23G90103309J23 23G90109899J23 23G90102063J23 23G90103956J23 23G90109879J23 23G90103954J23 23G90109885J23 23G90109896J23 23G90103844J23 23G90109886J23 23G90109887J23 23G90103576J23 23G90103564J23 23G90103574J23 23G90103214J23 23G90103562J23 23G90103581J23 23G90103592J23 24G90016039A24 24G90016029A24 24G90016024A24 23G90103066J23 24G90016042A24 24G90016019A24 24G90016041A24 24G90016022A24 24G90016040A24 24G90016028A24 24G90016020A24 24G90016044A24 24G90016030A24 24G90016021A24 23G90103100J23 23G90103542J23 23G90103095J23 24G90016046A24 24G90016037A24 24G90016018A24 24G90016038A24 24G90016031A24 24G90016025A24 24G90016026A24 24G90016017A24 24G90016032A24 24G90016027A24 24G90016034A24 24G90016033A24 24G90016036A24 24G90010725A24 24G90010733A24 24G90016721A24 24G90016720A24 24G90016722A24 24G90010723A24 24G90016724A24 24G90016726A24 24G90010734A24 24G90016728A24 24G90010834A24 24G90010828A24 24G90010836A24 24G90010837A24 24G90010815A24 24G90010816A24 24G90015905A24 24G90010817A24 24G90015765A24 24G90015908A24 24G90010808A24 24G90010814A24 24G90010761A24 24G90015904A24 24G90010810A24 24G90010811A24 24G90010819A24 24G90016080A24 24G90010771A24 24G90010824A24 24G90016086A24 24G90010820A24 24G90010791A24 24G90010772A24 24G90016084A24 24G90010793A24 24G90010821A24 24G90010800A24 24G90010804A24 24G90016079A24 24G90016088A24 24G90010805A24 24G90010794A24 24G90010781A24 24G90010749A24 24G90016716A24 24G90016709A24 24G90016585A24 24G90016584A24 24G90016583A24 24G90016710A24 24G90016712A24 24G90016580A24 24G90016718A24 24G90016579A24 24G90016703A24 24G90016699A24 24G90016696A24 24G90016689A24 24G90016694A24 24G90016704A24 24G90016695A24 24G90016692A24 24G90016702A24 24G90000135B24 24G90000134B24 24G90000141B24 24G90000142B24 24G90000133B24 24G90000138B24 24G90000139B24 24G90000132B24 24G90011286B24 24G90011062B24 24G90011056B24 24G90011063B24 24G90011069B24 24G90011068B24 24G90000109B24 24G90000108B24 24G90000107B24 24G90000113B24 24G90017604B24 24G90017605B24 24G90017616B24 24G90017618B24 24G90017606B24 24G90017059B24 24G90017058B24 24G90017060B24 24G90011055B24 24G90017081B24 24G90017079B24 24G90000117B24 24G90017080B24 24G90000119B24 24G90012576B24 24G90000146B24 24G90000150B24 24G90012573B24 24G90000152B24 24G90012572B24 24G90000147B24 24G90000144B24 24G90012507B24 24G90017233B24 24G90017236B24 24G90017237B24 24G90012504B24 24G90012511B24 24G90012564B24 24G90012570B24 24G90012508B24 24G90012509B24 24G90012505B24 24G90012510B24 24G90012551B24 24G90012568B24 24G90012506B24 24G90012512B24 24G90012571B24 24G90012567B24 24G90012569B24 24G90012562B24 24G90012563B24 24G90000032B24 24G90000014B24 24G90017252B24 24G90000049B24 24G90000035B24 24G90000034B24 24G90017082B24 24G90017083B24 24G90012557B24 24G90012559B24 24G90012556B24 24G90012555B24 24G90017013B24 24G90012552B24 24G90017011B24 24G90012554B24 24G90012558B24 24G90012560B24 24G90012561B24 24G90017012B24 24G90017015B24 24G90017014B24 24G90012553B24 24G90017017B24 24G90017641B24 24G90017018B24 24G90017102B24 24G90017103B24 24G90017640B24 24G90012501B24 24G90017650B24 24G90017639B24 24G90017632B24 24G90017645B24 24G90017095B24 24G90017633B24 24G90017649B24 24G90012503B24 24G90017646B24 24G90017643B24 24G90017097B24 24G90017100B24 24G90017265B24 24G90017651B24 24G90017099B24 24G90017638B24 24G90017648B24 24G90017105B24 24G90017644B24 24G90017637B24 24G90017109B24 24G90017110B24 24G90017101B24 24G90017107B24 24G90017263B24 24G90017636B24 24G90017108B24 24G90017635B24 24G90017104B24 24G90017634B24 24G90017264B24 24G90017647B24 24G90017098B24 24G90012502B24 24G90017642B24 24G90017278B24 24G90017448B24 24G90017449B24 24G90017280B24 24G90017271B24 24G90017446B24 24G90017268B24 24G90017270B24 24G90017279B24 24G90017523B24 24G90017503B24 24G90017507B24 24G90017504B24 24G90017506B24 24G90017512B24 24G90017511B24 24G90017509B24 24G90017510B24 24G90017505B24 24G90017508B24 24G90017091B24 24G90017608B24 24G90017087B24 24G90017090B24 24G90017056B24 24G90017053B24 24G90017057B24 24G90017055B24 24G90017610B24 24G90017609B24 24G90017540B24 24G90017549B24 24G90017534B24 24G90017537B24 24G90017518B24 24G90017535B24 24G90017094B24 24G90017514B24 24G90017539B24 24G90017516B24 24G90017547B24 24G90017521B24 24G90017541B24 24G90017552B24 24G90017550B24 24G90017548B24 24G90017522B24 24G90017536B24 24G90017513B24 24G90017520B24 24G90017551B24 24G90017542B24 24G90017543B24 24G90017538B24 24G90017545B24 24G90017546B24 24G90017544B24 24G90017515B24 24G90017530B24 24G90017085B24 24G90017531B24 24G90017527B24 24G90017529B24 24G90017524B24 24G90017084B24 24G90017526B24 24G90017528B24 24G90017525B24 24G90017532B24 24G90013283H24 24G90013284H24 24G90013285H24 24G90013281H24 24G90013213H24 24G90015930A24 24G90015928A24 24G90015912A24 24G90015922A24 24G90015926A24 24G90015915A24 24G90015796A24 24G90010661A24 24G90015948A24 24G90010659A24 24G90015944A24 24G90010718A24 24G90010717A24 24G90010701A24 24G90010700A24 24G90010720A24 24G90010707A24 24G90010716A24 24G90010706A24 24G90016687A24 24G90016685A24 24G90016681A24 24G90016684A24 24G90016679A24 24G90016683A24 24G90016686A24 23G90103802J23 23G90103607J23 23G90103588J23 23G90103550J23 23G90103553J23 23G90103549J23 23G90103656J23 23G90103707J23 23G90103554J23 23G90103560J23 23G90103557J23 23G90103556J23 23G90103558J23 23G90103697J23 23G90103696J23 24G90012099B24 24G90012091B24 24G90012093B24 24G90011536B24 24G90011801B24 24G90011913B24 24G90011912B24 24G90011793B24 24G90011804B24 24G90012235B24 24G90011797B24 24G90011795B24 24G90011925B24 24G90011926B24 24G90012071B24 24G90012066B24 24G90012063B24 24G90012069B24 24G90012067B24 24G90012065B24 24G90012229B24 24G90005706B24 24G90017736C24 24G90008497D24 24G90012749D24 24G90012747D24 24G90009475D24 24G90009483D24 24G90009571D24 24G90009550D24 24G90009460D24 24G90009576D24 24G90009582D24 24G90009578D24 24G90009581D24 24G90009575D24 24G90009574D24 24G90009579D24 24G90009466D24 24G90009444D24 24G90009436D24 24G90009442D24 24G90009479D24 24G90009464D24 24G90009482D24 24G90009572D24 24G90008256D24 24G90008250D24 24G90008246D24 24G90008245D24 24G90011960B24 24G90011970B24 24G90011967B24 24G90011709B24 24G90011703B24 24G90011707B24 24G90011708B24 24G90011705B24 24G90011710B24 24G90005646B24 24G90005672B24 24G90005673B24 24G90005670B24 24G90011792B24 24G90011975B24 24G90011979B24 24G90011973B24 24G90011977B24 24G90011790B24 24G90011811B24 24G90011535B24 24G90011789B24 24G90011799B24 24G90011922B24 24G90011920B24 24G90011814B24 24G90011921B24 24G90011816B24 24G90011820B24 24G90011766B24 24G90011767B24 24G90011817B24 24G90011903B24 24G90011821B24 24G90011818B24 24G90011531B24 24G90011779B24 24G90011777B24 24G90011534B24 24G90011771B24 24G90011773B24 24G90011947B24 24G90011938B24 24G90011939B24 24G90011945B24 24G90011940B24 24G90011952B24 24G90011802B24 24G90011808B24 24G90011971B24 24G90011950B24 24G90011955B24 24G90011810B24 24G90011969B24 24G90011803B24 24G90011809B24 24G90011949B24 24G90011948B24 24G90011953B24 24G90011956B24 24G90005308B24 24G90005307B24 24G90005311B24 24G90005663B24 24G90005304B24 24G90005305B24 24G90005298B24 24G90005668B24 24G90005666B24 24G90005661B24 24G90005669B24 24G90011785B24 24G90011786B24 24G90011781B24 24G90011780B24 24G90005701B24 24G90005708B24 24G90005684B24 24G90005687B24 24G90005691B24 24G90005694B24 24G90005696B24 24G90005688B24 24G90005266B24 24G90005262B24 24G90005693B24 24G90005282B24 24G90005273B24 24G90005704B24 24G90011676B24 24G90005275B24 24G90005278B24 24G90011931B24 24G90011933B24 24G90011932B24 24G90011934B24 24G90011539B24 24G90011696B24 24G90011698B24 24G90011700B24 24G90010623A24 24G90010624A24 24G90010622A24 24G90010613A24 24G90010612A24 24G90015855A24 24G90015861A24 24G90015865A24 24G90015870A24 24G90011194B24 24G90011005B24 24G90011004B24 24G90010992B24 24G90011006B24 24G90011187B24 24G90010988B24 24G90011029B24 24G90011030B24 24G90011035B24 24G90011226B24 24G90011034B24 24G90011076B24 24G90010980B24 24G90011044B24 24G90011051B24 24G90011038B24 24G90011050B24 24G90011039B24 24G90011037B24 24G90011047B24 24G90011045B24 24G90011212B24 24G90011214B24 24G90011024B24 24G90011344B24 24G90011346B24 24G90011221B24 24G90011206B24 24G90011205B24 24G90010649A24 24G90010651A24 24G90010641A24 24G90010647A24 24G90010643A24 24G90010645A24 24G90010646A24 24G90015801A24 24G90010650A24 24G90010640A24 24G90010626A24 24G90013178H24 24G90012910H24 24G90012961H24 24G90012964H24 24G90001606D24 24G90001603D24 24G90001629D24 24G90001604D24 24G90001609D24 24G90001613D24 24G90012921H24 24G90001666D24 24G90001670D24 24G90001549D24 24G90001563D24 24G90001564D24 24G90001567D24 24G90001552D24 24G90001466D24 24G90001568D24 24G90001545D24 24G90001546D24 24G90001586D24 24G90001587D24 24G90001656D24 24G90001639D24 24G90001641D24 24G90001642D24 24G90001635D24 24G90001644D24 24G90001657D24 24G90001584D24 24G90001593D24 24G90001647D24 24G90001640D24 24G90001636D24 24G90001659D24 24G90001637D24 24G90001638D24 24G90001664D24 24G90001658D24 24G90001588D24 24G90001591D24 24G90001599D24 24G90001598D24 24G90001626D24 24G90001654D24 24G90001652D24 24G90001601D24 24G90001597D24 24G90001590D24 24G90001594D24 24G90001589D24 24G90010736A24 24G90010726A24 24G90015860A24 24G90016004A24 24G90016000A24 24G90010731A24 24G90010737A24 24G90016003A24 24G90010727A24 24G90016002A24 24G90010728A24 24G90010729A24 24G90015859A24 24G90016006A24 24G90015858A24 24G90015857A24 24G90010735A24 24G90016005A24 24G90015856A24 24G90010724A24 24G90016001A24 24G90010732A24 24G90015869A24 24G90015864A24 24G90010835A24 24G90015866A24 24G90015867A24 24G90015851A24 24G90015862A24 24G90015863A24 24G90015853A24 24G90010830A24 24G90015854A24 24G90015852A24 24G90010833A24 24G90016069A24 24G90015881A24 24G90010843A24 24G90010841A24 24G90010842A24 24G90010839A24 24G90016072A24 24G90016075A24 24G90010845A24 24G90016068A24 24G90010813A24 24G90016078A24 24G90010809A24 24G90016070A24 24G90010844A24 24G90010758A24 24G90010818A24 24G90010780A24 24G90010822A24 24G90010807A24 24G90010796A24 24G90010785A24 24G90010827A24 24G90010792A24 24G90010788A24 24G90010803A24 24G90010826A24 24G90010823A24 24G90010991B24 24G90010755A24 24G90010750A24 24G90010752A24 24G90016521A24 24G90016515A24 24G90011188B24 24G90015880A24 24G90015877A24 24G90015872A24 24G90015879A24 24G90015871A24 24G90015875A24 24G90015876A24 24G90011192B24 24G90011189B24 24G90010990B24 24G90017074B24 24G90011196B24 24G90010989B24 24G90010997B24 24G90010999B24 24G90010987B24 24G90011336B24 24G90017624B24 24G90017621B24 24G90000129B24 24G90011183B24 24G90011179B24 24G90011186B24 24G90017495B24 24G90011330B24 24G90000136B24 24G90000137B24 24G90011013B24 24G90011031B24 24G90011033B24 24G90011032B24 24G90011293B24 24G90000128B24 24G90011074B24 24G90010984B24 24G90000089B24 24G90000087B24 24G90000091B24 24G90010978B24 24G90011061B24 24G90011011B24 24G90011058B24 24G90011059B24 24G90011012B24 24G90011070B24 24G90000084B24 24G90000085B24 24G90000123B24 24G90011025B24 24G90000106B24 24G90000111B24 24G90011054B24 24G90011046B24 24G90011049B24 24G90012582B24 24G90011213B24 24G90000121B24 24G90000118B24 24G90000122B24 24G90017075B24 24G90017078B24 24G90011340B24 24G90000145B24 24G90012575B24 24G90012589B24 24G90012577B24 24G90000148B24 24G90012596B24 24G90012593B24 24G90012597B24 24G90012600B24 24G90012594B24 24G90012599B24 24G90012598B24 24G90012595B24 24G90012592B24 24G90012513B24 24G90011359B24 24G90011018B24 24G90011341B24 24G90011017B24 24G90011352B24 24G90011016B24 24G90000101B24 24G90000050B24 24G90017496B24 24G90017486B24 24G90017488B24 24G90017484B24 24G90017493B24 24G90017595B24 24G90017073B24 24G90017076B24 24G90017077B24 24G90011701B24 24G90011702B24 24G90011211B24 24G90011208B24 24G90011706B24 24G90011704B24 24G90005650B24 24G90011225B24 24G90005676B24 24G90005649B24 24G90011216B24 24G90005647B24 24G90005648B24 24G90005659B24 24G90005658B24 24G90005674B24 24G90011207B24 24G90011198B24 24G90005671B24 24G90005678B24 24G90011204B24 24G90005656B24 24G90011202B24 24G90005655B24 24G90011220B24 24G90005679B24 24G90005653B24 24G90005675B24 24G90005677B24 24G90005657B24 24G90017586B24 24G90017585B24 24G90017587B24 24G90017594B24 24G90017592B24 24G90011755B24 24G90017588B24 24G90011794B24 24G90011761B24 24G90011923B24 24G90011822B24 24G90017591B24 24G90011533B24 24G90011776B24 24G90011778B24 24G90011521B24 24G90017574B24 24G90017578B24 24G90011942B24 24G90017573B24 24G90017566B24 24G90011944B24 24G90011936B24 24G90017583B24 24G90017580B24 24G90017565B24 24G90017567B24 24G90011943B24 24G90017581B24 24G90017569B24 24G90017568B24 24G90017579B24 24G90017576B24 24G90017584B24 24G90017575B24 24G90017572B24 24G90017582B24 24G90011951B24 24G90017564B24 24G90011954B24 24G90017562B24 24G90011744B24 24G90017559B24 24G90011746B24 24G90012516B24 24G90017557B24 24G90017560B24 24G90017561B24 24G90017563B24 24G90012522B24 24G90011747B24 24G90005494B24 24G90012528B24 24G90005493B24 24G90012530B24 24G90012525B24 24G90005492B24 24G90012537B24 24G90012523B24 24G90012527B24 24G90012533B24 24G90012531B24 24G90012526B24 24G90012524B24 24G90011957B24 24G90005491B24 24G90005664B24 24G90005260B24 24G90005299B24 24G90005667B24 24G90005261B24 24G90005665B24 24G90005660B24 24G90005662B24 24G90005301B24 24G90005300B24 24G90005303B24 24G90011788B24 24G90011783B24 24G90017481B24 24G90012521B24 24G90017602B24 24G90017596B24 24G90012519B24 24G90012515B24 24G90012517B24 24G90017600B24 24G90017483B24 24G90017599B24 24G90012520B24 24G90017601B24 24G90012514B24 24G90010644A24 24G90010642A24 24G90010631A24 24G90010658A24 24G90015943A24 24G90010715A24 24G90010692A24 24G90010722A24 24G90010694A24 24G90010691A24 24G90010687A24 24G90010683A24 24G90010721A24 24G90010719A24 24G90010710A24 24G90010693A24 24G90010699A24 24G90010684A24 24G90010698A24 24G90010711A24 24G90010703A24 24G90010712A24 24G90010708A24 24G90010704A24 24G90010713A24 24G90010685A24 24G90010686A24 24G90010695A24 24G90010689A24 24G90010705A24 24G90017554B24 24G90017629B24 24G90017631B24 24G90017630B24 24G90017555B24 24G90005692B24 24G90005690B24 24G90005699B24 24G90005695B24 24G90005685B24 24G90005707B24 24G90005697B24 24G90005689B24 24G90005686B24 24G90005682B24 24G90005683B24 24G90005680B24 24G90005681B24 24G90005703B24 24G90005277B24 24G90005269B24 24G90005700B24 24G90005705B24 24G90005702B24 24G90005281B24 24G90005279B24 24G90005267B24 24G90011550B24 24G90011930B24 24G90011935B24 24G90011695B24 24G90011547B24 24G90011699B24 24G90012913H24 24G90017999H24 24G90012874H24 24G90012926H24 24G90012868H24 24G90012900H24 24G90012945H24 24G90012875H24 24G90012893H24 24G90012955H24 24G90012907H24 24G90012908H24 24G90012909H24 24G90012870H24 24G90012871H24 24G90017996H24 24G90018002H24 24G90012902H24 24G90018000H24 24G90012928H24 24G90012892H24 24G90012896H24 24G90012924H24 24G90012887H24 24G90012957H24 24G90012923H24 24G90018030H24 24G90018021H24 24G90018025H24 24G90018034H24 24G90011427H24 24G90011429H24 24G90018033H24 24G90011430H24 24G90013672H24 24G90018017H24 24G90011428H24 24G90018031H24 24G90011426H24 24G90011431H24 24G90018024H24 24G90018028H24 24G90018018H24 24G90018032H24 24G90018009H24 24G90013678H24 24G90006847D24 24G90006848D24 24G90006839D24 24G90006837D24 24G90006854D24 24G90006836D24 24G90006838D24 24G90001671D24 24G90001682D24 24G90001684D24 24G90001673D24 24G90001676D24 24G90001675D24 24G90001677D24 24G90001674D24 24G90001683D24 24G90001678D24 24G90001672D24 24G90001681D24 24G90001679D24 24G90001562D24 24G90001542D24 24G90001560D24 24G90001538D24 24G90001551D24 24G90001557D24 24G90001565D24 24G90001559D24 24G90001536D24 24G90001547D24 24G90001555D24 24G90001548D24 24G90001556D24 24G90001561D24 24G90001537D24 24G90010350D24 24G90010345D24 24G90010347D24 24G90010346D24 24G90010344D24 24G90010348D24 24G90010351D24 24G90010349D24 24G90001883D24 24G90001884D24 24G90001686D24 24G90001891D24 24G90001655D24 24G90001543D24 24G90001886D24 24G90001544D24 24G90001888D24 24G90001885D24 24G90001887D24 24G90001687D24 24G90012952H24 24G90012939H24 24G90012894H24 23G90100150B24 24G90001669D24 24G90001668D24 24G90011980B24 24G90011964B24 24G90011966B24 24G90011965B24 24G90011968B24 24G90011976B24 24G90011974B24 24G90011978B24 24G90011774B24 24G90011775B24 24G90011946B24 24G90011937B24 24G90011806B24 24G90011958B24 24G90011784B24 24G90011782B24 24G90011787B24 24G90005268B24 24G90005276B24 24G90001651D24 24G90013295H24 24G90013220H24 24G90013257H24 24G90013183H24 24G90013271H24 24G90013286H24 24G90013175H24 24G90013293H24 24G90013179H24 24G90013174H24 24G90013217H24 24G90007313H24 24G90012917H24 24G90012912H24 24G90012873H24 24G90012876H24 24G90012932H24 24G90013265H24 24G90012891H24 24G90012897H24 24G90013258H24 24G90012869H24 24G90012931H24 24G90012918H24 24G90013269H24 24G90012898H24 24G90013375H24 24G90012919H24 24G90012953H24 24G90012906H24 24G90012888H24 24G90013210H24 24G90012904H24 24G90012940H24 24G90012905H24 24G90012903H24 24G90012901H24 24G90013322H24 24G90013323H24 24G90013357H24 24G90013368H24 24G90012960H24 24G90013327H24 24G90013326H24 24G90013317H24 24G90013374H24 24G90012937H24 24G90013373H24 24G90012966H24 24G90013367H24 24G90013321H24 24G90013372H24 24G90013328H24 24G90012879H24 24G90012885H24 24G90013319H24 24G90013316H24 24G90013318H24 24G90013229H24 24G90013225H24 24G90013227H24 24G90013211H24 24G90013222H24 24G90013226H24 24G90013297H24 24G90013331H24 24G90013254H24 24G90013260H24 24G90013255H24 24G90013248H24 24G90013249H24 24G90013230H24 23G90109799J23 23G90103856J23 23G90102098J23 23G90103852J23 23G90103850J23 23G90102095J23 23G90102087J23 23G90102086J23 23G90103848J23 23G90102096J23 23G90102458K23 23G90102460K23 23G90102462K23 23G90102461K23 23G90102455K23 23G90102464K23 23G90102295K23 23G90102838K23 23G90102463K23 23G90102836K23 23G90109816J23 24G90016443A24 24G90016350A24 24G90016352A24 23G90103310J23 23G90103308J23 23G90103307J23 23G90103316J23 23G90109861J23 23G90102843K23 23G90103339J23 23G90102848K23 23G90102860K23 23G90109897J23 23G90102846K23 23G90102858K23 23G90102280K23 23G90102276K23 23G90102844K23 23G90102853K23 23G90102212K23 23G90102277K23 23G90102856K23 23G90102851K23 23G90102281K23 23G90102279K23 23G90103341J23 23G90103511J23 23G90103991J23 23G90103652J23 23G90103998J23 23G90103958J23 23G90103957J23 23G90109893J23 23G90103858J23 23G90103959J23 23G90109884J23 23G90102103J23 23G90102061J23 23G90102062J23 23G90103649J23 23G90103668J23 23G90109880J23 23G90103953J23 23G90103999J23 23G90103997J23 23G90103955J23 23G90104000J23 23G90109878J23 23G90103996J23 23G90103364J23 23G90103983J23 23G90102368K23 23G90102365K23 23G90102864K23 23G90102473K23 23G90102477K23 23G90102467K23 23G90102862K23 23G90102469K23 23G90102255K23 23G90103518J23 23G90103594J23 23G90103570J23 23G90103369J23 23G90103215J23 23G90103216J23 23G90103572J23 23G90103543J23 23G90103084J23 23G90103222J23 23G90103537J23 23G90103544J23 23G90103539J23 23G90103540J23 23G90103087J23 23G90103541J23 23G90103234J23 23G90103536J23 23G90103538J23 24G90016023A24 23G90102228K23 23G90102239K23 23G90102223K23 23G90102182K23 23G90102224K23 23G90102230K23 23G90102232K23 23G90102229K23 23G90102222K23 23G90102241K23 23G90102237K23 23G90102213K23 23G90102184K23 23G90102233K23 24G90016627A24 23G90102236K23 23G90102226K23 23G90102225K23 23G90102235K23 23G90102221K23 23G90102231K23 23G90102234K23 23G90102274K23 23G90102227K23 23G90102238K23 24G90016723A24 24G90010730A24 24G90016727A24 24G90016077A24 24G90016076A24 24G90016082A24 24G90016087A24 24G90016760A24 24G90016763A24 24G90016765A24 24G90016758A24 24G90016767A24 24G90016757A24 24G90016768A24 24G90016764A24 24G90016762A24 24G90016761A24 24G90016766A24 24G90016715A24 24G90016713A24 24G90016586A24 24G90016582A24 24G90016717A24 24G90016505A24 24G90016517A24 24G90016511A24 24G90016504A24 24G90016520A24 24G90016440A24 24G90016496A24 24G90016503A24 24G90016392A24 24G90016502A24 24G90016394A24 24G90016439A24 24G90016380A24 24G90016494A24 24G90016438A24 24G90016512A24 24G90016385A24 24G90016497A24 24G90016509A24 24G90016390A24 24G90016381A24 24G90016389A24 24G90016388A24 24G90016488A24 24G90016490A24 24G90016384A24 24G90016386A24 24G90016441A24 24G90016437A24 24G90016500A24 24G90016510A24 24G90016495A24 24G90016491A24 24G90016492A24 24G90016379A24 24G90016393A24 24G90016516A24 24G90016387A24 24G90016501A24 24G90016487A24 24G90016391A24 24G90016506A24 24G90016382A24 24G90016508A24 24G90016493A24 24G90016507A24 24G90016486A24 24G90016499A24 24G90016513A24 24G90016383A24 24G90016519A24 24G90016489A24 24G90016498A24 24G90016518A24 24G90016334A24 24G90016474A24 24G90016475A24 24G90016376A24 24G90016332A24 24G90016325A24 24G90016473A24 24G90016338A24 24G90016339A24 24G90016335A24 24G90016373A24 24G90016479A24 24G90016331A24 24G90016341A24 24G90016481A24 24G90016330A24 24G90016327A24 24G90016480A24 24G90016482A24 24G90016485A24 24G90016378A24 24G90016377A24 24G90016337A24 24G90016484A24 24G90016324A24 24G90016326A24 24G90016375A24 24G90016374A24 24G90016476A24 24G90016329A24 24G90016478A24 24G90016483A24 24G90016781A24 24G90016782A24 24G90016769A24 24G90016771A24 24G90016770A24 24G90016778A24 24G90016773A24 24G90016776A24 24G90016777A24 24G90016774A24 24G90016775A24 24G90016780A24 24G90001667D24 24G90017613B24 24G90017612B24 24G90017620B24 24G90011588B24 24G90011600B24 24G90011591B24 24G90011589B24 24G90011590B24 24G90017611B24 24G90017623B24 24G90017619B24 24G90017502B24 24G90000130B24 24G90011290B24 24G90011292B24 24G90011075B24 24G90000090B24 24G90000086B24 24G90010979B24 24G90011067B24 24G90011060B24 24G90000099B24 24G90000112B24 24G90000098B24 24G90000104B24 24G90017052B24 24G90017050B24 24G90017043B24 24G90017048B24 24G90017044B24 24G90017047B24 24G90017045B24 24G90017051B24 24G90017603B24 24G90017046B24 24G90000116B24 24G90000120B24 24G90000115B24 24G90000114B24 24G90000143B24 24G90000149B24 24G90000151B24 24G90017498B24 24G90017500B24 24G90000102B24 24G90000103B24 24G90000051B24 24G90017061B24 24G90017501B24 24G90017062B24 24G90017499B24 24G90017497B24 24G90017489B24 24G90017490B24 24G90017491B24 24G90017492B24 24G90017487B24 24G90017069B24 24G90017066B24 24G90017071B24 24G90017072B24 24G90017065B24 24G90017068B24 24G90017028B24 24G90017041B24 24G90017037B24 24G90017036B24 24G90017039B24 24G90017022B24 24G90017025B24 24G90017038B24 24G90017019B24 24G90017026B24 24G90017020B24 24G90017027B24 24G90017024B24 24G90017040B24 24G90017021B24 24G90017029B24 24G90017031B24 24G90017030B24 24G90017450B24 24G90011529B24 24G90011596B24 24G90011712B24 24G90011752B24 24G90011753B24 24G90011598B24 24G90011764B24 24G90011733B24 24G90011593B24 24G90011594B24 24G90011592B24 24G90011751B24 24G90017590B24 24G90011763B24 24G90017593B24 24G90011759B24 24G90011765B24 24G90011599B24 24G90011512B24 24G90011511B24 24G90011760B24 24G90017589B24 24G90011757B24 24G90011762B24 24G90011750B24 24G90011758B24 24G90011768B24 24G90011769B24 24G90011739B24 24G90011736B24 24G90011528B24 24G90011595B24 24G90011754B24 24G90011737B24 24G90011524B24 24G90011526B24 24G90011530B24 24G90011714B24 24G90011711B24 24G90011525B24 24G90011522B24 24G90011713B24 24G90011507B24 24G90011516B24 24G90011520B24 24G90011506B24 24G90017494B24 24G90011504B24 24G90011514B24 24G90017571B24 24G90011505B24 24G90011502B24 24G90017577B24 24G90011519B24 24G90011509B24 24G90011513B24 24G90011518B24 24G90011515B24 24G90011501B24 24G90011508B24 24G90011510B24 24G90011503B24 24G90011517B24 24G90011749B24 24G90017042B24 24G90011748B24 24G90017558B24 24G90017556B24 24G90011741B24 24G90011740B24 24G90011743B24 24G90011745B24 24G90017607B24 24G90011742B24 24G90017064B24 24G90017067B24 24G90017482B24 24G90017598B24 24G90017597B24 24G90008404D24 24G90008387D24 24G90008358D24 24G90008383D24 24G90008380D24 24G90008363D24 24G90008360D24 24G90008364D24 24G90008361D24 24G90008359D24 24G90008378D24 24G90013282H24 23G90107047J23 23G90107058J23 23G90107057J23 23G90107712J23 23G90102166K23 23G90107046J23 23G90107054J23 23G90107060J23 23G90107061J23 23G90107053J23 23G90107056J23 23G90105394K23 23G90107713J23 23G90105644K23 24G90010702A24 24G90010697A24 24G90010696A24 23G90102248K23 23G90102250K23 23G90102466K23 23G90102449K23 23G90103012J23 23G90103804J23 23G90103801J23 23G90103803J23 23G90103015J23 23G90102445K23 23G90102465K23 23G90102447K23 23G90103800J23 23G90103799J23 23G90103712J23 23G90103605J23 23G90103606J23 23G90103609J23 23G90103597J23 23G90103659J23 23G90103684J23 23G90103660J23 23G90103608J23 23G90102354K23 23G90102448K23 23G90107563K23 23G90102246K23 23G90102377K23 23G90102243K23 23G90102247K23 23G90102244K23 23G90107561K23 23G90103545J23 23G90103552J23 23G90102440K23 23G90102438K23 23G90102351K23 23G90102825J23 23G90103555J23 23G90103694J23 23G90103705J23 23G90107448J23 23G90107447J23 23G90102633J23 23G90102822J23 23G90102375K23 23G90103686J23 23G90103688J23 23G90103687J23 23G90103681J23 23G90103680J23 23G90103658J23 23G90103678J23 23G90103677J23 23G90103691J23 23G90103689J23 23G90103683J23 23G90102398K23 23G90102395K23 23G90103692J23 23G90103690J23 23G90103703J23 23G90103699J23 23G90103702J23 23G90102832K23 23G90102385K23 23G90102376K23 23G90102383K23 23G90102363K23 23G90102249K23 23G90102380K23 23G90102381K23 23G90102378K23 23G90107710J23 23G90102831J23 24G90017035B24 24G90017034B24 24G90017033B24 24G90017628B24 24G90017032B24 24G90017553B24 24G90011677B24 24G90011678B24 24G90011667B24 24G90011664B24 24G90011663B24 24G90011675B24 24G90011673B24 24G90011680B24 24G90011671B24 24G90011672B24 24G90011679B24 24G90011662B24 24G90011661B24 24G90011674B24 24G90011545B24 24G90011543B24 24G90011542B24 24G90011541B24 24G90011684B24 24G90011685B24 24G90011655B24 24G90011691B24 24G90011665B24 24G90011666B24 24G90011669B24 24G90011670B24 24G90011668B24 24G90011656B24 24G90011657B24 24G90011694B24 24G90011693B24 24G90011692B24 24G90011654B24 24G90011546B24 24G90011548B24 24G90011653B24 24G90011651B24 24G90011690B24 24G90011683B24 24G90011658B24 24G90011652B24 24G90011686B24 24G90011660B24 24G90011659B24 24G90011689B24 24G90011688B24 24G90011687B24 24G90011681B24 24G90001661D24 24G90001663D24 24G90001653D24 24G90001649D24 24G90001619D24 24G90001632D24 24G90001650D24 24G90001600D24 24G90012956H24 24G90012944H24 24G90013069H24 24G90013071H24 24G90013068H24 24G90012949H24 24G90012942H24 24G90018108H24 24G90018109H24 24G90018106H24 24G90018089H24 24G90018091H24 24G90018086H24 24G90001616D24 24G90011191B24 24G90011197B24 24G90010994B24 24G90010995B24 24G90010993B24 24G90010985B24 24G90011066B24 24G90011077B24 24G90011010B24 24G90011009B24 24G90011048B24 24G90011036B24 24G90011360B24 24G90011357B24 24G90011351B24 24G90011354B24 24G90011021B24 24G90011358B24 24G90011356B24 24G90012095B24 24G90011209B24 24G90012097B24 24G90012105B24 24G90012098B24 24G90011200B24 24G90011199B24 24G90011203B24 24G90011201B24 24G90011910B24 24G90011904B24 24G90005274B24 24G90005265B24 24G90001595D24 24G90001628D24 24G90001585D24 24G90001631D24 24G90001633D24 24G90001643D24 24G90001646D24 24G90001630D24 24G90001634D24 24G90001648D24 24G90001618D24 24G90001592D24 24G90004246F24 24G90004248F24 24G90007430F24 24G90007429F24 24G90007422F24 24G90007425F24 24G90007428F24 24G90007426F24 24G90004244F24 24G90007496F24 24G90007424F24 24G90004249F24 24G90001831F24 24G90007431F24 24G90007494F24 24G90001825F24 24G90001834F24 24G90007432F24 24G90007499F24 24G90007500F24 24G90007490F24 24G90007493F24 24G90001830F24 24G90007497F24 24G90007423F24 24G90007378F24 24G90007427F24 24G90004243F24 24G90009500F24 24G90009499F24 24G90004242F24 24G90004245F24 24G90007419F24 24G90004247F24 24G90001827F24 24G90001828F24 24G90001823F24 24G90001824F24 24G90001826F24 24G90001822F24 24G90010355F24 24G90001829F24 24G90001833F24 24G90007495F24 24G90001608D24 23G90000314H23 23G90000312H23 23G90000316H23 23G90000313H23 23G90102256K23 23G90009051H23 23G90009407H23 23G90009403H23 23G90009410H23 23G90009413H23 23G90009449H23 23G90009626H23 23G90009425H23 23G90009624H23 23G90009629H23 23G90009769H23 23G90009771H23 23G90009630H23 23G90009622H23 23G90009053H23 23G90009073H23 23G90009071H23 23G90009429H23 23G90009550H23 23G90009473H23 23G90009472H23 23G90009614H23 23G90000371H23 23G90009474H23 23G90009619H23 23G90009668H23 23G90000373H23 23G90000375H23 23G90009554H23 23G90009476H23 23G90009618H23 23G90009610H23 23G90009611H23 23G90009556H23 23G90000372H23 23G90009549H23 23G90009606H23 23G90009609H23 23G90009612H23 23G90009552H23 23G90009553H23 23G90009669H23 23G90000374H23 23G90009030H23 23G90000306H23 23G90009524H23 23G90009632H23 23G90000308H23 23G90000369H23 23G90000304H23 23G90000370H23 23G90009593H23 23G90000303H23 23G90000302H23 23G90009280H23 23G90009275H23 23G90009279H23 23G90009586H23 23G90009281H23 23G90009282H23 23G90009589H23 23G90009585H23 23G90009597H23 23G90009594H23 23G90009276H23 23G90009277H23 23G90009599H23 23G90009753H23 23G90009528H23 23G90009526H23 23G90009640H23 23G90009651H23 23G90009652H23 23G90009650H23 23G90009271H23 23G90009038H23 23G90009285H23 23G90009443H23 23G90009445H23 23G90009603H23 23G90009601H23 23G90009607H23 23G90009613H23 23G90009608H23 23G90009604H23 23G90009522H23 23G90009748H23 23G90009034H23 23G90009283H23 23G90009438H23 23G90009436H23 23G90009602H23 23G90009284H23 23G90009649H23 23G90009646H23 23G90009536H23 23G90009653H23 23G90009535H23 23G90009537H23 23G90009525H23 23G90009529H23 23G90009760H23 23G90009268H23 23G90009272H23 23G90009273H23 23G90009639H23 23G90009523H23 23G90009664H23 23G90009663H23 23G90009662H23 23G90009655H23 23G90009661H23 23G90009581H23 23G90009579H23 23G90009660H23 23G90009656H23 23G90009659H23 23G90009657H23 23G90009658H23 23G90009583H23 23G90009584H23 23G90009582H23 24G90010323D24 24G90010321D24 24G90010310D24 24G90009619D24 24G90010035D24 24G90010029D24 24G90010027D24 24G90010449D24 24G90003355D24 24G90010032D24 24G90010033D24 24G90018912D24 24G90007433F24 24G90012089B24 24G90012094B24 24G90012120B24 24G90012118B24 24G90011227B24 24G90015878A24 24G90015874A24 24G90015868A24 24G90008247D24 24G90001625D24 24G90011961B24 24G90011770B24 24G90001627D24 24G90001615D24 24G90011800B24 24G90011772B24 24G90001596D24 24G90012061B24 24G90001607D24 24G90012965H24 24G90001554D24 24G90001645D24 24G90012121B24 24G90009715D24 24G90011798B24 24G90011807B24 24G90011537B24 24G90011532B24 24G90011193B24 24G90011190B24 24G90011007B24 24G90011000B24 24G90010996B24 24G90011001B24 24G90011002B24 24G90011332B24 24G90010986B24 24G90011310B24 24G90011283B24 24G90011311B24 24G90011312B24 24G90011313B24 24G90017211B24 24G90011307B24 24G90011309B24 24G90011185B24 24G90011182B24 24G90011180B24 24G90011181B24 24G90011178B24 24G90011329B24 24G90011334B24 24G90011284B24 24G90011288B24 24G90011285B24 24G90011316B24 24G90011287B24 24G90011071B24 24G90011299B24 24G90011297B24 24G90011015B24 24G90011301B24 24G90011014B24 24G90011008B24 24G90017232B24 24G90011296B24 24G90011303B24 24G90011300B24 24G90011304B24 24G90011027B24 24G90011298B24 24G90011210B24 24G90011337B24 24G90011215B24 24G90000005B24 24G90000002B24 24G90000003B24 24G90017234B24 24G90017235B24 24G90000006B24 24G90011020B24 24G90011328B24 24G90011019B24 24G90011022B24 24G90011023B24 24G90000052B24 24G90000027B24 24G90017257B24 24G90000062B24 24G90000019B24 24G90000037B24 24G90000068B24 24G90000028B24 24G90000077B24 24G90000074B24 24G90017241B24 24G90000057B24 24G90000073B24 24G90000072B24 24G90000008B24 24G90000054B24 24G90000044B24 24G90000061B24 24G90000022B24 24G90000056B24 24G90000041B24 24G90000063B24 24G90000023B24 24G90000012B24 24G90000055B24 24G90000070B24 24G90000020B24 24G90000060B24 24G90000043B24 24G90017250B24 24G90000075B24 24G90000013B24 24G90017249B24 24G90000040B24 24G90000047B24 24G90017244B24 24G90017256B24 24G90000053B24 24G90000059B24 24G90000046B24 24G90000007B24 24G90017240B24 24G90000058B24 24G90017245B24 24G90017243B24 24G90017251B24 24G90000045B24 24G90017255B24 24G90000029B24 24G90000033B24 24G90000031B24 24G90000009B24 24G90000071B24 24G90000069B24 24G90000036B24 24G90017261B24 24G90017262B24 24G90017258B24 24G90017260B24 24G90017259B24 24G90017267B24 24G90017276B24 24G90017274B24 24G90017286B24 24G90011224B24 24G90011218B24 24G90017287B24 24G90011223B24 24G90017275B24 24G90017273B24 24G90011222B24 24G90017225B24 24G90017231B24 24G90017219B24 24G90017212B24 24G90017230B24 24G90017220B24 24G90017228B24 24G90017229B24 24G90017224B24 24G90017218B24 24G90017215B24 24G90017222B24 24G90017213B24 24G90017221B24 24G90017227B24 24G90017223B24 24G90017216B24 24G90017217B24 24G90000010B24 24G90011295B24 24G90000039B24 24G90011219B24 24G90012297B24 24G90012119B24 24G90012920H24 24G90012962H24 24G90001623D24 24G90012968H24 24G90001566D24 24G90012889H24 24G90001602D24

Other devices under the same FDA submission

16 devices cleared under P160008 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page