Skip to content

Pad-Pak Pad-Pak-01Recalls & reported problems

Heartsine Technologies Ltd · Battery and Electrode Cartridge · Class 3

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Heartsine Pad-Pak battery and electrode cartridge for a heart defibrillator has 529 adverse event reports, with no deaths and no injuries. Most reports describe a bent locator pin, along with depleted batteries that stop the device from powering on. One recall is open, begun in April 2023, because depleted battery cells may leave the cartridge unable to let the defibrillator analyze a patient or deliver therapy. Reports rose from 11 in the prior 12 months to 505 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
529
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 529 adverse event reports that list this exact device (DI 05060167120466), received between Feb 21, 2020 and Apr 23, 2026 — classified as 529 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (512 reports).

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (P160008); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 9 reportsNov 2025: 82 reportsDec 2025: 205 reportsJan 2026: 110 reportsFeb 2026: 102 reportsMar 2026: 5 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026205
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 26, 2023Open, Classified

    Single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator from analyzing patient condition or delivering therapy correctly.

    Z-2143-2023 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Material Deformation512
  • Battery Problem10
  • Power Problem6
  • Therapeutic or Diagnostic Output Failure5
  • Failure to Power Up3
  • Protective Measures Problem2
  • Audible Prompt/Feedback Problem1
  • Connection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions528
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 529 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Material Deformation

    THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968412 ↗
  • MalfunctionMar 30, 2026Failure to Power Up; Power Problem

    THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE'S BATTERY WAS DEPLETED. AS A RESULT, THE DEVICE MAY NOT BE ABLE TO POWER ON TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24735693 ↗
  • MalfunctionMar 23, 2026Battery Problem

    THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE'S PAD-PAK WAS DEAD, AND DOES NOT POWER ON THE DEVICE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24671448 ↗
  • MalfunctionMar 23, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24665318 ↗
  • MalfunctionMar 23, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24665137 ↗
  • MalfunctionMar 23, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24665133 ↗
  • MalfunctionFeb 26, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455242 ↗
  • MalfunctionFeb 26, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24454834 ↗
  • MalfunctionFeb 26, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452688 ↗
  • MalfunctionFeb 26, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452392 ↗
  • MalfunctionFeb 26, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452391 ↗
  • MalfunctionFeb 26, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452331 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24443000 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442986 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442980 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442978 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442977 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442976 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442835 ↗
  • MalfunctionFeb 25, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442833 ↗
  • MalfunctionFeb 23, 2026Failure to Power Up

    THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR PAD-PAK DID NOT POWER ON AN ASSOCIATED SAM PAD DEVICE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24414116 ↗
  • MalfunctionFeb 23, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413942 ↗
  • MalfunctionFeb 23, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413385 ↗
  • MalfunctionFeb 23, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413144 ↗
  • MalfunctionFeb 20, 2026Material Deformation

    IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24406199 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160008. One submission often covers many models, so check the product and lot details in each record.

  • Z-3120-2026 · Jul 27, 2026 · Open, Classified
    HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (American Airlines) - 2 Min CPR UDI-DI code: 0506016712207 350P U
  • Z-3270-2026 · Jul 27, 2026 · Open, Classified
    HeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (Americ
  • Z-0288-2026 · Sep 18, 2025 · Open, Classified
    HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J
  • Z-2174-2025 · Jun 30, 2025 · Open, Classified
    HeartSine SAM 350P, HeartSine SAM 360P, and HeartSine SAM 450P The HeartSine samaritan PAD (Public Access Defibrillator)
  • Z-1217-2023 · Feb 1, 2023 · Open, Classified
    Heartsine Samaritan PAD: SAM 350P semi-automatic defibrillator, SAM 360P fully automatic defibrillator, SAM 500P semi-automatic defibrillator with CPR Advisor
  • Z-0745-2022 · Jan 13, 2022 · Terminated
    HeartSine Samaritan PAD, SAM 350P Semi-Automatic Defibrillator, used with Pad-Pak

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.