Ciaglia Blue Rhino recallZ-1781-2026 · Cook Incorporated · initiated Mar 5, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
FDA root cause category: Incorrect or no expiration date
What the company is doing
Cook Medical issued an URGENT: MEDICAL DEVICE RECALL to its consignee on 3/5 2026 via courier or email. The notice explained the issue, potential risk to health, and requested the following: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Recall details
- Recall number
- Z-1781-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 5, 2026
- Posted by the FDA
- Apr 6, 2026
- Recalling firm
- Cook Incorporated
- Quantity
- 484 units
- Product code
- JOH
- FDA submissions
- K133597
- Distribution
- Worldwide distribution.
Product description
COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Order Number G12115.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270812(10)NS16200246, Lot Number NS16200246; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270803(10)NS16189601, Lot Number NS16189601; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270827(10)NS16226797, Lot Number NS16226797; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270812(10)NS16200247, Lot Number NS16200247; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270803(10)NS16189598, Lot Number NS16189598; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270803(10)NS16189586, Lot Number NS16189586; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270803(10)NS16189600, Lot Number NS16189600; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270620(10)16120672, Lot Number 16120672; Reference Part Number C-PTIS-100-FLEX-HC, UDI (01)00827002320925(17)270803(10)NS16189607, Lot Number NS16189607; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189604, Lot Number 16189604; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189588, Lot Number 16189588; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189603, Lot Number 16189603; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189608, Lot Number 16189608; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189589, Lot Number 16189589; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189602, Lot Number 16189602; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270620(10)16120666, Lot Number 16120666; Reference Part Number C-PTIS-100-HC, UDI (01)00827002121157(17)270803(10)16189594, Lot Number 16189594.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ciaglia Blue Rhino G12115Cook Incorporated · Tube tracheostomy and tube cuff1 recall2 reports
- Ciaglia Blue Rhino G32092Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98495).
- Z-1775-2026 · Open, ClassifiedCOOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
- Z-1776-2026 · Open, ClassifiedCOOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
- Z-1777-2026 · Open, ClassifiedCOOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
- Z-1778-2026 · Open, ClassifiedCOOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
- Z-1779-2026 · Open, ClassifiedCOOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA…
- Z-1780-2026 · Open, ClassifiedCOOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-…
- Z-1783-2026 · Open, ClassifiedCOOK MEDICAL Cook Spectrum Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-…
- Z-1784-2026 · Open, ClassifiedCOOK MEDICAL Cook Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
- Z-1785-2026 · Open, ClassifiedMelker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number …
- Z-1786-2026 · Open, ClassifiedCOOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
- Z-1787-2026 · Open, ClassifiedCOOK MEDICAL NCompass Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
- Z-1788-2026 · Open, ClassifiedCOOK MEDICAL Spectrum Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24…
- Z-1789-2026 · Open, ClassifiedCOOK MEDICAL Spectrum Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-R…
- Z-1790-2026 · Open, ClassifiedCOOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
- Z-1791-2026 · Open, ClassifiedCOOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
- Z-1792-2026 · Open, ClassifiedCOOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page