Blue Rhino recallZ-1779-2026 · Cook Incorporated · initiated Mar 5, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
FDA root cause category: Incorrect or no expiration date
What the company is doing
Cook Medical issued an URGENT: MEDICAL DEVICE RECALL to its consignee on 3/5 2026 via courier or email. The notice explained the issue, potential risk to health, and requested the following: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Recall details
- Recall number
- Z-1779-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 5, 2026
- Posted by the FDA
- Apr 6, 2026
- Recalling firm
- Cook Incorporated
- Quantity
- 916 units
- Product code
- JOH
- FDA submissions
- K193133
- Distribution
- Worldwide distribution.
Product description
COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687; Reference Part Number C-PTIS-100-UNS-HC-G-EU, Order Number G57686.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)270802(10)16189446, Lot Number 16189446; Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)270803(10)16189593, Lot Number 16189593; Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)270725(10)16175956, Lot Number 16175956; Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)270725(10)16175957, Lot Number 16175957; Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)270803(10)16189597, Lot Number 16189597; Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)270725(10)16175953, Lot Number 16175953; Reference Part Number C-PTIS-100-HC-G-EU, UDI (01)00827002576858(17)271001(10)16460934, Lot Number 16460934; Reference Part Number C-PTIS-100-HC-G-NA, UDI (01)00827002576827(17)280318(10)16514949, Lot Number 16514949; Reference Part Number C-PTIS-100-HC-G-NA, UDI (01)00827002576827(17)270725(10)16175948, Lot Number 16175948; Reference Part Number C-PTIS-100-UNL-HC-G-EU, UDI (01)00827002576872(17)260101(10)15151627, Lot Number 15151627; Reference Part Number C-PTIS-100-UNL-HC-G-EU, UDI (01)00827002576872(17)270627(10)16131781, Lot Number 16131781; Reference Part Number C-PTIS-100-UNL-HC-G-EU, UDI (01)00827002576872(17)280218(10)16471138, Lot Number 16471138; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)270725(10)16175951, Lot Number 16175951; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)270725(10)16175949, Lot Number 16175949; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)270724(10)16174015, Lot Number 16174015; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)270725(10)16175955, Lot Number 16175955; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)270724(10)16174023, Lot Number 16174023; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)270801(10)16187746, Lot Number 16187746; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)260327(10)15548776, Lot Number 15548776; Reference Part Number C-PTIS-100-UNS-HC-G-EU, UDI (01)00827002576865(17)280121(10)16429827, Lot Number 16429827.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Blue Rhino G57686Cook Incorporated · Tube tracheostomy and tube cuff1 recall9 reports
- Blue Rhino G57685Cook Incorporated · Tube tracheostomy and tube cuff1 recall4 reports
- Blue Rhino G57682Cook Incorporated · Tube tracheostomy and tube cuff1 recall1 report
- Blue Rhino G57687Cook Incorporated · Tube tracheostomy and tube cuff1 recall1 report
Product lines
Other devices under the same FDA submission
12 devices cleared under K193133 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Blue Rhino G57703Cook Incorporated · Tube tracheostomy and tube cuff1 recall8 reports
- Blue Rhino G57696Cook Incorporated · Tube tracheostomy and tube cuff1 recall2 reports
- Blue Rhino G57716Cook Incorporated · Tube tracheostomy and tube cuff1 recall2 reports
- Blue Rhino G57717Cook Incorporated · Tube tracheostomy and tube cuff1 recall2 reports
- Blue Rhino G57704Cook Incorporated · Tube tracheostomy and tube cuff1 recall1 report
- Blue Rhino G57721Cook Incorporated · Tube tracheostomy and tube cuff1 recall1 report
- Blue Rhino G57691Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
- Blue Rhino G57692Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
- Blue Rhino G57695Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
- Blue Rhino G57706Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
- Blue Rhino G57720Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
- Blue Rhino G57725Cook Incorporated · Tube tracheostomy and tube cuff1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98495).
- Z-1775-2026 · Open, ClassifiedCOOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
- Z-1776-2026 · Open, ClassifiedCOOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
- Z-1777-2026 · Open, ClassifiedCOOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
- Z-1778-2026 · Open, ClassifiedCOOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
- Z-1780-2026 · Open, ClassifiedCOOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-…
- Z-1781-2026 · Open, ClassifiedCOOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Orde…
- Z-1783-2026 · Open, ClassifiedCOOK MEDICAL Cook Spectrum Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-…
- Z-1784-2026 · Open, ClassifiedCOOK MEDICAL Cook Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
- Z-1785-2026 · Open, ClassifiedMelker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number …
- Z-1786-2026 · Open, ClassifiedCOOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
- Z-1787-2026 · Open, ClassifiedCOOK MEDICAL NCompass Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
- Z-1788-2026 · Open, ClassifiedCOOK MEDICAL Spectrum Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24…
- Z-1789-2026 · Open, ClassifiedCOOK MEDICAL Spectrum Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-R…
- Z-1790-2026 · Open, ClassifiedCOOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
- Z-1791-2026 · Open, ClassifiedCOOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
- Z-1792-2026 · Open, ClassifiedCOOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page