1)(2) LeadCare II Blood Lead Test Kit, (3)LeadCare II 5X… recallZ-1893-2026 · Magellan Diagnostics, Inc. · initiated Mar 13, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Magellan Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
FDA root cause category: Device Design
What the company is doing
On March 13, 2026, MEDICAL DEVICE FIELD CORRECTION letters were sent to customers. REQUIRED ACTIONS: This field action does NOT require removal of product from the field. The LeadCare Test Systems continue to perform as intended when used according to the product labeling and Instructions for Use. The regulatory status of all products noted above remains unchanged and the product supply uninterrupted. Technical and customer support will continue as normal. LEADCARE II: The only blood collection devices validated for use with the Magellan LeadCare II Blood Lead Test Kit are the capillary tubes provided in the LeadCare II Test Kit. No other collection device should be used for blood sample collection for testing with the LeadCare II System. LEADCARE PLUS AND ULTRA: End users must validate the use of micro-collection devices with LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems in accordance with their laboratory policies and procedures. REGARDING PREVIOUS RESULTS: Patients who have had elevated results when tested on the LeadCare Blood Lead Test Systems (LeadCare II, LeadCare Plus, and LeadCare Ultra) should have confirmatory testing performed in accordance with CDC or other applicable standards and guidelines.
Recall details
- Recall number
- Z-1893-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 13, 2026
- Posted by the FDA
- Apr 17, 2026
- Recalling firm
- Magellan Diagnostics, Inc.
- Quantity
- 197893
- Product code
- DOF
- FDA submissions
- K052549
- Distribution
- Worldwide - US Nationwide and the countries of Argentina, Australia, Bhutan, Bolivia, Canada, Chile, Diego Garcia, Ecuador, Eritrea, Georgia, Ghana, Guam, Guatemala, Haiti, Hong Kong, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kyrgyzstan, Mexico, Nepal, Peru, Philippines, South Africa, South Korea, Spain, UAE, Uruguay, Vietnam.
Product description
(1)(2) LeadCare II Blood Lead Test Kit, (3)LeadCare II 5X Blood Lead Test Kit, (4)LeadCare II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 70-6762RUO, (3) 70-8404 (Box containing 5 Test Kits), (4) 70-9000 (Box containing 9 test kits); The LeadCare II kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50 @) " Transfer Droppers (50 @) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Code information (models, lots, serial numbers)
Catalog Number(s): (1) 70-6762, (2) 70-6762RUO; UDI-DI: 00850355006000; ALL lots and ALL sublots; Catalog Number(s): (3) 70-8404; UDI-DI: 00850355006086; ALL lots and ALL sublots; Catalog Number(s): (4) 70-9000; UDI-DI: 00850355006093; ALL lots and ALL sublots;
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- LeadCare® II Blood Lead Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption4 recalls353 reports
- LeadCare® II 5X Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption1 recall0 reports
- LeadCare® II Blood Lead Testing System 9X Test KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption1 recall0 reports
Other devices under the same FDA submission
2 devices cleared under K052549 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LeadCare® II Blood Lead Analyzer KitMagellan Diagnostics, Inc. · Lead, Atomic Absorption137 reports
- LeadCare® II Blood Lead Analyzer Kit, RefurbishedMagellan Diagnostics, Inc. · Lead, Atomic Absorption21 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98502).
- Z-1894-2026 · Open, ClassifiedLeadCare Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatm…
- Z-1895-2026 · Open, ClassifiedLeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the following materials: " Sensors (8 containers of 24 each; 192 tests) " Tre…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page