RefleXion X1 Radiotherapy System Model RXM1000, Part… recallZ-1656-2026 · Reflexion Medical, Inc. · initiated Feb 9, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Reflexion Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

FDA root cause category: Software change control

What the company is doing

On 2/9/2026, recall notices were emailed to customers who were informed of the following: How to recognize that the device may fail: Any patient treatment with a roll offset applied in a localization followed by a subsequent repeat localization within the same treatment session will be affected. For patient treatments that have already occurred, measuring the approximate distance off-axis (R) and the maximum roll offset value (., in radians) can provide an approximate displacement value (S) through a simple mathematical formula, which may be helpful in estimating the severity of the dose impact for the patient. Customers were asked to do the following: 1) Notify your clinical staff of the recall, including posting this notification on or near the system until this issue can be resolved. 2) Halt utilization of the Repeat Localization feature for all treatment and imaging sessions. a. Note: Standard Roll corrections may still be used on the system if they are only applied in the localization immediately prior to the treatment delivery or imaging session commencing, as those are unaffected by this issue. 3) If a Roll correction is applied in an initial Localization and a Repeat Localization is initiated, the user should end the treatment session without continuing to treatment delivery or PET imaging. The user should restart the treatment session as a new treatment session to properly apply the Roll correction in the initial Localization without performing a Repeat Localization afterwards. 4) Complete and return the acknowledgement form. Questions or additional assistance can be sent to the firm at [email protected]

Recall details

Recall number
Z-1656-2026
Status
Open, Classified
Initiated by the company
Feb 9, 2026
Posted by the FDA
Mar 27, 2026
Recalling firm
Reflexion Medical, Inc.
Quantity
12
Product code
QVA
Distribution
US: CA, TX, PA, CT, NJ, OR, LA, OH

Product description

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Code information (models, lots, serial numbers)

Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.