Airway Exam Kit, DYKE1796 recallZ-1705-2026 · Medline Industries, LP · initiated Feb 16, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
Firm notified customers on 2/16/26. Consignees were provided with a copy of Olympus's notification, which includes reminders to follow the IFU. According to the firm, the component can continue to be utilized by the customer, along with the remainder of the kit.
Recall details
- Recall number
- Z-1705-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 16, 2026
- Posted by the FDA
- Apr 1, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 14,379 kits total
- Product code
- CAI
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Product description
Airway Exam Kit, DYKE1796
Code information (models, lots, serial numbers)
UDI-DI 10193489874716 (ea) 40193489874717 (case) Lots 24CBT023 24FBB701 24LBL343 25GBP141
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYKE1796Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98530).
- Z-1706-2026 · Open, ClassifiedMedline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
- Z-1707-2026 · Open, ClassifiedTHORACIC ROBOTS, DYNJ908777B
- Z-1708-2026 · Open, ClassifiedMedline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
- Z-1709-2026 · Open, ClassifiedMedline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ…
- Z-1710-2026 · Open, ClassifiedMedline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0…
- Z-1711-2026 · Open, ClassifiedMEDLINE ANTERIOR HIP PACK DYNJ64672B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.