Skip to content

Medline Anterior Hip Pack recallZ-1711-2026 · Medline Industries, LP · initiated Feb 16, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

FDA root cause category: Under Investigation by firm

What the company is doing

Firm notified customers on 2/16/26. Consignees were provided with a copy of Olympus's notification, which includes reminders to follow the IFU. According to the firm, the component can continue to be utilized by the customer, along with the remainder of the kit.

Recall details

Recall number
Z-1711-2026
Status
Open, Classified
Initiated by the company
Feb 16, 2026
Posted by the FDA
Apr 1, 2026
Recalling firm
Medline Industries, LP
Quantity
14,379 kits total
Product code
OJH
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

Product description

MEDLINE ANTERIOR HIP PACK DYNJ64672B

Code information (models, lots, serial numbers)

UDI-DI 10195327377496 (EA) 40195327377497 (CS) LOT 23LMD300

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98530).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page