Hemodialysis Bloodlines recallZ-1799-2026 · B Braun Medical Inc · initiated Mar 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
FDA root cause category: Device Design
What the company is doing
B. Braun notified consignees on about 03/19/2026 via letter titled URGENT MEDICAL DEVICE CORRECTION. Consignees were instructed to use alternative dialysis equipment and disposables, if possible. If not, BBMI recommends the following actions that can help mitigate potential harm should you need to use sets impacted by this field action on patients: 1. Ensure full adherence to both the bloodline and dialysis machine Instructions for Use (IFU). a. Streamline bloodlines - strict adherence to the priming process as outlined in the IFU, with special emphasis on Step 26. b. B3 bloodlines - strict adherence to the priming process as outlined in the IFU, with special emphasis on Step 12. 2. If possible, avoid high flow during treatment if small air bubbles are present. 3. Avoid retrograde rinseback if small air bubbles are present. 4. Be alert for a follow-up communication indicating the issue is resolved. Consignees were also instructed to review the customer notification its entirety, ensure all affected personnel are notified and aware of the issue, post the notification where the affected products are stored, and complete and return the response form. For distributors, they were instructed to forward the notification to all consignees as this recall extends to the consumer level. B. Braun will provide a follow up communication to users when lots with corrective actions are available, and users may resume high flow treatment and use of retrograde rinseback where indicated.
Recall details
- Recall number
- Z-1799-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 19, 2026
- Posted by the FDA
- Apr 15, 2026
- Recalling firm
- B Braun Medical Inc
- Quantity
- 2,653,711 units
- Product code
- FJK
- FDA submissions
- K080807
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Product description
B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096
Code information (models, lots, serial numbers)
Model Number: SL-2010M2096; UDI-DI Primary: 04046964367786; UDI-DI Unit: 04046964367779; All lots manufactured since 25MAY2025.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hemodialysis Bloodlines SL-2010M2096B. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve3 recalls3 deaths1,000 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98552).
- Z-1798-2026 · Open, ClassifiedB. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A.
- Z-1800-2026 · Open, ClassifiedB. Braun Hemodialysis Bloodlines, STREAMLINE FRESENIUS FOR DAVITA. Model Number: SL-2000M2095DA.
- Z-1801-2026 · Open, ClassifiedB. Barun Hemodialysis Bloodlines, StreamLine Bloodline Long Version, FMC. Model Number: SL-2000M2095L.
- Z-1802-2026 · Open, ClassifiedB. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705.
- Z-1803-2026 · Open, ClassifiedB Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page