Hemodialysis BloodlinesRecalls & reported problems across all models
B. Braun Medical Inc. · 9 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
14 open recalls name at least one Hemodialysis Bloodlines model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2,179
- Deaths reported
- 5
- Injuries reported
- 6
- Recalls
- 14
We found 9 models of Hemodialysis Bloodlines by B. Braun Medical Inc. in the FDA's device registry (GUDID). Together they are named in 2,179 adverse event reports in the FDA's MAUDE database, received between Dec 4, 2020 and Sep 29, 2026 — classified as 5 deaths, 6 injuries, 2,168 malfunctions.
The most frequently reported specific device problem is “Air/Gas in Device” (1,468 reports).
14 FDA recall records name at least one of these models by its device identifier, 14 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Hemodialysis Bloodlines recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 19, 2026Open, Classified
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Z-1803-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 19, 2026Open, Classified
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Z-1802-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 19, 2026Open, Classified
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Z-1801-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 19, 2026Open, Classified
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Z-1800-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 19, 2026Open, Classified
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Z-1799-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 19, 2026Open, Classified
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Z-1798-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Sep 30, 2025Open, Classified
The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.
Z-0070-2026 · Root cause: Process control · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3179-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3178-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3177-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3176-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3175-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3174-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 15, 2024Open, Classified
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Z-3173-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported Hemodialysis Bloodlines problems
Device problems
- Air/Gas in Device1,468
- Leak/Splash638
- Detachment of Device or Device Component236
- Material Puncture/Hole160
- Device Alarm System60
- Loose or Intermittent Connection42
- Contamination /Decontamination Problem32
- Crack25
Patient problems
- No Clinical Signs, Symptoms or Conditions1,782
- Hemorrhage/Blood Loss/Bleeding390
- Dyspnea2
- Insufficient Information2
- Lethargy2
- Low Blood Pressure/ Hypotension2
- Air Embolism1
- Anemia1
Hemodialysis Bloodlines models
9 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Hemodialysis Bloodlines SL-2010M2096Set, tubing, blood, with and without anti-regurgitation valve3 recalls3 deaths1,000 reports
- Hemodialysis Bloodlines SL-2010M2096ASet, tubing, blood, with and without anti-regurgitation valve2 recalls1 death483 reports
- Hemodialysis Bloodlines SL-2000M2095DASet, tubing, blood, with and without anti-regurgitation valve2 recalls1 death214 reports
- Hemodialysis Bloodlines SL-2000M2095Set, tubing, blood, with and without anti-regurgitation valve1 recall219 reports
- Hemodialysis Bloodlines SL-2000M2095DSet, tubing, blood, with and without anti-regurgitation valve1 recall158 reports
- Hemodialysis Bloodlines SL-2000M2095LSet, tubing, blood, with and without anti-regurgitation valve2 recalls97 reports
- Hemodialysis Bloodlines B3-4630M4705Set, tubing, blood, with and without anti-regurgitation valve1 recall5 reports
- Hemodialysis Bloodlines B3-3632M3705Set, tubing, blood, with and without anti-regurgitation valve1 recall2 reports
- Hemodialysis Bloodlines SL-2000M2095WSet, tubing, blood, with and without anti-regurgitation valve1 recall1 report
Recent Hemodialysis Bloodlines adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Difficult to Remove
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927919 ↗MalfunctionSep 29, 2026Difficult to Remove
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927915 ↗MalfunctionSep 29, 2026Difficult to Remove
As reported by the user facility:Event 1:Multiple incidents (at least 6) where the venous locking ring threads are slightly mismatched or not aligned correctly for connecting the inlet set to dialysis catheter. This makes it extremely difficult to screw and disconnect the venous line from the catheter lumen at end of treatment. In one case, a patient's catheter lumen was cracked/piece chipped off because the tech used a lot of force.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927908 ↗MalfunctionSep 29, 2026Difficult to Remove
As reported by the user facility:Event 4:Multiple incidents (at least 6) where the venous locking ring threads are slightly mismatched or not aligned correctly for connecting the inlet set to dialysis catheter. This makes it extremely difficult to screw and disconnect the venous line from the catheter lumen at end of treatment. In one case, a patient's catheter lumen was cracked/piece chipped off because the tech used a lot of force.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927905 ↗MalfunctionSep 29, 2026Difficult to Remove
As reported by the user facility:Event 5:Multiple incidents (at least 6) where the venous locking ring threads are slightly mismatched or not aligned correctly for connecting the inlet set to dialysis catheter. This makes it extremely difficult to screw and disconnect the venous line from the catheter lumen at end of treatment. In one case, a patient's catheter lumen was cracked/piece chipped off because the tech used a lot of force.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927901 ↗MalfunctionSep 29, 2026Difficult to Remove
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927894 ↗MalfunctionSep 15, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50777881 ↗MalfunctionSep 4, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50611603 ↗MalfunctionJun 10, 2026Air/Gas in Device
AS REPORTED BY THE USER FACILITY: BUBBLES IN THE ARTERIAL LINE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25461664 ↗MalfunctionJun 10, 2026Air/Gas in Device
AS REPORTED BY THE USER FACILITY: MACRO BUBBLES CONTINUE IN LINES. CONNECTIONS CHECKED AND TIGHTENED. BUBBLES ALSO APPEARING IN THE SALINE LINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25456973 ↗MalfunctionJun 10, 2026Air/Gas in Device
AS REPORTED BY THE USER FACILITY: MACROBUBBLES NOTED IN ARTERIAL LINE AND SALINE LINE OF TUBING DURING TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25456441 ↗MalfunctionJun 10, 2026Air/Gas in Device
AS REPORTED BY THE USER FACILITY: MACROBUBBLES LOCATED AT THE SALINE TO BLOODLINE CONNECTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25456385 ↗MalfunctionJun 8, 2026Leak/Splash
AS REPORTED BY THE USER FACILITY: ARTERIAL LINE CRACKED AND LEAKING BLOOD. WAS CRACKED RIGHT WHERE CONNECT TO FILTER.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25428867 ↗MalfunctionJun 8, 2026Leak/Splash
AS REPORTED BY THE USER FACILITY: INFORMING YOUR DEPARTMENT THAT OUR NURSES CAUGHT A LEAK WHILE THE FOLLOWING BLOODLINE WAS USED DURING A HEMODIALYSIS TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25428830 ↗MalfunctionJun 8, 2026Air/Gas in Device
AS REPORTED BY THE USER FACILITY: MACROBUBBLES PRESENT AT THE SALINE LINE BLOODLINE CONNECTOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25428786 ↗
Other B. Braun Medical Inc. products
- Infusomat54 models · 2,783 reports · 30 recalls
- IV Administration Set153 models · 281 reports · 68 recalls
- Outlook25 models · 195 reports · 24 recalls
- Perifix57 models · 224 reports · 20 recalls
- Extension Set41 models · 101 reports · 16 recalls
- Design Options116 models · 132 reports · 14 recalls
- Blood Administration Set24 models · 177 reports · 13 recalls
- Caresite23 models · 420 reports · 4 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page