MicroScan recallZ-1609-2026 · Beckman Coulter, Inc. · initiated Jul 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to the likely presence of contamination in well(s).

FDA root cause category: Material/Component Contamination

What the company is doing

On 07/24/2025, the firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter has received reports that when testing clinical isolates there was no growth observed in Ceftazidime (Caz) 1, 8 and 16 mg/mL wells; however, growth was observed in the 4 mg/mL well with multiple panels from this lot. Upon further investigation, it was confirmed the growth was Bacillus spp. and not the clinical isolates being tested. Customers are instructed to: -Consider reviewing previous Caz antimicrobial test results from this lot for any spot growth detection (e.g., isolates with no growth in in Caz 1, 8 and 16 mg/mL wells and growth in the 4 mg/mL well or if there is an unusual increase in Caz MIC values). -Discontinue use of MicroScan Neg MIC 56 Lot 2026-01-03 For any questions regarding this notice, contact Customer Technical Support: -From our website: http://www.beckmancoulter.com -By phone: call 800 677-7226 option 1 in the United States. If replacement product is needed: -Complete the attached Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR -Call Client Services at (800) 526-3821

Recall details

Recall number
Z-1609-2026
Status
Open, Classified
Initiated by the company
Jul 24, 2025
Posted by the FDA
Mar 20, 2026
Recalling firm
Beckman Coulter, Inc.
Quantity
685 boxes
Product code
LTT
FDA submissions
K202343
Distribution
U.S. Nationwide distribution in the states of CA, CO, IN, MA, MD, MN, MO, NC, NJ, OK, PA, and TX.

Product description

MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.

Code information (models, lots, serial numbers)

Catalog Number: C42464 UDI-DI code: 15099590731212 Lot Number: 2026-01-03