Skip to content

MicroScanRecalls & reported problems across all models

Beckman Coulter, Inc. · 10 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

7 open recalls name at least one MicroScan model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
1
Injuries reported
3
Recalls
9

We found 10 models of MicroScan by Beckman Coulter, Inc. in the FDA's device registry (GUDID). Together they are named in 7 adverse event reports in the FDA's MAUDE database, received between May 22, 2015 and May 1, 2026 — classified as 1 death, 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).

9 FDA recall records name at least one of these models by its device identifier, 7 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

MicroScan recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 26, 2026Open, Classified

    Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

    Z-2889-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

  2. Jun 26, 2026Open, Classified

    Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

    Z-2888-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

  3. Jul 24, 2025Open, Classified

    Due to the likely presence of contamination in well(s).

    Z-1609-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

  4. Oct 3, 2022Open, Classified

    Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.

    Z-0269-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  5. Dec 2, 2020Terminated

    Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.

    Z-0720-2021 · Root cause: Process control · Recall details · FDA record ↗

  6. Oct 21, 2019Open, Classified

    WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

    Z-1086-2022 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  7. Oct 21, 2019Open, Classified

    WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

    Z-1085-2022 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  8. Oct 21, 2019Open, Classified

    WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

    Z-1084-2022 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  9. Sep 4, 2019Terminated

    Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

    Z-1371-2021 · Root cause: Process control · Recall details · FDA record ↗

Most reported MicroScan problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Device Handling Problem1
  • Device Misassembled During Manufacturing /Shipping1
  • Manufacturing, Packaging or Shipping Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Patient Involvement2
  • Chemical Exposure1
  • Injury1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Not Applicable1
  • Septic Shock1

MicroScan models

10 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent MicroScan adverse event reports

The 7 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Beckman Coulter, Inc. products

All Beckman Coulter, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page