CELLEX recallZ-1822-2026 · Therakos Development Limited · initiated May 16, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Therakos Development Limited or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
FDA root cause category: Process control
What the company is doing
On May 28, 2025 MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Determine if your facility has any of the impacted lots listed. 1. If you have received any of the kit lot numbers listed, quarantine the product and contact our Customer Care Team at 1-833-223-4ECP (1-833-223-4327) to receive instructions on how to return the affected product. 2. To report a complaint or adverse event or if you have any other questions related to the recall, please call the Customer Care Team at 1-833-223-4ECP (1-833-223-4327). Always follow the instructions for use as detailed in the Operator s Manual when installing the THERAKOS CELLEX Photopheresis System Procedural Kits. Ensuring the correct installation of the Procedural Kits is critical for system performance and avoiding treatment interruptions. A MEDICAL DEVICE RECALL notification dated 6/23/26 was communicated to additional consignees due to the expansion of this recall to include additional lots. Consignees are to quarantine affected product and coordinate with Therakos return of devices.
Recall details
- Recall number
- Z-1822-2026
- Status
- Open, Classified
- Initiated by the company
- May 16, 2025
- Posted by the FDA
- Apr 16, 2026
- Recalling firm
- Therakos Development Limited
- Quantity
- 11,499 units
- Product code
- LNR
- FDA submissions
- P860003
- Distribution
- US Nationwide distribution.
Product description
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable
Code information (models, lots, serial numbers)
Model/Catalog Number: CLXUSA; UDI-DI: 20705030200003; Lot Numbers: N301, N339, N340, N341, N342, N343, N344, N345, N346, N347; P302, P303, P304, P305, P306, P307, P308, P309, P310, P311, P312, P313, P314, P319, P320.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CELLEX CLXUSATherakos LLC · SYSTEM, PHOTOPHERESIS, EXTRACORPOREAL2 recalls564 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under P860003 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CELLEX CLXSP-ITherakos LLC · SYSTEM, PHOTOPHERESIS, EXTRACORPOREAL23 reports
- CELLEX CLXCC-ITherakos LLC · SYSTEM, PHOTOPHERESIS, EXTRACORPOREAL4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.