NAMIC recallZ-1836-2026 · Medline Industries, LP · initiated Feb 25, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.

FDA root cause category: No Marketing Application

What the company is doing

An URGENT MEDICAL DEVICE RECALL dated 2/25/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and lot number(s) listed in the recall portal. Quarantine all affected product. 2. Complete your response form. List the quantity of affected product you have in inventory on the form. Even if you do not have any affected product, submit the form to Website link: https://recalls.medline.com Recall Reference #: R-26-020-FGX1 Recall Code: 3. When we receive your response form, we will send over-labels via FedEx overnight to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Include your customers quantities on your response form. When we receive your response form, we will send over-labels to provide to your customers. Any questions? Contact the Recall Department at 866-359-1704 or [email protected]. *** An Urgent Medical Device Recall notification letter dated 3/2/26 was sent to customers. This is a follow-up to a Medline notification you may have received on February 25, 2026. New item numbers and lot numbers have been included with this communication. Recipients receiving this notification for the first time were not affected by the original notification. REQUIRED ACTIONS: same as for 2/25/26 letter

Recall details

Recall number
Z-1836-2026
Status
Open, Classified
Initiated by the company
Feb 25, 2026
Posted by the FDA
Apr 17, 2026
Recalling firm
Medline Industries, LP
Quantity
19189 units
Product code
OEZ
Distribution
US Nationwide distribution.

Product description

Medline Kits: 1) MTS LEFT HEART KIT-CAROLINAS HOS, Model Number: 60021775; 2) DBD-MTS LHK BHS BAPTIST. LOUISVI, Model Number: 60022026; 3) MTO LEFT HEART KIT SUMMA BARBERT, Model Number: 60022673; 4) MTS LH KIT-UNIV OF MIAMI PG, Model Number: 60030817; 5) MTS LEFT HEART KIT SPRINGFIELD R, Model Number: 60031245; 6) MTS PERIPHERAL KIT SPRINGFIELD R, Model Number: 60031247; 7) MTS LEFT HEART KIT-COLISEUM MED, Model Number: 60031928; 8) MTS LEFT HEART KIT -2-CALIFORNIA, Model Number: 60033011; 9) MTS LEFT HEART KIT JOHN MUIR CON, Model Number: 60040273; 10) MTS EP MANIFOLD KIT TEXAS HLTH P, Model Number: 60041268; 11) MTS 3V MANIFOLD KIT-SPARROW HOSP, Model Number: 60050097; 12) MTS LEFT HEART KIT FORT SANDERS, Model Number: 60060387; 13) MTS LEFT HEART KIT-GOOD SAMARITA, Model Number: 60070435; 14) DBD-MTS ANGIO KIT-MOUNT HOOD MED, Model Number: 60070521; 15) MTS LFT HRT KIT-GRAND VIEW PG, Model Number: 60071743; 16) MTS CUSTOM MANIFOLD KIT-INGALLS, Model Number: 60090078; 17) MTS CATH LAB KIT - HEARTLAND PG, Model Number: 60090542; 18) MTS LEFT HEART KIT-W.C.A PG, Model Number: 60100071; 19) MTS MANIFOLD KIT-GENESYS REG MED, Model Number: 60100254; 20) MTS 4V W/PRO STA-ST.JOHNS REG ME, Model Number: 60100439; 21) MTS LEFT HEART KIT JEANES HOSPIT, Model Number: 60101773; 22) MTS LEFT HEART KIT-ST. MARYS OF, Model Number: 60131099; 23) DBD-MTS LEFT HEART KIT-LEHIGH VA, Model Number: 60132653; 24) MTS LEFT HEART KIT (3V DT WASTE), Model Number: 60138326; 25) MTS LF HRT KIT-PRESBYTERIAN UNIV, Model Number: 60160148; 26) MTS LEFT HEART KIT-PENINSULA REG, Model Number: 60160428; 27) MTS LEFT HEART KIT BARNES ST.PET, Model Number: 60161135; 28) MTO OR ENDO KIT - PARK RIDGE HOS, Model Number: 60163618; 29) MTS 3V W/DT-CORPUS CHRISTI M.C., Model Number: 60181531; 30) MTS 3V PERC MANIFOLD KIT WEST. A, Model Number: 60182137; 31) MTO CATH KIT PAMPA REGIONAL MED, Model Number: 60182661; 32) MTO HOPKINS LEFT HEART HOWARD CO, Model Number: 60183662; 33) MTS-CARDIOLOGY KIT-HARLINGEN MED, Model Number: 60184001; 34) MTS PERCEPTOR KIT METHODIST. RIC, Model Number: 60185432; 35) MTS LEFT HEART KIT SADDLEBACK ME, Model Number: 60191312; 36) MTS-CARD MAN KIT MARIAN MED CENT, Model Number: 60194273; 37) LEFT HEART KIT SOUTHERN HILLS HO, Model Number: 60195202; 38) MTS DR KHAIRALLAH LEFT HEART KIT, Model Number: 60200189; 39) MTS,4V,DIAG,KIT UNIV OF CONNECTICUT, Model Number: 60210014; 40) MTO L/H KIT-UNIV OF CLEVELAND -, Model Number: 60210047; 41) MTS LEFT HEART KIT-BAPTIST HLTH, Model Number: 60230503; 42) MTS CAN KIT ACCESSOIRE DIAGNOSTI, Model Number: 60700527; 43) MTS LEFT HEART W/SYRINGE-EMORY U, Model Number: 600503712; 44) MTS ACCESSORY KIT SAINT JOSEPH L, Model Number: 601010412; 45) MTS 3 PORT WHITE KIT LEESBURG RE, Model Number: 601222712; 46) MTS PEDS LFT RT HRT-SUNRISE HO, Model Number: 601905414; 47) MTS LHK ST. LOUIS CARDIO INST, Model Number: 606119072; 48) LHK CENTURA HLTH CORP (ST. ANTHO, Model Number: 6010030361; 49) MTS LHK CHAMPLAIN VALLEY, Model Number: 6010128411; 50) MTS LHK VANDERBILT COFFEE HARTON, Model Number: 6019705921; 51) DBD-LHK EPIC SURGERY CENTER LLC, Model Number: 6020410581; 52) DBD-MTS LHK MS HEALTHCARE MGMT, Model Number: 6061073001; 53) LHK - WESLEY MEDICAL CENTER, Model Number: 6010088152A; 54) IR ANGIOGRAPHY PACK-LF, Model Number: DYNJVB1077D; 55) RADIOLOGY SPECIALS TRAY, Model Number: DYNJVB1302B; 56) ANGIO LEFT HEART PACK, Model Number: DYNJ40632D; 57) ANGIO RIGHT HEART PACK, Model Number: DYNJ40633C; 58) ANGIO TRAY, Model Number: DYNJ43278C; 59) CARDIAC LEG PACK, Model Number: DYNJ44065C; 60) EP PACK, Model Number: DYNJ55512A; 61) VASCULAR PACK, Model Number: DYNJ58522A; 62) PERIPHERAL CATH, Model Number: DYNJ59220C; 63) CARDIAC CATH - NO SYR, Model Number: DYNJ62409A; 64) C V MINOR PACK, Model Number: DYNJ64862B; 65) ENDOVASCULAR PACK, Model Number: DYNJ65926

Code information (models, lots, serial numbers)

Show the full code information (13k characters)

1) 60021775, UDI-DI: 10193489041385(each), 20193489041382(case), Lot Number: 183372; 2) 60021775, UDI-DI: 10193489041385(each), 20193489041382(case), Lot Number: 185190; 3) 60022026, UDI-DI: 10193489087253(each), 20193489087250(case), Lot Number: 106008; 4) 60022026, UDI-DI: 10193489087253(each), 20193489087250(case), Lot Number: 117378; 5) 60022673, UDI-DI: 10193489041538(each), 20193489041535(case), Lot Number: 176180; 6) 60030817, UDI-DI: 10193489041972(each), 20193489041979(case), Lot Number: 185193; 7) 60031245, UDI-DI: 10193489042214(each), 20193489042211(case), Lot Number: 183374; 8) 60031247, UDI-DI: 10193489042238(each), 20193489042235(case), Lot Number: 174370; 9) 60031247, UDI-DI: 10193489042238(each), 20193489042235(case), Lot Number: 184424; 10) 60031247, UDI-DI: 10193489042238(each), 20193489042235(case), Lot Number: 185080; 11) 60031928, UDI-DI: 10193489042276(each), 20193489042273(case), Lot Number: 185195; 12) 60033011, UDI-DI: 10193489042429(each), 20193489042426(case), Lot Number: 181365; 13) 60040273, UDI-DI: 10193489042832(each), 20193489042839(case), Lot Number: 179872; 14) 60040273, UDI-DI: 10193489042832(each), 20193489042839(case), Lot Number: 183375; 15) 60041268, UDI-DI: 10193489043051(each), 20193489043058(case), Lot Number: 140365; 16) 60041268, UDI-DI: 10193489043051(each), 20193489043058(case), Lot Number: 170084; 17) 60050097, UDI-DI: 10193489043242(each), 20193489043249(case), Lot Number: 176186; 18) 60060387, UDI-DI: 10193489043693(each), 20193489043690(case), Lot Number: 183376; 19) 60060387, UDI-DI: 10193489043693(each), 20193489043690(case), Lot Number: 185194; 20) 60060387, UDI-DI: 10193489043693(each), 20193489043690(case), Lot Number: 187370; 21) 60070435, UDI-DI: 10193489044133(each), 20193489044130(case), Lot Number: 183377; 22) 60070435, UDI-DI: 10193489044133(each), 20193489044130(case), Lot Number: 187371; 23) 60070521, UDI-DI: 10193489044164(each), 20193489044161(case), Lot Number: 185196; 24) 60071743, UDI-DI: 10193489044348(each), 20193489044345(case), Lot Number: 176190; 25) 60090078, UDI-DI: 10193489044829(each), 20193489044826(case), Lot Number: 185197; 26) 60090542, UDI-DI: 10193489044867(each), 20193489044864(case), Lot Number: 181373; 27) 60090542, UDI-DI: 10193489044867(each), 20193489044864(case), Lot Number: 185198; 28) 60100071, UDI-DI: 10193489044959(each), 20193489044956(case), Lot Number: 176193; 29) 60100071, UDI-DI: 10193489044959(each), 20193489044956(case), Lot Number: 185189; 30) 60100254, UDI-DI: 10193489045024(each), 20193489045021(case), Lot Number: 185188; 31) 60100439, UDI-DI: 10193489045130(each), 20193489045137(case), Lot Number: 181369; 32) 60100439, UDI-DI: 10193489045130(each), 20193489045137(case), Lot Number: 185203; 33) 60101773, UDI-DI: 10193489045451(each), 20193489045458(case), Lot Number: 181364; 34) 60131099, UDI-DI: 10193489046717(each), 20193489046714(case), Lot Number: 181370; 35) 60131099, UDI-DI: 10193489046717(each), 20193489046714(case), Lot Number: 185204; 36) 60132653, UDI-DI: 10193489047332(each), 20193489047339(case), Lot Number: 181381; 37) 60138326, UDI-DI: 10193489047653(each), 20193489047650(case), Lot Number: 185205; 38) 60160148, UDI-DI: 10193489048704(each), 20193489048701(case), Lot Number: 183382; 39) 60160148, UDI-DI: 10193489048704(each), 20193489048701(case), Lot Number: 185207; 40) 60160428, UDI-DI: 10193489048773(each), 20193489048770(case), Lot Number: 181100; 41) 60161135, UDI-DI: 10193489049046(each), 20193489049043(case), Lot Number: 177936; 42) 60161135, UDI-DI: 10193489049046(each), 20193489049043(case), Lot Number: 185208; 43) 60163618, UDI-DI: 10193489049251(each), 20193489049258(case), Lot Number: 140392; 44) 60181531, UDI-DI: 10193489049619(each), 20193489049616(case), Lot Number: 185209; 45) 60182137, UDI-DI: 10193489049695(each), 20193489049692(case), Lot Number: 185522; 46) 60182137, UDI-DI: 10193489049695(each), 20193489049692(case), Lot Number: 187343; 47) 60182661, UDI-DI: 10193489049794(each), 20193489049791(case), Lot Number: 176203; 48) 60183662, UDI-DI: 10193489050080(each), 20193489050087(case), Lot Number: 181371; 49) 60184001, UDI-DI: 10193489050110(each), 20193489050117(case), Lot Number: 177939; 50) 60184001, UDI-DI: 10193489050110(each), 20193489050117(case), Lot Number: 185210; 51) 60185432, UDI-DI: 10193489050325(each), 20193489050322(case), Lot Number: 184739; 52) 60191312, UDI-DI: 10193489050820(each), 20193489050827(case), Lot Number: 185212; 53) 60194273, UDI-DI: 10193489051094(each), 20193489051091(case), Lot Number: 133064; 54) 60195202, UDI-DI: 10193489051322(each), 20193489051329(case), Lot Number: 181368; 55) 60195202, UDI-DI: 10193489051322(each), 20193489051329(case), Lot Number: 187374; 56) 60200189, UDI-DI: 10193489051513(each), 20193489051510(case), Lot Number: 176208; 57) 60200189, UDI-DI: 10193489051513(each), 20193489051510(case), Lot Number: 181375; 58) 60200189, UDI-DI: 10193489051513(each), 20193489051510(case), Lot Number: 185213; 59) 60210014, UDI-DI: 10193489051766(each), 20193489051763(case), Lot Number: 187375; 60) 60210047, UDI-DI: 10193489051797(each), 20193489051794(case), Lot Number: 181378; 61) 60230503, UDI-DI: 10193489053098(each), 20193489053095(case), Lot Number: 185214; 62) 60700527, UDI-DI: 10193489057171(each), 20193489057178(case), Lot Number: 185215; 63) 600503712, UDI-DI: 10193489070095(each), 20193489070092(case), Lot Number: 185217; 64) 601010412, UDI-DI: 10193489070613(each), 20193489070610(case), Lot Number: 185220; 65) 601222712, UDI-DI: 10193489070910(each), 20193489070917(case), Lot Number: 176212; 66) 601905414, UDI-DI: 10193489072105(each), 20193489072102(case), Lot Number: 181379; 67) 601905414, UDI-DI: 10193489072105(each), 20193489072102(case), Lot Number: 185221; 68) 606119072, UDI-DI: 10193489073461(each), 20193489073468(case), Lot Number: 106750; 69) 6010030361, UDI-DI: 10193489108033(each), 20193489108030(case), Lot Number: 184480; 70) 6010128411, UDI-DI: 10193489093728(each), 20193489093725(case), Lot Number: 181376; 71) 6010128411, UDI-DI: 10193489093728(each), 20193489093725(case), Lot Number: 185223; 72) 6019705921, UDI-DI: 10193489096156(each), 20193489096153(case), Lot Number: 185352; 73) 6019705921, UDI-DI: 10193489096156(each), 20193489096153(case), Lot Number: 185354; 74) 6020410581, UDI-DI: 10193489131598(each), 20193489131595(case), Lot Number: 155372; 75) 6061073001, UDI-DI: 10193489095692(each), 20193489095699(case), Lot Number: 167702; 76) 6010088152A, UDI-DI: 10193489141276(each), 20193489141273(case), Lot Number: 185224; 77) DYNJVB1077D, UDI-DI: 10198459589270(each), 40198459589271(case), Lot Number: 25KLA735; 78) DYNJVB1302B, UDI-DI: 10195327625429(each), 40195327625420(case), Lot Number: 25LLA560; 79) DYNJVB1302B, UDI-DI: 10195327625429(each), 40195327625420(case), Lot Number: 25KLA369; 80) DYNJ40632D, UDI-DI: 10198459198212(each), 40198459198213(case), Lot Number: 25JML198; 81) DYNJ40633C, UDI-DI: 10198459198229(each), 40198459198220(case), Lot Number: 24KMG184; 82) DYNJ40633C, UDI-DI: 10198459198229(each), 40198459198220(case), Lot Number: 25EMF185; 83) DYNJ43278C, UDI-DI: 10193489957358(each), 40193489957359(case), Lot Number: 25DBU258; 84) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 22ABO333; 85) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 22DBN254; 86) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 22IBJ531; 87) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23ABN017; 88) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23BBI303; 89) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23DBG355; 90) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23FBG335; 91) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23GBR632; 92) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23IBB714; 93) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 23JBU288; 94) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24ABA045; 95) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24BBH849; 96) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24EBA787; 97) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24FBN083; 98) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24GBX937; 99) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24IBU804; 100) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24KBU070; 101) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 25ABB637; 102) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 25BBT277; 103) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 25DBL306; 104) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 22HBI765; 105) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 24BBE157; 106) DYNJ44065C, UDI-DI: 10195327080631(each), 40195327080632(case), Lot Number: 25EBU657; 107) DYNJ55512A, UDI-DI: 10889942765003(each), 40889942765004(case), Lot Number: 22FBI713; 108) DYNJ55512A, UDI-DI: 10889942765003(each), 40889942765004(case), Lot Number: 22FBR989; 109) DYNJ55512A, UDI-DI: 10889942765003(each), 40889942765004(case), Lot Number: 23DBR938; 110) DYNJ55512A, UDI-DI: 10889942765003(each), 40889942765004(case), Lot Number: 23DBR937; 111) DYNJ58522A, UDI-DI: 10198459135781(each), 40198459135782(case), Lot Number: 25CBC459; 112) DYNJ59220C, UDI-DI: 10198459484551(each), 40198459484552(case), Lot Number: 25JLA710; 113) DYNJ62409A, UDI-DI: 10193489218671(each), 40193489218672(case), Lot Number: 25CBD234; 114) DYNJ64862B, UDI-DI: 10195327053703(each), 40195327053704(case), Lot Number: 24IBJ708; 115) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 21HBB198; 116) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 21IBK707; 117) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 21LBH265; 118) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 22ABZ745; 119) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 22EBO031; 120) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 22FBK919; 121) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 23ABG784; 122) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 23BBM271; 123) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 23DBJ192; 124) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 23EBF015; 125) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 23GBC290; 126) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24BBH874; 127) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24CBD055; 128) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24DBO541; 129) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24FBQ427; 130) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24IBL349; 131) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24KBE756; 132) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 24KBR816; 133) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 25DBI909; 134) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 21HBO230; 135) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 23EBS139; 136) DYNJ65926, UDI-DI: 10193489305456(each), 40193489305457(case), Lot Number: 25GBJ642; 137) DYNJ82863C, UDI-DI: 10198459103315(each), 40198459103316(case), Lot Number: 25CBP219; 138) DYNJ83913D, UDI-DI: 10198459515378(each), 40198459515379(case), Lot Number: 25LMI880; 139) DYNJ85073B, UDI-DI: 10198459167164(each), 40198459167165(case), Lot Number: 25EMF051; 140) DYNJ88292, UDI-DI: 10198459142079(each), 40198459142070(case), Lot Number: 24JBQ452; 141) DYNJ88292, UDI-DI: 10198459142079(each), 40198459142070(case), Lot Number: 25EBR618; 142) DYNJ89865, UDI-DI: 10198459296765(each), 40198459296766(case), Lot Number: 25CBP998; 143) DYNJ906951D, UDI-DI: 10198459132490(each), 40198459132491(case), Lot Number: 25JBR478; 144) DYNJ908389B, UDI-DI: 10198459122903(each), 40198459122904(case), Lot Number: 25GBI395; 145) VASC1068, UDI-DI: 10889942527908(each), 40889942527909(case), Lot Number: 21ABE418; 146) VASC1068, UDI-DI: 10889942527908(each), 40889942527909(case), Lot Number: 21ABA917; 147) VASC1127, UDI-DI: 10889942834273(each), 40889942834274(case), Lot Number: 25HBQ004; 148) VASC1324, UDI-DI: 10195327126773(each), 40195327126774(case), Lot Number: 25FBD245; 149) VASC1509, UDI-DI: 10195327686581(each), 40195327686582(case), Lot Number: 24IBE174; 150) VASC1824, UDI-DI: 10198459435805(each), 40198459435806(case), Lot Number: 25GBT688; 151) 00-399024AC, UDI-DI: 10198459162442(each), 40198459162443(case), Lot Number: 25KMB256; 152) 00-399024AC, UDI-DI: 10198459162442(each), 40198459162443(case), Lot Number: 25HMD974

Devices named in this recall

74 devices in the FDA device registry whose identifier appears in the recall's code information.