Medline Kits recallZ-1842-2026 · Medline Industries, LP · initiated Feb 25, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.

FDA root cause category: No Marketing Application

What the company is doing

An URGENT MEDICAL DEVICE RECALL dated 2/25/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and lot number(s) listed in the recall portal. Quarantine all affected product. 2. Complete your response form. List the quantity of affected product you have in inventory on the form. Even if you do not have any affected product, submit the form to Website link: https://recalls.medline.com Recall Reference #: R-26-020-FGX1 Recall Code: 3. When we receive your response form, we will send over-labels via FedEx overnight to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Include your customers quantities on your response form. When we receive your response form, we will send over-labels to provide to your customers. Any questions? Contact the Recall Department at 866-359-1704 or [email protected]. *** An Urgent Medical Device Recall notification letter dated 3/2/26 was sent to customers. This is a follow-up to a Medline notification you may have received on February 25, 2026. New item numbers and lot numbers have been included with this communication. Recipients receiving this notification for the first time were not affected by the original notification. REQUIRED ACTIONS: same as for 2/25/26 letter

Recall details

Recall number
Z-1842-2026
Status
Open, Classified
Initiated by the company
Feb 25, 2026
Posted by the FDA
Apr 17, 2026
Recalling firm
Medline Industries, LP
Quantity
38014 units
Product code
LRO
Distribution
US Nationwide distribution.

Product description

Medline Kits: 1) NEWBORN ADMISSION, Model Number: DYKB1038; 2) DRAINAGE TRAY, Model Number: DYNDA3188; 3) VA ORLANDO SHOULDER PACK, Model Number: DYNJG001027B; 4) TUMESCENT SYRINGE KIT, Model Number: DYNJTUMSYR; 5) OMC PTCA PACK, Model Number: DYNJT3719; 6) PORT PACK, Model Number: DYNJT5667; 7) WASTE MANAGEMENT KIT - 2, Model Number: DYNJWASTEKIT2; 8) CARDIAC CATH, Model Number: DYNJ20175J; 9) D&C PACK-POB, Model Number: DYNJ33521D; 10) BASIC PACK, Model Number: DYNJ38533F; 11) CATH PACK IMPLANT, Model Number: DYNJ39230D; 12) CERVICAL SPINE PACK, Model Number: DYNJ40886B; 13) MV-IR ANGIO DRAPE PACK-LF, Model Number: DYNJ41553F; 14) TV ENDOVASCULAR-LF, Model Number: DYNJ41558D; 15) MAJOR PACK-LF, Model Number: DYNJ44816J; 16) ANGIOGRAPHY PACK, Model Number: DYNJ48683A; 17) EXTREMITY PACK, Model Number: DYNJ55922C; 18) NEURO PACK, Model Number: DYNJ57785C; 19) PACEMAKER PACK, Model Number: DYNJ61377B; 20) VASCULAR PACK, Model Number: DYNJ62683D; 21) LINQ INSERTION PK W/OUT LIDO, Model Number: DYNJ63668B; 22) BASIC PACK, Model Number: DYNJ83324; 23) OMF MINOR PACK, Model Number: DYNJ83991B; 24) ARTERIOGRAM SUMMIT PACK, Model Number: DYNJ84040B; 25) HMWB NEURO NAMIC PACK, Model Number: DYNJ86600; 26) VCH CARDIAC CATH PACK, Model Number: DYNJ89064; 27) RADIOLOGY PACK, Model Number: DYNJ89080; 28) LITHOTOMY PACK, Model Number: DYNJ89959; 29) AHT SPINE PACK, Model Number: DYNJ903566D; 30) LAP, Model Number: DYNJ904576B; 31) EP MERCY JANESVILLE, Model Number: DYNJ907013; 32) LHK, US, RUBY MEMORIAL HOSP, Model Number: NAM0032; 33) CHERIAN CATH KIT, Model Number: VASC1086; 34) RIGHT HEART KIT - LAKE NONA, Model Number: VASC1120; 35) KIT ANGIO CSTM-SHAWNEE, Model Number: VASC1151; 36) LEFT HEART KIT - OAKBEND, Model Number: VASC1155; 37) IR 2 PORT-ARROWHEAD REGIONAL, Model Number: VASC1404; 38) LHK - ARROWHEAD REGIONAL MED, Model Number: VASC1405; 39) LHK - UMC DOUBLE TRANSDUCER, Model Number: VASC1415A; 40) SDMC, EP TWO PORT CUSTOM KIT, Model Number: VASC1533; 41) ARTERIOGRAM TRAY, Model Number: 00-402001V

Code information (models, lots, serial numbers)

Show the full code information (17k characters)

1) DYKB1038, UDI-DI: 10193489502428(each), 40193489502429(case), Lot Number: 22GBJ869; 2) DYNDA3188, UDI-DI: 10198459303579(each), 40198459303570(case), Lot Number: 25DBF256; 3) DYNJG001027B, UDI-DI: 10198459065453(each), 40198459065454(case), Lot Number: 25ELA998; 4) DYNJG001027B, UDI-DI: 10198459065453(each), 40198459065454(case), Lot Number: 25ELA723; 5) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21BBH579; 6) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21BBS213; 7) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21DBS220; 8) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21HBK178; 9) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21IBV194; 10) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21JBJ063; 11) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21JBM583; 12) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22ABG678; 13) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22ABZ724; 14) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22DBL342; 15) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22IBS978; 16) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22JBV884; 17) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22LBH141; 18) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22OBE023; 19) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22OBE024; 20) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 22OBE025; 21) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23CBP389; 22) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23EBB429; 23) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23EBQ437; 24) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23GBB278; 25) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23HBD985; 26) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23JBN242; 27) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23LBP034; 28) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23LBP432; 29) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24ABA416; 30) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24EBF244; 31) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24FBB647; 32) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24FBO960; 33) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24GBE249; 34) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24GBW947; 35) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24HBN439; 36) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24JBC123; 37) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24KBI920; 38) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24LBO860; 39) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 21JBV362; 40) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 23BBU890; 41) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 24ABN982; 42) DYNJTUMSYR, UDI-DI: 10884389882054(each), 40884389882055(case), Lot Number: 25HBM032; 43) DYNJT3719, UDI-DI: 10198459347054(each), 40198459347055(case), Lot Number: 25EBU996; 44) DYNJT5667, UDI-DI: 10198459492129(each), 40198459492120(case), Lot Number: 25HDB174; 45) DYNJWASTEKIT2, UDI-DI: 10193489548204(each), 40193489548205(case), Lot Number: 22HBK640; 46) DYNJ20175J, UDI-DI: 10198459023750(each), 40198459023751(case), Lot Number: 25DMI117; 47) DYNJ20175J, UDI-DI: 10198459023750(each), 40198459023751(case), Lot Number: 25BMC155; 48) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 24FDB534; 49) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 24HDC006; 50) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 24IDB770; 51) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 24KDA884; 52) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 24LDA938; 53) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 25ADB529; 54) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 25BDB719; 55) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 25DDB454; 56) DYNJ33521D, UDI-DI: 10198459015656(each), 40198459015657(case), Lot Number: 25HDA895; 57) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24DBD251; 58) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24EBF801; 59) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24FBD925; 60) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24GBY008; 61) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24HBN127; 62) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24JBT025; 63) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24KBF095; 64) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24LBK225; 65) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 25BBA827; 66) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 25CBF946; 67) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 25DBH070; 68) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 25EBK903; 69) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 24GBC776; 70) DYNJ38533F, UDI-DI: 10195327609931(each), 40195327609932(case), Lot Number: 25GBS410; 71) DYNJ39230D, UDI-DI: 10198459223679(each), 40198459223670(case), Lot Number: 25BBB333; 72) DYNJ39230D, UDI-DI: 10198459223679(each), 40198459223670(case), Lot Number: 25CBI608; 73) DYNJ40886B, UDI-DI: 10193489768282(each), 40193489768283(case), Lot Number: 25ELA632; 74) DYNJ41553F, UDI-DI: 10198459079634(each), 40198459079635(case), Lot Number: 25EMI492; 75) DYNJ41558D, UDI-DI: 10889942811663(each), 40889942811664(case), Lot Number: 25EMC747; 76) DYNJ44816J, UDI-DI: 10198459087318(each), 40198459087319(case), Lot Number: 24HBE996; 77) DYNJ44816J, UDI-DI: 10198459087318(each), 40198459087319(case), Lot Number: 24IBD804; 78) DYNJ44816J, UDI-DI: 10198459087318(each), 40198459087319(case), Lot Number: 24JBH568; 79) DYNJ44816J, UDI-DI: 10198459087318(each), 40198459087319(case), Lot Number: 25EBR969; 80) DYNJ48683A, UDI-DI: 10889942491377(each), 40889942491378(case), Lot Number: 24LBO769; 81) DYNJ55922C, UDI-DI: 10198459368318(each), 40198459368319(case), Lot Number: 25FMA909; 82) DYNJ55922C, UDI-DI: 10198459368318(each), 40198459368319(case), Lot Number: 25EMJ828; 83) DYNJ57785C, UDI-DI: 10195327543624(each), 40195327543625(case), Lot Number: 24BBN969; 84) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24CBB176; 85) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24DBR422; 86) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24FBO638; 87) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24HBT927; 88) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24IBP969; 89) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24JBH473; 90) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 24KBG036; 91) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 25CBH396; 92) DYNJ61377B, UDI-DI: 10195327592554(each), 40195327592555(case), Lot Number: 25EBE869; 93) DYNJ62683D, UDI-DI: 10195327556518(each), 40195327556519(case), Lot Number: 24ABP681; 94) DYNJ63668B, UDI-DI: 10195327550851(each), 40195327550852(case), Lot Number: 24HME110; 95) DYNJ83324, UDI-DI: 10195327260781(each), 40195327260782(case), Lot Number: 23BMD400; 96) DYNJ83324, UDI-DI: 10195327260781(each), 40195327260782(case), Lot Number: 23AMA569; 97) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 24HBS684; 98) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 24JBP876; 99) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 24JBS147; 100) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 24LBL583; 101) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 25CBP938; 102) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 25DBQ919; 103) DYNJ83991B, UDI-DI: 10198459087325(each), 40198459087326(case), Lot Number: 25FBD407; 104) DYNJ84040B, UDI-DI: 10198459254161(each), 40198459254162(case), Lot Number: 25BMG397; 105) DYNJ86600, UDI-DI: 10198459021565(each), 40198459021566(case), Lot Number: 25BBT025; 106) DYNJ89064, UDI-DI: 10198459230820(each), 40198459230821(case), Lot Number: 25ABL031; 107) DYNJ89080, UDI-DI: 10198459232749(each), 40198459232740(case), Lot Number: 25ABS218; 108) DYNJ89080, UDI-DI: 10198459232749(each), 40198459232740(case), Lot Number: 25BBC477; 109) DYNJ89080, UDI-DI: 10198459232749(each), 40198459232740(case), Lot Number: 25CBC491; 110) DYNJ89080, UDI-DI: 10198459232749(each), 40198459232740(case), Lot Number: 25DBC830; 111) DYNJ89959, UDI-DI: 10198459300912(each), 40198459300913(case), Lot Number: 25DBD397; 112) DYNJ903566D, UDI-DI: 10195327560829(each), 40195327560820(case), Lot Number: 25HBU069; 113) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 21FBN722; 114) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 21HBI274; 115) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 21IBL093; 116) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 21KBE009; 117) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 21LBG338; 118) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 22GBZ027; 119) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 22JMB507; 120) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 22KMC060; 121) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23BMF719; 122) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23CME266; 123) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23FMI182; 124) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23GMI509; 125) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23JME302; 126) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23KMI130; 127) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23LMH540; 128) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 24BMA155; 129) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 24GME966; 130) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 24IMC690; 131) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 24KMG557; 132) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 21GBO411; 133) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 23FMC753; 134) DYNJ904576B, UDI-DI: 10888277776760(each), 40888277776761(case), Lot Number: 25GMD612; 135) DYNJ907013, UDI-DI: 10193489318630(each), 40193489318631(case), Lot Number: 25FLA929; 136) NAM0032, UDI-DI: 10195327694357(each), 40195327694358(case), Lot Number: 24LBG464; 137) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 21DBF029; 138) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 21GBV041; 139) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 21IBQ150; 140) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 21KBK856; 141) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 21LBA326; 142) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 22CBX005; 143) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 22EBO968; 144) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 22HBF991; 145) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 22IBR901; 146) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23ABL896; 147) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23ABN264; 148) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23DBA069; 149) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23DBM493; 150) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23EBH545; 151) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23FBF982; 152) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23HBE007; 153) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23IBE867; 154) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23JBW353; 155) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24ABL399; 156) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24CBB750; 157) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24CBP944; 158) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24EBQ247; 159) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24FBB644; 160) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24HBV207; 161) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 24KBI912; 162) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 25CBQ305; 163) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 25DBK904; 164) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 22JBD604; 165) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 23KBG280; 166) VASC1086, UDI-DI: 10889942668144(each), 40889942668145(case), Lot Number: 25EBH405; 167) VASC1120, UDI-DI: 10889942816316(each), 40889942816317(case), Lot Number: 25EBV199; 168) VASC1120, UDI-DI: 10889942816316(each), 40889942816317(case), Lot Number: 24FBD352; 169) VASC1151, UDI-DI: 10889942937523(each), 40889942937524(case), Lot Number: 25EBM012; 170) VASC1151, UDI-DI: 10889942937523(each), 40889942937524(case), Lot Number: 25EBD891; 171) VASC1155, UDI-DI: 10888277760493(each), 40888277760494(case), Lot Number: 25GBN921; 172) VASC1404, UDI-DI: 10195327366568(each), 40195327366569(case), Lot Number: 23EBI565; 173) VASC1405, UDI-DI: 10195327367015(each), 40195327367016(case), Lot Number: 25GBS655; 174) VASC1405, UDI-DI: 10195327367015(each), 40195327367016(case), Lot Number: 25DBJ392; 175) VASC1415A, UDI-DI: 10195327678272(each), 40195327678273(case), Lot Number: 24LBH191; 176) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 24GBM956; 177) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 24HBN734; 178) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 24JBS224; 179) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 24LBL061; 180) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 25BBQ097; 181) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 25CBK254; 182) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 25CBN756; 183) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 25CBU919; 184) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 25DBG067; 185) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 24FBQ508; 186) VASC1533, UDI-DI: 10198459035623(each), 40198459035624(case), Lot Number: 25EBQ528; 187) 00-402001V, UDI-DI: 10198459235269(each), 40198459235260(case), Lot Number: 25ABG557; 188) 00-402001V, UDI-DI: 10198459235269(each), 40198459235260(case), Lot Number: 25BBE970; 189) 00-402001V, UDI-DI: 10198459235269(each), 40198459235260(case), Lot Number: 25CBN926; 190) 00-402001V, UDI-DI: 10198459235269(each), 40198459235260(case), Lot Number: 25EBD516; 191) 00-402001V, UDI-DI: 10198459235269(each), 40198459235260(case), Lot Number: 25EBP919

Devices named in this recall

37 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98599).