i-STAT EG7+ cartridge recallZ-1999-2026 · Abbott Point Of Care Inc. · initiated Apr 1, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

FDA root cause category: Process change control

What the company is doing

On April 1, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. This Urgent Medical Device Corrective is to customers in the U.S. with the following recommended actions: " Discontinue use of the affected i-STAT EG7+, EG6+, and G3+ cartridge lots for reporting PCO and pH results. " Use an alternate cartridge type, alternate lot, or testing method. " Complete and return the Business Reply Card included with this notification to confirm receipt and understanding. If you have forwarded any of the specified i-STAT EG7+, EG6+ or G3+ cartridges to another facility, we request that you please provide a copy of this letter to them.

Recall details

Recall number
Z-1999-2026
Status
Open, Classified
Initiated by the company
Apr 1, 2026
Posted by the FDA
Apr 29, 2026
Recalling firm
Abbott Point Of Care Inc.
Quantity
369650 units (150875 US, 218775 OUS)
Product code
CHL
FDA submissions
K940918
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.

Product description

i-STAT EG7+ cartridge; List Number: 03P76-25;

Code information (models, lots, serial numbers)

List Number: 03P76-25; UDI-DI: 00054749000043; Lot numbers: N25206, N25207, N25209, N25213, N25218, N25219, N25233;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

4 devices cleared under K940918 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98635).