ARTIS pheno and ARTIS Icono biplane, floor and ceiling.… recallZ-1653-2026 · Siemens Medical Solutions USA, Inc · initiated Mar 10, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice, dated March 9, 2026, to all affected customers.

Recall details

Recall number
Z-1653-2026
Status
Open, Classified
Initiated by the company
Mar 10, 2026
Posted by the FDA
Apr 7, 2026
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
U.S. 758, OUS 2077
Product code
OWB
FDA submissions
K241572
Distribution
Worldwide - US Nationwide distribution.

Product description

ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

Code information (models, lots, serial numbers)

ARTIS pheno UDI: 4056869046877. ARTIS icono biplane UDI: 4056869063317. ARTIS icono ceiling UDI: 4056869295923. ARTIS icono floor UDI: 4056869149325.

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.