ARTIS icono ceiling 11328100Recalls & reported problems
Siemens Healthcare GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
9 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 9
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056869295923), received between Apr 15, 2026 and Apr 15, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Activation Failure” (1 report).
9 FDA recall records name this device's identifier, 9 still open.
7 other recalls cover devices cleared under the same FDA submission (K221516, K241572, K254173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jun 8, 2026Open, Classified
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
Z-2648-2026 · Root cause: Device Design · Recall details · FDA record ↗
- May 20, 2026Open, Classified
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
Z-2642-2026 · Root cause: Software design · Recall details · FDA record ↗
- Mar 10, 2026Open, Classified
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Z-1653-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗
- Mar 9, 2026Open, Classified
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Z-2206-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗
- May 15, 2025Open, Classified
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
Z-2021-2025 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗
- Jun 26, 2024Open, Classified
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
Z-2240-2024 · Root cause: Software design · Recall details · FDA record ↗
- Jun 26, 2024Open, Classified
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
Z-2241-2024 · Root cause: Software design · Recall details · FDA record ↗
- Jun 26, 2024Open, Classified
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
Z-2242-2024 · Root cause: Software design · Recall details · FDA record ↗
- Oct 17, 2023Open, Classified
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
Z-0523-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure1
- Computer Software Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 15, 2026Computer Software Problem; Activation Failure
SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WITH THE ARTIS ICONO CEILING SYSTEM. DURING SYSTEM BOOTUP, IT WAS DETERMINED THAT STAND MOVEMENTS WERE NOT POSSIBLE, AND A STAND/TABLE ERROR MESSAGE WAS DISPLAYED TO THE USER. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24885278 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221516, K241572, K254173. One submission often covers many models, so check the product and lot details in each record.
- Z-2641-2026 · May 20, 2026 · Open, ClassifiedARTIS icono biplane. Model Number: 11327600.
- Z-2643-2026 · May 20, 2026 · Open, ClassifiedARTIS icono floor. Model Number: 11327700.
- Z-1814-2026 · Mar 12, 2026 · Open, ClassifiedARTIS Pheno VE30A and VE40A, Model 10849000
- Z-2205-2026 · Mar 9, 2026 · Open, ClassifiedARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
- Z-2529-2025 · Aug 12, 2025 · Open, ClassifiedArtis Pheno. Image-Intensified Flouroscopic X-Ray System.
- Z-2020-2025 · May 15, 2025 · Open, Classifiedinterventional fluoroscopic x-ray system
- Z-2022-2025 · May 15, 2025 · Open, Classifiedinterventional fluoroscopic x-ray system
Other System, x-ray, angiographic devices
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis zee biplane 10094141Siemens Healthcare GmbH · System, x-ray, angiographic3 recalls63 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis Q.zen floor 10848353Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls5 reports
- Artis zee floor MN 10094142Siemens Healthcare GmbH · System, x-ray, angiographic2 recalls0 reports
- Allia MoveoGe Medical Systems · System, X-Ray, Angiographic1 recall0 reports
- iMRX System iMRX System AR-DR MR30RSIMRIS Imaging, Inc. · System, X-Ray, Angiographic1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page