ARTIS icono ceiling 11328100Recalls & reported problems

Siemens Healthcare GmbH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

9 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
9

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056869295923), received between Apr 15, 2026 and Apr 15, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

9 FDA recall records name this device's identifier, 9 still open.

7 other recalls cover devices cleared under the same FDA submission (K221516, K241572, K254173); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 8, 2026Open, Classified

    Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.

    Z-2648-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. May 20, 2026Open, Classified

    Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.

    Z-2642-2026 · Root cause: Software design · Recall details · FDA record ↗

  3. Mar 10, 2026Open, Classified

    During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.

    Z-1653-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  4. Mar 9, 2026Open, Classified

    During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

    Z-2206-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  5. May 15, 2025Open, Classified

    In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.

    Z-2021-2025 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  6. Jun 26, 2024Open, Classified

    A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

    Z-2240-2024 · Root cause: Software design · Recall details · FDA record ↗

  7. Jun 26, 2024Open, Classified

    A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

    Z-2241-2024 · Root cause: Software design · Recall details · FDA record ↗

  8. Jun 26, 2024Open, Classified

    A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

    Z-2242-2024 · Root cause: Software design · Recall details · FDA record ↗

  9. Oct 17, 2023Open, Classified

    Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.

    Z-0523-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure1
  • Computer Software Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 15, 2026Computer Software Problem; Activation Failure

    SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WITH THE ARTIS ICONO CEILING SYSTEM. DURING SYSTEM BOOTUP, IT WAS DETERMINED THAT STAND MOVEMENTS WERE NOT POSSIBLE, AND A STAND/TABLE ERROR MESSAGE WAS DISPLAYED TO THE USER. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885278 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K221516, K241572, K254173. One submission often covers many models, so check the product and lot details in each record.

  • Z-2641-2026 · May 20, 2026 · Open, Classified
    ARTIS icono biplane. Model Number: 11327600.
  • Z-2643-2026 · May 20, 2026 · Open, Classified
    ARTIS icono floor. Model Number: 11327700.
  • Z-1814-2026 · Mar 12, 2026 · Open, Classified
    ARTIS Pheno VE30A and VE40A, Model 10849000
  • Z-2205-2026 · Mar 9, 2026 · Open, Classified
    ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
  • Z-2529-2025 · Aug 12, 2025 · Open, Classified
    Artis Pheno. Image-Intensified Flouroscopic X-Ray System.
  • Z-2020-2025 · May 15, 2025 · Open, Classified
    interventional fluoroscopic x-ray system
  • Z-2022-2025 · May 15, 2025 · Open, Classified
    interventional fluoroscopic x-ray system

Other System, x-ray, angiographic devices

All System, x-ray, angiographic devices →