Skip to content

Paradigm REAL-Time Revel recallZ-1752-2026 · Medtronic MiniMed, Inc. · initiated Feb 13, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic MiniMed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

FDA root cause category: Device Design

What the company is doing

On 2/13/2026, communications began via email, SMS, certified postal mail, and in-app push notifications. Correction notices were sent informing customers of the following: User Guides will be updated to include: "It is important to keep the location of your pump stable relative to your infusion site. Do not wear or place your pump more than 14 in (35.5 cm) above your infusion site. Doing so can cause an over-delivery of insulin, which may result in hypoglycemia." User guides that are updated will be available at: https://www.medtronicdiabetes.com/download-library visit https://info.medtronicdiabetes.com/pump-placement for affected user guide versions. Required Actions: Keep the location of your insulin pump close to your infusion site. - If you need to raise the pump above your infusion site, such as when a parent or caregiver is using the pump for a child, limit the time that the pump is held above the infusion site. - Pay attention to any alerts from your pump as well as symptoms of hypoglycemia and hyperglycemia; follow your healthcare professional's treatment instructions in these situations. - If you have any concerns about how this may impact insulin delivery and blood glucose, please contact your healthcare professional. You may also contact firm at 1-855-587-4950 with any questions about this notification. -Complete and return the acknowledgement form by visiting https://info.medtronicdiabetes.com/pump-placement Distributor Partner / Service Provider: Send existing patients the Patient Notification. To HCPs send the HCP Notification and Patient Notification

Recall details

Recall number
Z-1752-2026
Status
Open, Classified
Initiated by the company
Feb 13, 2026
Posted by the FDA
Apr 2, 2026
Recalling firm
Medtronic MiniMed, Inc.
Quantity
55,724
Product code
N/A
Distribution
Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.

Product description

Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Veo: 00613994717801, 00613994713834, 00613994713841, 00613994870414, 00613994870377, 00613994870407, 00643169378636, 00613994924957, 00613994717672, 00613994715531, 00613994715555, 00613994715562, 00613994409423, 00613994951502, 00643169026131, 00613994622600, 00613994407740, 00643169077737, 00643169464537, 00643169305038, 00643169500259, 00643169503229, 00643169528789, 00643169533776, 00613994151766, 00643169711440, 00763000084370, 00763000161224, 00763000161262, 00643169464520, 00613994963376, 00643169026100, 00613994924247, 00643169299757, 00643169400931, 00643169400986, 00643169535749, 00643169535794, 00643169519190, 00643169579507, 00643169519091, 00763000084349, 00763000084356, 00763000161248, 00763000048853, 00763000161231, 00643169729445, 00643169846036, 00643169579453, 00613994714183, 00613994716156, 00613994657671, 00643169515741, 00643169579514, 00643169579538, 00643169579477, 00643169579569, 00643169579392, 00643169579583, 00643169579552, 00643169579576, 00643169846012, 00643169845909, 00643169845947, 00643169845992, 00643169846043, 00763000067663, 00643169846005, 00763000001247, 00643169578470, 00763000059729, 00763000048860,00763000048914, 00763000048839, 00763000048884, 00763000048822, 00763000048808, 00763000048938, 00763000048815, 00763000048921, 00763000048891, 00763000048785, 00763000048945, 00763000048877, 00763000048846, 00613994924438, 00613994924223, 00613994924391, 00613994924339, 00613994924346, 00613994924216, 00613994924278, 00763000084387, 00643169196766, 00613994611697, 00613994715944, 00613994714190, 00613994716002, 00613994716248, 00613994717719, 00613994716194, 00613994716361, 00613994716422, 00613994716415, 00613994716200, 00613994716354, 00613994716187, 00613994716323, 00613994716408, 00613994716385, 00613994716378, 00613994716217, 00613994863560, 00613994863515, 00613994863607, 00613994863409, 00613994863614, 00613994863355, 00613994863393, 00613994863430, 00613994863416, 00613994863553, 00613994863577, 00613994863539, 00613994863584, 00613994974273, 00613994870278, 00613994863621, 00613994924032, 00613994924254, 00613994863638, 00613994924353, 00613994924414, 00613994924445, 00613994863522, 00613994924407, 00613994863546, 00613994924384, 00613994924230, 00613994924421, 00613994750969, 00643169196674, 00643169196612, 00643169196704, 00643169196735, 00643169196698, 00643169196711, 00643169196773, 00643169196681, 00643169196728, 00643169196759, 00643169196742, 00643169196780, 00643169196636, 00643169196650, 00643169341654, 00643169317550, 00643169317574, 00643169317611, 00643169317567, 00643169317659, 00643169317642, 00643169317666, 00643169317536, 00643169306912, 00643169317581, 00643169341647, 00643169341661, 00643169324039, 00643169318960, 00643169317604, 00643169317598, 00643169317628, 00643169341630, 00643169324084, 00643169326668, 00643169317635, 00643169319370, 00643169381827, 00643169378995, 00643169379008, 00643169381810, 00643169378988, 00643169381780, 00643169381834, 00643169381773, 00643169381889, 00643169381872, 00643169381803, 00643169381797, 00643169381896, 00643169381766, 00643169381841, 00643169381902, 00643169428232, 00643169383838, 00643169379572, 00643169378704, 00643169411692, 00643169382725, 00643169495593, 00643169381865, 00763000084363, 00643169376090, 00643169579545, 00643169579521, 00643169593404, 00643169579439, 00643169579415, 00643169579484, 00643169579491, 00643169742048, 00643169579408, 00643169579422, 00643169845985, 00763000161279, 00763000217396, 00643169845954, 00643169845930, 00643169846067, 00643169846050, 00643169845961, 00643169846029, 00763000161255, 00643169845923, 00763000048907, 00763000048792: Labeling/Packaging; INSTR, INSULIN PUMP 554/754 (6025362 1) INSTRUCTIONS FOR USE, 554/754, DE (6025362-071 4); INSTR,INSULIN PUMP, 554/754 (6025362-022 7); INSTR,INSULIN PUMP, 554/754 (6025362-021 2); INSTR, 554/754, EN, A52 (6025362-024 1); INSTR. FOR USE, 554/754, EN (6025362-023 2); INSTR., INSULIN PUMP, 554/754 (6025362-026 3); INSTR, 554/754, EN (6025362-025 2); Artwork, New Product, Feature, (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-554/754 PRDGM VEO PUMP W/ENLITE, Nordic Cover, EN, OUS (MP6025679-025 1); Artwork, CRB D Process; Instr, X54, Paradigm, Nordic, EN (MP6025679-023 3); Artwork, CRB A Process; Instr, X54, Paradigm, Nordic, EN (MP6025679-024 1)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98186).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page