MiniMed 530G recallZ-1751-2026 · Medtronic MiniMed, Inc. · initiated Feb 13, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
FDA root cause category: Device Design
What the company is doing
On 2/13/2026, communications began via email, SMS, certified postal mail, and in-app push notifications. Correction notices were sent informing customers of the following: User Guides will be updated to include: "It is important to keep the location of your pump stable relative to your infusion site. Do not wear or place your pump more than 14 in (35.5 cm) above your infusion site. Doing so can cause an over-delivery of insulin, which may result in hypoglycemia." User guides that are updated will be available at: https://www.medtronicdiabetes.com/download-library visit https://info.medtronicdiabetes.com/pump-placement for affected user guide versions. Required Actions: Keep the location of your insulin pump close to your infusion site. - If you need to raise the pump above your infusion site, such as when a parent or caregiver is using the pump for a child, limit the time that the pump is held above the infusion site. - Pay attention to any alerts from your pump as well as symptoms of hypoglycemia and hyperglycemia; follow your healthcare professional's treatment instructions in these situations. - If you have any concerns about how this may impact insulin delivery and blood glucose, please contact your healthcare professional. You may also contact firm at 1-855-587-4950 with any questions about this notification. -Complete and return the acknowledgement form by visiting https://info.medtronicdiabetes.com/pump-placement Distributor Partner / Service Provider: Send existing patients the Patient Notification. To HCPs send the HCP Notification and Patient Notification
Recall details
- Recall number
- Z-1751-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 13, 2026
- Posted by the FDA
- Apr 2, 2026
- Recalling firm
- Medtronic MiniMed, Inc.
- Quantity
- 13,429
- Product code
- OZO
- FDA submissions
- P120010
- Distribution
- Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
Product description
MiniMed 530G Insulin Pump (MMT-551, MMT-751)
Code information (models, lots, serial numbers)
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 530G: 00643169507784, 00643169736061, 00643169736092, 00643169736085, 00643169933224, 00763000161170: Artwork, CRB A Process, Instr, 530G, One Press, Launch, US(MP6026081-012DOC); Artwork, CRB A Process, Instr, 530G, FDA comments, add TST(MP6025813-013DOC); Artwork, CRB A Process, Instr, 530G, Remove Velcro, Icons, US(MP6025813-014DOC);
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- MiniMed® 530G MMT-751LNASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend1 recall7 deaths3,201 reports
- MiniMed® 530G MMT-751LNASMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend1 recall2 deaths2,218 reports
- MiniMed® 530G MMT-751LNASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend1 recall5 deaths1,923 reports
- MiniMed® 530G MMT-751NALMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend1 recall1 death58 reports
- MiniMed® 530G MMT-751NABMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend1 recall49 reports
- MiniMed® 530G MMT-751NAPMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend1 recall29 reports
Product lines
Other devices under the same FDA submission
70 devices cleared under P120010 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed® 530G MMT-751NALMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend15 deaths3,405 reports
- MiniMed® 530G MMT-551NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend5 deaths2,944 reports
- MiniMed® 530G MMT-751NABMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend11 deaths2,700 reports
- MiniMed® 530G MMT-751NAPMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend12 deaths2,637 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend16 deaths1,995 reports
- MiniMed® 530G MMT-551NALMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend6 deaths1,938 reports
- MiniMed® 530G MMT-751NAHMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend5 deaths1,368 reports
- Enlite® Sensor MMT-7008AMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend1,312 reports
- MiniMed® 530G MMT-751LNALMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend7 deaths1,267 reports
- MiniMed® 530G MMT-551NAPMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend5 deaths1,246 reports
- Enlite ® MMT-7008AMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend994 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98186).
- Z-1739-2026 · Open, ClassifiedMiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
- Z-1740-2026 · Open, ClassifiedMiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-1741-2026 · Open, ClassifiedMiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
- Z-1742-2026 · Open, ClassifiedMiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1748-2026 · Open, ClassifiedParadigm Insulin Pump (MMT-712, MMT-715)
- Z-1749-2026 · Open, ClassifiedParadigm REAL-Time Insulin Pump (MMT-522, MMT-722)
- Z-1750-2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1752-2026 · Open, ClassifiedParadigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.