Skip to content

NAMIC recallZ-2138-2026 · Medline Industries, LP · initiated Mar 24, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline has identified the presence of particulate within the fluid path of the Manifolds.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries, LP issued an URGENT MEDICAL DEVICE notice to its consignees on 3/24/2026 via email and mail. The notice explained the issue, potential risk and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-046 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2138-2026
Status
Open, Classified
Initiated by the company
Mar 24, 2026
Posted by the FDA
May 8, 2026
Recalling firm
Medline Industries, LP
Quantity
7075 units
Product code
DTL
FDA submissions
K782095
Distribution
US (nationwide) and AE, AU, CA, JP, KR, NL, SG, SK.

Product description

Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.

Code information (models, lots, serial numbers)

UDI/DI each 10193489059144, UDI/Di case 30193489059148, Lot Numbers: 90184, 94779, 97786, 106351, 108597, 109876, 114056, 114989, 115459, 115931, 116358, 116865, 118245, 122974, 126077, 127795, 129531, 130557, 136114, 150501, 151591, 152772, 154514, 156045, 160318, 163255, 168053.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98735).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page