NAMIC recallZ-2147-2026 · Medline Industries, LP · initiated Mar 24, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline has identified the presence of particulate within the fluid path of the Manifolds.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline Industries, LP issued an URGENT MEDICAL DEVICE notice to its consignees on 3/24/2026 via email and mail. The notice explained the issue, potential risk and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-046 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2147-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 24, 2026
- Posted by the FDA
- May 8, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 21119 kits
- Product code
- LRO
- Distribution
- US (nationwide) and AE, AU, CA, JP, KR, NL, SG, SK.
Product description
Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. MTS,RT,HANDED/LFT,HRT,KIT, Medline Kit Number/SKU 60010508; 2. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60160075; 3. CARDIAC CATH PACK, Medline Kit Number/SKU DYNJ04636N; 4. CATH LAB TRAY, Medline Kit Number/SKU DYNJ24250G; 5. CATH LAB PACK-LF, Medline Kit Number/SKU DYNJ38333D; 6. CATH LAB PACK-LF, Medline Kit Number/SKU DYNJ38333F; 7. CATH LAB ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ50779G; 8. CATH LAB ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ50779I; 9. CLOT RETRIEVAL PACK, Medline Kit Number/SKU DYNJ56140C; 10. TAVR PACK NO SYR, Medline Kit Number/SKU DYNJ62310C; 11. TAVR PACK, Medline Kit Number/SKU DYNJ62310D; 12. TAVR PACK, Medline Kit Number/SKU DYNJ62310F; 13. DIAGNOSTIC 2 PACK, Medline Kit Number/SKU DYNJ81911; 14. DIAGNOSTIC 2 PACK, Medline Kit Number/SKU DYNJ81911A; 15. CATH PACK, Medline Kit Number/SKU DYNJT6544; 16. LHK, Medline Kit Number/SKU NAM0015; 17. SLU LHK, Medline Kit Number/SKU SAMPC0706.
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23KBN009; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23LBI829; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24ABH036; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24BBG055; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24EBO156; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24GBS605; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24IBK765; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25ABR102; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25CBK686; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25DBR341; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25EBB568; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25FBB624; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25GBC747; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25HBF702; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25JBJ140; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 26ABB718; Medline Kit Number/SKU 60010508: UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 26BBI625; Medline Kit Number/SKU 60160075: UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23IBC156; Medline Kit Number/SKU 60160075: UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23KBB857; Medline Kit Number/SKU 60160075: UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24ABL640; Medline Kit Number/SKU 60160075: UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24CBD613; Medline Kit Number/SKU 60160075: UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24EBQ349; Medline Kit Number/SKU DYNJ04636N: UDI/DI each 10198459580918, UDI/DI case 40198459580919, Lot Number: 26AMC406; Medline Kit Number/SKU DYNJ04636N: UDI/DI each 10198459580918, UDI/DI case 40198459580919, Lot Number: 26CMA066; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25GBI702; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25GBQ931; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25JBG417; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25KBH544; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26AMC371; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26AMC433; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26CMC609; Medline Kit Number/SKU DYNJ24250G: UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26CMC660; Medline Kit Number/SKU DYNJ38333D: UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24GMH056; Medline Kit Number/SKU DYNJ38333D: UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24HMB740; Medline Kit Number/SKU DYNJ38333D: UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24JME459; Medline Kit Number/SKU DYNJ38333D: UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24KMJ018; Medline Kit Number/SKU DYNJ38333D: UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 25AMG765; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25CME621; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25EMB706; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25EME362; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25HME818; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25KME742; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 26AME732; Medline Kit Number/SKU DYNJ38333F: UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 26BME198; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24EMG321; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24GMA788; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24HMA312; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24IMI472; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24KMC197; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25BMC031; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25BMF563; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25CMI786; Medline Kit Number/SKU DYNJ50779G: UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25DMH709; Medline Kit Number/SKU DYNJ50779I: UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25GMF741; Medline Kit Number/SKU DYNJ50779I: UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25HME177; Medline Kit Number/SKU DYNJ50779I: UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25IME495; Medline Kit Number/SKU DYNJ50779I: UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25KMB250; Medline Kit Number/SKU DYNJ50779I: UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25LMC853; Medline Kit Number/SKU DYNJ50779I: UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 26CMD408; Medline Kit Number/SKU DYNJ56140C: UDI/DI each 10198459477218, UDI/DI case 40198459477219, Lot Number: 25HBV231; Medline Kit Number/SKU DYNJ56140C: UDI/DI each 10198459477218, UDI/DI case 40198459477219, Lot Number: 26ABK843; Medline Kit Number/SKU DYNJ56140C: UDI/DI each 10198459477218, UDI/DI case 40198459477219, Lot Number: 26CBE251; Medline Kit Number/SKU DYNJ62310C: UDI/DI each 10198459104299, UDI/DI case 40198459104290, Lot Number: 24JMG837; Medline Kit Number/SKU DYNJ62310C: UDI/DI each 10198459104299, UDI/DI case 40198459104290, Lot Number: 24JMH356; Medline Kit Number/SKU DYNJ62310C: UDI/DI each 10198459104299, UDI/DI case 40198459104290, Lot Number: 24KMA339; Medline Kit Number/SKU DYNJ62310C: UDI/DI each 10198459104299, UDI/DI case 40198459104290, Lot Number: 24KMC940; Medline Kit Number/SKU DYNJ62310D: UDI/DI each 10198459254727, UDI/DI case 40198459254728, Lot Number: 25BMA421; Medline Kit Number/SKU DYNJ62310D: UDI/DI each 10198459254727, UDI/DI case 40198459254728, Lot Number: 25BME140; Medline Kit Number/SKU DYNJ62310D: UDI/DI each 10198459254727, UDI/DI case 40198459254728, Lot Number: 25CMH080; Medline Kit Number/SKU DYNJ62310D: UDI/DI each 10198459254727, UDI/DI case 40198459254728, Lot Number: 25DMB438; Medline Kit Number/SKU DYNJ62310D: UDI/DI each 10198459254727, UDI/DI case 40198459254728, Lot Number: 25EMA045; Medline Kit Number/SKU DYNJ62310F: UDI/DI each 10198459393327, UDI/DI case 40198459393328, Lot Number: 25GMI604; Medline Kit Number/SKU DYNJ62310F: UDI/DI each 10198459393327, UDI/DI case 40198459393328, Lot Number: 25HME419; Medline Kit Number/SKU DYNJ62310F: UDI/DI each 10198459393327, UDI/DI case 40198459393328, Lot Number: 25JMF984; Medline Kit Number/SKU DYNJ62310F: UDI/DI each 10198459393327, UDI/DI case 40198459393328, Lot Number: 25KMC303; Medline Kit Number/SKU DYNJ62310F: UDI/DI each 10198459393327, UDI/DI case 40198459393328, Lot Number: 25KMJ733; Medline Kit Number/SKU DYNJ62310F: UDI/DI each 10198459393327, UDI/DI case 40198459393328, Lot Number: 26BMD977; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23HBE467; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23HBW249; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23HBW251; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23IBC125; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23KBJ681; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23KBT543; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 23LBM984; Medline Kit Number/SKU DYNJ81911: UDI/DI each 10195327158699, UDI/DI case 40195327158690, Lot Number: 24BBR688; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 24EBP711; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 24GBP073; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 24IBF171; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 24KBH753; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 24LBH071; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 25ABM562; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 25CBM896; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 25FBJ466; Medline Kit Number/SKU DYNJ81911A: UDI/DI each 10198459005640, UDI/DI case 40198459005641, Lot Number: 25HBJ125; Medline Kit Number/SKU DYNJT6544: UDI/DI each 10198459558689, UDI/DI case 40198459558680, Lot Number: 25KMF858; Medline Kit Number/SKU DYNJT6544: UDI/DI each 10198459558689, UDI/DI case 40198459558680, Lot Number: 26BMK015; Medline Kit Number/SKU NAM0015: UDI/DI each 10195327374198, UDI/DI case 40195327374199, Lot Number: 23GBQ457; Medline Kit Number/SKU SAMPC0706: UDI/DI each 10195327586157, UDI/DI case 40195327586158, Lot Number: 24ADA538.
Devices named in this recall
13 devices in the FDA device registry whose identifier appears in the recall's code information.
- NAMIC 60010508Medline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. DYNJ38333DMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ38333FMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. 60160075Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Medline Industries, Inc. DYNJ50779GMedline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ24250GMedline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ50779IMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ81911Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. NAM0015Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. SAMPC0706Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ81911AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62310CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62310DMedline Industries, Inc. · General surgery tray1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98735).
- Z-2138-2026 · Open, ClassifiedNamic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
- Z-2139-2026 · Open, ClassifiedNamic Angiographic Manifold, custom, Medline Product Number/SKU 64038301; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
- Z-2140-2026 · Open, ClassifiedNamic Angiographic Manifold, custom, Medline Product Number/SKU 64038402; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
- Z-2141-2026 · Open, ClassifiedNamic Preceptor Manifold, custom, Medline Product Number/SKU 64037107; Extravascular blood pressure transducer
- Z-2142-2026 · Open, ClassifiedNamic Preceptor Manifold, custom, Medline Product Number/SKU 64039307; Extravascular blood pressure transducer
- Z-2143-2026 · Open, ClassifiedMedline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ6…
- Z-2144-2026 · Open, Classified1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 3. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 4. MTS,LEFT …
- Z-2145-2026 · Open, Classified1. 3 VALVE MAN KIT, Medline Kit SKU Number: 65101544; 2. 3V MANIFOLD KIT, Medline Kit SKU Number: 65110351; 3. HEART CATH PACK-LF, Medline Kit SKU Number: DYNJ0373285M; 4. AN…
- Z-2146-2026 · Open, ClassifiedMedline medical convenience kits, containing Namic Manifold, Labeled as: HEART CATH PACK-LF, Medline Kit Number/SKU DYNJ36478B
- Z-2148-2026 · Open, ClassifiedMedline medical convenience kits, containing Namic Manifold, Labeled as: TAVI, Medline Kit Number/SKU DYNJ904981C
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.