ARTIS Pheno VE30A and VE40A, Model 10849000 recallZ-1814-2026 · Siemens Medical Solutions USA, Inc · initiated Mar 12, 2026

Class II Open, Classified All recalls posted in April 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers dated March 9, 2026.

Recall details

Recall number
Z-1814-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Mar 12, 2026
Posted by the FDA
Apr 16, 2026
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
73
Product code
OWB
FDA submissions
K230949, K241572
Distribution
U.S. and O.U.S.

Product description

ARTIS Pheno VE30A and VE40A, Model 10849000

Code information (models, lots, serial numbers)

UDI 04056869046877

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K230949, K241572 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page