Ingenia recallZ-1968-2026 · Philips North America · initiated Apr 14, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
FDA root cause category: Under Investigation by firm
What the company is doing
Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.
Recall details
- Recall number
- Z-1968-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 14, 2026
- Posted by the FDA
- Apr 27, 2026
- Recalling firm
- Philips North America
- Quantity
- 3 units
- Product code
- LNH
- FDA submissions
- K193215, K213583, K251397
- Distribution
- Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
Product description
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
Code information (models, lots, serial numbers)
1. Model Number (REF): 782118. UDI: (01)00884838099746(21)85655. Serial Numbers: 85655. 2. Model Number (REF): 782144. UDI: (01)00884838108677(21)71087. Serial Numbers: 71087. 3. Model Number (REF): 782163. UDI: (01)00884838115743(21)42299. Serial Numbers: 42299.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia SmartPath to Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging9 recalls0 reports
- Ingenia SmartPath to Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging5 recalls0 reports
- Ingenia SmartPath to Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98779).
- Z-1954-2026 · Open, ClassifiedPhilips Evolution Upgrade 1.5T with MR Elastography (MRE). 1. Model Number (REF): 782116 2. Model Number (REF): 782148. 3. Model Number (REF): 782166.
- Z-1955-2026 · Open, ClassifiedPhilips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.
- Z-1956-2026 · Open, ClassifiedPhilips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115.…
- Z-1957-2026 · Open, ClassifiedPhilips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
- Z-1958-2026 · Open, ClassifiedPhilips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K1730…
- Z-1959-2026 · Open, ClassifiedPhilips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458…
- Z-1960-2026 · Open, ClassifiedPhilips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.
- Z-1961-2026 · Open, ClassifiedPhilips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 7…
- Z-1962-2026 · Open, ClassifiedPhilips Ingenia Elition S with MR Elastography (MRE). 1. Model Number (REF): 781357. 2. Model Number (REF): 782106. 3. Model Number (REF): 782137.
- Z-1963-2026 · Open, ClassifiedPhilips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516…
- Z-1964-2026 · Open, ClassifiedPhilips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
- Z-1965-2026 · Open, ClassifiedPhilips SmartPath to dStream for 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781260. 2. Model Number (REF): 782112.
- Z-1966-2026 · Open, ClassifiedPhilips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
- Z-1967-2026 · Open, ClassifiedPhilips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
- Z-1969-2026 · Open, ClassifiedPhilips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page