Skip to content

Ingenia recallZ-1963-2026 · Philips North America · initiated Apr 14, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

FDA root cause category: Under Investigation by firm

What the company is doing

Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.

Recall details

Recall number
Z-1963-2026
Status
Open, Classified
Initiated by the company
Apr 14, 2026
Posted by the FDA
Apr 27, 2026
Recalling firm
Philips North America
Quantity
103 units
Product code
LNH
FDA submissions
K213516, K251808
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.

Product description

Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

1. Model Number (REF): 781358. UDI: (01)00884838088115(21)45121, (01)00884838088115(21)45148, (01)00884838088115(21)45160, (01)00884838088115(21)45438, (01)00884838088115(21)45034, (01)00884838088115(21)45049, (01)00884838088115(21)45066, (01)00884838088115(21)45090, (01)00884838088115(21)45141, (01)00884838088115(21)45149, (01)00884838088115(21)45170, (01)00884838088115(21)45250, (01)00884838088115(21)45290, (01)00884838088115(21)45085, (01)00884838088115(21)45312, (01)00884838088115(21)45376, (01)00884838088115(21)45122, (01)00884838088115(21)45169, (01)00884838088115(21)45191, (01)00884838088115(21)45192, (01)00884838088115(21)45203, (01)00884838088115(21)45204, (01)00884838088115(21)45207, (01)00884838088115(21)45220, (01)00884838088115(21)45228, (01)00884838088115(21)45229, (01)00884838088115(21)45249, (01)00884838088115(21)45278, (01)00884838088115(21)45279, (01)00884838088115(21)45320, (01)00884838088115(21)45028, (01)00884838088115(21)45195, (01)00884838088115(21)45174, (01)00884838088115(21)45165, (01)00884838088115(21)45382, (01)00884838088115(21)45011, (01)00884838088115(21)45055, (01)00884838088115(21)45075, (01)00884838088115(21)45113, (01)00884838088115(21)45147, (01)00884838088115(21)45152, (01)00884838088115(21)45235, (01)00884838088115(21)45237, (01)00884838088115(21)45245, (01)00884838088115(21)45264, (01)00884838088115(21)45269, (01)00884838088115(21)45277, (01)00884838088115(21)45316, (01)00884838088115(21)45322, (01)00884838088115(21)45337, (01)00884838088115(21)45338, (01)00884838088115(21)45353, (01)00884838088115(21)45398, (01)00884838088115(21)45414, (01)00884838088115(21)45429, (01)00884838088115(21)45051. Serial Numbers: 45121, 45148, 45160, 45438, 45034, 45049, 45066, 45090, 45141, 45149, 45170, 45250, 45290, 45085, 45312, 45376, 45122, 45169, 45191, 45192, 45203, 45204, 45207, 45220, 45228, 45229, 45249, 45278, 45279, 45320, 45028, 45195, 45174, 45165, 45382, 45011, 45055, 45075, 45113, 45147, 45152, 45235, 45237, 45245, 45264, 45269, 45277, 45316, 45322, 45337, 45338, 45353, 45398, 45414, 45429, 45051. 2. Model Number (REF): 782107. UDI: (01)00884838098336(21)45590, (01)00884838098336(21)62158, (01)00884838098336(21)62180, (01)00884838098336(21)62044, (01)00884838098336(21)45567, (01)00884838098336(21)62031, (01)00884838098336(21)62166, (01)00884838098336(21)62184, (01)00884838098336(21)62190, (01)00884838098336(21)62240, (01)00884838098336(21)45519, (01)00884838098336(21)45533, (01)00884838098336(21)62024, (01)00884838098336(21)62047, (01)00884838098336(21)62048, (01)00884838098336(21)62051, (01)00884838098336(21)62052, (01)00884838098336(21)45554, (01)00884838098336(21)62145, (01)00884838098336(21)62146, (01)00884838098336(21)62148, (01)00884838098336(21)45500, (01)00884838098336(21)45501, (01)00884838098336(21)45510, (01)00884838098336(21)45511, (01)00884838098336(21)45514, (01)00884838098336(21)45516, (01)00884838098336(21)45518, (01)00884838098336(21)45589, (01)00884838098336(21)62017, (01)00884838098336(21)62039, (01)00884838098336(21)62073, (01)00884838098336(21)62081, (01)00884838098336(21)62088, (01)00884838098336(21)45521, (01)00884838098336(21)45600, (01)00884838098336(21)45563, (01)00884838098336(21)45570, (01)00884838098336(21)62153, (01)00884838098336(21)62181, (01)00884838098336(21)62182, (01)00884838098336(21)62183. Serial Numbers: 45590, 62158, 62180, 62044, 45567, 62031, 62166, 62184, 62190, 62240, 45519, 45533, 62024, 62047, 62048, 62051, 62052, 45554, 62145, 62146, 62148, 45500, 45501, 45510, 45511, 45514, 45516, 45518, 45589, 62017, 62039, 62073, 62081, 62088, 45521, 45600, 45563, 45570, 62153, 62181, 62182, 62183. 3. Model Number (REF): 782136. UDI: (01)00884838108608(21)28568, (01)00884838108608(21)28503, (01)00884838108608(21)28515, (01)00884838108608(21)28528, (01)00884838108608(21)28577. Serial Numbers: 28568, 28503, 28515, 28528, 28577.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98779).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page