Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT… recallZ-2296-2026 · Bayer Medical Care, Inc. · initiated Apr 24, 2026
Class II Open, Classified All recalls posted in June 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bayer Medical Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
FDA root cause category: Process control
What the company is doing
On April 24, 2026, URGENT MEDICAL DEVICE RECALL letters were sent to customers. Customers were instructed to immediately halt the use of the affected product and review inventory for impacted product and quarantine. A Return Goods Authorization will be sent to return any remaining affected inventory. Place the affected product in a box and label the outside of the container with the RGA number provided and RECALL ID: SA-2026-RAD-03 in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return the affected product as soon as possible. Please distribute this letter to other appropriate departments and personnel within your facility who may need awareness of this recall.
Recall details
- Recall number
- Z-2296-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 24, 2026
- Posted by the FDA
- Jun 3, 2026
- Recalling firm
- Bayer Medical Care, Inc.
- Quantity
- 11,300 units (1650 US, 9650 OUS)
- Product code
- DXT
- FDA submissions
- K050456
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
Product description
Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;
Code information (models, lots, serial numbers)
Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007, 60729458, 86566648; Affected batch numbers: 251702, 252302, 252502, 252602, 252702, 252802, 253304, 253702, 253704, 253804, 254202, 254304, 254704, 254802;
Other devices under the same FDA submission
3 devices cleared under K050456 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MEDRAD® Avanta Fluid Management Injection SystemBayer Medical Care Inc. · Injector And Syringe, Angiographic1 death5 reports
- MEDRAD® Avanta Fluid Management Injection SystemBayer Medical Care Inc. · Injector And Syringe, Angiographic1 death3 reports
- MEDRAD® Avanta Single Patient Disposable Set – Transducer Compatible AVA 500 SPAT LBayer Medical Care Inc. · Injector And Syringe, Angiographic1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.