Skip to content

Neptune recallZ-2282-2026 · Stryker Corporation · initiated May 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Stryker issued an URGENT Medical Device Field Action notice to its consignee on 5/6/2026 via FedEx. The notice explained the issue, potential risk and requested the consignee locate and quarantine the device. The consignee was directed to contact their Stryker Sales Rep to help schedule replacement of the affected device. For questions of concerns, contact Brooke Thompson at [email protected].

Recall details

Recall number
Z-2282-2026
Status
Open, Classified
Initiated by the company
May 6, 2026
Posted by the FDA
Jun 1, 2026
Recalling firm
Stryker Corporation
Quantity
1 unit
Product code
JCX
FDA submissions
K132671
Distribution
US Nationwide distribution in the state of Indiana.

Product description

Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System

Code information (models, lots, serial numbers)

GTIN 07613327634914, Serial Number 2521008873

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines