Skip to content

Neptune 0770100000Recalls & reported problems

Stryker Corporation · SMART DOCKING STATION (120V) · Class 2

All models of Neptune (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 6, 2026Class IIOpen, Classified

    It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.

    Z-2282-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Other APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED devices

All APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page