Skip to content

Atlan A350. Model Number: 8211500. anesthesia workstation recallZ-2207-2026 · Draeger, Inc. · initiated May 7, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

FDA root cause category: Process change control

What the company is doing

Draeger notified consignees of the recall update and expansion on about 05/07/2026 via emailed letter titled Urgent Medical Device Correction Update. Consignees were made aware that due to continuous monitoring, Draeger has determined that an additional number of Atlan devices are affected by the issue and they are replacing ventilator motors. Consignees were instructed that a Draeger Service Representative will contact consignees to arrange a date for the ventilator motor assembly replacement, and until the corrective action is implemented, the device may continue to be used with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Consignees were also instructed to complete and return the provided Medical Device Correction Return Response Acknowledgment and Receipt Form and if products have been made available to third parties, please forward this Urgent Medical Device Correction notice to them.

Recall details

Recall number
Z-2207-2026
Status
Open, Classified
Initiated by the company
May 7, 2026
Posted by the FDA
Jun 4, 2026
Recalling firm
Draeger, Inc.
Quantity
3,515 units
Product code
BSZ
FDA submissions
K230931
Distribution
US distribution in WI, AR, MN, WI, CO, MA, PA, AZ, FL, NY, SD, ND. International distribution to: Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chili, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Gabon, Germany, Ghana, Greece, Greenland, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Italy, Israel, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mauritius, Mexico, Morocco, Mozambique, Namibia, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Uzbekistan, Venezuela, Vietnam, Yemen

Product description

Atlan A350. Model Number: 8211500. anesthesia workstation

Code information (models, lots, serial numbers)

Model Number: 8211500. UDI-DI Number: 04048675556176.Serial Numbers: ASSL-0347, ASSL-0348, ASSL-0351, ASSH-0220, ASSH-0221, ASSH-0224, ASSH-0225, ASSH-0228, ASSH-0229, ASSH-0230, ASSH-0234, ASSH-0235, ASTA-0086, ASSL-0349, ASSL-0350, ASSL-0352, ASSL-0353, ASSH-0222, ASSH-0223, ASSH-0226, ASSH-0227, ASSH-0231, ASSH-0232, ASSH-0233, ASSH-0236, ASSH-0237, ASSH-0238, ASSN-0185, ASSN-0186, ASSN-0187, ASSN-0188, ASSN-0189, ASSN-0190, ASTB-0037, ASTB-0040, ASTA-0087, ASTA-0088, ASSM-0130, ASSM-0131, ASSM-0132.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K230931 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98914).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page