MAC recallZ-2306-2026 · GE Medical Systems China Co., Ltd. · initiated Apr 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems China Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
Firm notified consignees beginning April 21, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION". Customers were given instructions on how to check their software version. If a customer's software version begins with 3.00, the customer should not use the system until a GE HealthCare representative has updated the device with the latest licensed software version.
Recall details
- Recall number
- Z-2306-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2026
- Posted by the FDA
- Jun 3, 2026
- Recalling firm
- GE Medical Systems China Co., Ltd.
- Quantity
- 16
- Product code
- DXH
- FDA submissions
- K221321
- Distribution
- Worldwide - US Nationwide distribution in the states of Florida and Georgia. The countries of India, Italy, Republic of Korea, and United Kingdom.
Product description
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
Code information (models, lots, serial numbers)
Software version 3.00 855001-001, UDI-DI 00195278276070 8855002-001, UDI-DI 00195278276124
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.