Skip to content

MACRecalls & reported problems across all models

Ge Medical Systems Information Technologies, Inc. · 4 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name at least one MAC model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls
4

We found 4 models of MAC by Ge Medical Systems Information Technologies, Inc. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Sep 17, 2020 and Sep 17, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

4 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

MAC recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 29, 2026Open, Classified

    Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.

    Z-2876-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Apr 21, 2026Open, Classified

    Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.

    Z-2306-2026 · Root cause: Software Manufacturing/Software Deployment · names 2 of these models · Recall details · FDA record ↗

  3. Aug 22, 2025Open, Classified

    If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

    Z-0231-2026 · Root cause: Device Design · Recall details · FDA record ↗

  4. Jun 28, 2019Terminated

    Two issues: 1) The MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient s barcode. This may result in the ECG report being assigned to the incorrect patient. 2) The MAC VU360 may have the incorrect patient demographics appear on the patient banner. This may result in the ECG report being assigned to the incorrect patient.

    Z-2060-2019 · Root cause: Software design · Recall details · FDA record ↗

Most reported MAC problems

Device problems

  • Device Handling Problem1
  • Smoking1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

MAC models

4 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent MAC adverse event reports

The 1 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Ge Medical Systems Information Technologies, Inc. products

All Ge Medical Systems Information Technologies, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page