Brand Name: CGuard Prime Carotid Stent System, 135cm… recallZ-2328-2026 · INSPIREMD Inc · initiated May 1, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from INSPIREMD Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Stent delivery system may encounter complications during deployment which may result in high resistance or inability to deploy the stent.
FDA root cause category: Component design/selection
What the company is doing
On May 1, 2026, InspireMD, Inc. issued a Urgent Medical Device Recall notification to affected consignees. InspireMD asked consignees to take the following actions: 1. Cease further use of all affected CGuard Prime Carotid Stent Systems identified by the model numbers listed in the letter. 2. Quarantine and segregate all affected units currently in your inventory. Place affected product in a clearly labeled quarantine area to prevent inadvertent use. 3. This notice needs to be passed on to all those who need to be aware within your organization or facilities where the affected devices have been transferred. 4. Please retain a copy of this letter, and keep a copy of the Acknowledgement Form. 5. Complete the attached Acknowledgement Form and return it by email to [email protected] within 2 business days. 6. InspireMD will contact each affected customer to facilitate the return and disposition process.
Recall details
- Recall number
- Z-2328-2026
- Status
- Open, Classified
- Initiated by the company
- May 1, 2026
- Posted by the FDA
- Jun 5, 2026
- Recalling firm
- INSPIREMD Inc
- Quantity
- 291 units
- Product code
- NIM
- FDA submissions
- P240029
- Distribution
- US Nationwide distribution in the states of AR, AZ, CA, DC, FL, GA, HI, IL, IN, KS, MA, MD, MI, MO, NC, NJ, NV, NY, OH, OR, RI, SD, TN, TX, VA, WA, WI, WV.
Product description
Brand Name: CGuard Prime Carotid Stent System, 135cm, 9mx40mm Model/Catalog Number: CND0940
Code information (models, lots, serial numbers)
UDI: 07290120281745; all lots/serial numbers manufactured between 02/03/2025 and 02/23/2026
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CGuard Prime Carotid Stent System CND0940Inspire M.D Ltd · Stent, Carotid1 recall31 reports
Other devices under the same FDA submission
5 devices cleared under P240029 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CGuard Prime Carotid Stent System CND0840Inspire M.D Ltd · Stent, Carotid1 recall1 death41 reports
- CGuard Prime Carotid Stent System CND0830Inspire M.D Ltd · Stent, Carotid1 recall15 reports
- CGuard Prime Carotid Stent System CND1040Inspire M.D Ltd · Stent, Carotid1 recall6 reports
- CGuard Prime Carotid Stent System CND1030Inspire M.D Ltd · Stent, Carotid1 recall2 reports
- CGuard Prime Carotid Stent System CND0930Inspire M.D Ltd · Stent, Carotid1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98922).
- Z-2325-2026 · Open, ClassifiedBrand Name: CGuard Prime Carotid Stent System, 135cm, 8mmx30mm Model/Catalog Number: CND0830
- Z-2326-2026 · Open, ClassifiedBrand Name: CGuard Prime Carotid Stent System, 135cm, 8mmx40mm Model/Catalog Number: CND0840
- Z-2327-2026 · Open, ClassifiedCGuard Prime Carotid Stent System, 135cm, 9mx30mm Model/Catalog Number: CND0930
- Z-2329-2026 · Open, ClassifiedCGuard Prime Carotid Stent System, 135cm, 10mmx30mm Model/Catalog Number: CND1030
- Z-2330-2026 · Open, ClassifiedCGuard Prime Carotid Stent System, 135cm, 10mmx40mm Model/Catalog Number: CND1040
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page