Ivenix, Inc. recallZ-2223-2026 · Fresenius Kabi USA, LLC · initiated May 6, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
FDA root cause category: Software design (manufacturing process)
What the company is doing
Consignees were notified on about 05/06/2026 via letter. Consignees were informed of the issue, risk associated, and instructed to keep the Ivenix pump plugged in at all times rather than rely on battery and if a "Battery Extremely Low" alarm occurs while on battery, immediately connect the pump to AC. For any transport of a patient on an affected pump, have a plan for AC access. For patients receiving critical or short half-life infusion therapies, ensure an RN accompanies the patient during transport. Consignees were also instructed to notify other users if applicable, notify customers if affected units were transferred or distributed, and post the provided Safety Tip Sheet, Battery Health Software Anomaly in the Ivenix Large Volume Pump (LVP) notice alongside this letter in all areas where affected pumps are stored and used. consignees were asked to complete and return the provided Customer Reply Form.
Recall details
- Recall number
- Z-2223-2026
- Status
- Open, Classified
- Initiated by the company
- May 6, 2026
- Posted by the FDA
- Jun 3, 2026
- Recalling firm
- Fresenius Kabi USA, LLC
- Quantity
- 32 systems
- Product code
- FRN
- FDA submissions
- K183311
- Distribution
- US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Product description
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Code information (models, lots, serial numbers)
Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ivenix, Inc. 2.0Fresenius Kabi AG · Pump, Infusion10 recalls0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Ivenix Infusion System LVP-0004Fresenius Kabi AG · Pump, Infusion7 recalls2 deaths4,328 reports
- Ivenix Infusion System 030-0028-08Fresenius Kabi AG · Pump, Infusion7 reports
- Ivenix Infusion System 030-0027-06Fresenius Kabi AG · Infusion Safety Management Software3 reports
- Ivenix Infusion System 030-0028-07Fresenius Kabi AG · Pump, Infusion2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page