Why it was recalled

Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.

FDA root cause category: Software design (manufacturing process)

What the company is doing

Consignees were notified on about 05/06/2026 via letter. Consignees were informed of the issue, risk associated, and instructed to keep the Ivenix pump plugged in at all times rather than rely on battery and if a "Battery Extremely Low" alarm occurs while on battery, immediately connect the pump to AC. For any transport of a patient on an affected pump, have a plan for AC access. For patients receiving critical or short half-life infusion therapies, ensure an RN accompanies the patient during transport. Consignees were also instructed to notify other users if applicable, notify customers if affected units were transferred or distributed, and post the provided Safety Tip Sheet, Battery Health Software Anomaly in the Ivenix Large Volume Pump (LVP) notice alongside this letter in all areas where affected pumps are stored and used. consignees were asked to complete and return the provided Customer Reply Form.

Recall details

Recall number
Z-2223-2026
Status
Open, Classified
Initiated by the company
May 6, 2026
Posted by the FDA
Jun 3, 2026
Recalling firm
Fresenius Kabi USA, LLC
Quantity
32 systems
Product code
FRN
FDA submissions
K183311
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.

Product description

LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

Code information (models, lots, serial numbers)

Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates