Intracept Access Instruments, Description/REF recallZ-2694-2026 · Boston Scientific Neuromodulation Corporation · initiated Apr 22, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Neuromodulation Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection.

FDA root cause category: Packaging process control

What the company is doing

On 4/22/2026, removal notices were mailed to materials managers and health care professionals who were asked to do the following: 1) Segregate affected devices until they can be returned to firm. 2) Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 3) Forward this notice to any healthcare professional from your organization for awareness and to any organization to which affected devices have been transferred. 4) Complete and return the enclosed Reply Verification Tracking Form via email to [email protected] Any adverse events or quality concerns associated with use of these devices should be reported to firm via email at mailto: [email protected] or

Recall details

Recall number
Z-2694-2026
Status
Open, Classified
Initiated by the company
Apr 22, 2026
Posted by the FDA
Jul 8, 2026
Recalling firm
Boston Scientific Neuromodulation Corporation
Quantity
37,764
Product code
GXI
FDA submissions
K222281
Distribution
Worldwide - US Nationwide distribution including in the states of NJ, WI, MI, MN, AR, UT, OH, FL, CA, TX, PA, AZ, IN, LA, NE, OK, NY, MD, WY, CO, NC, IL, GA, WA, KS, AL, MO, VA, TN, KY, MS, MA, SC, OR, WV, MT, ID, SD, CT, NV, ND, NH, DE, NM, AK, IA, DC, HI, ME, VT, RI and the countries of Great Britain, Germany, United Arab Emirates.

Product description

Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESS INSTRUMENTS- DRILL EU/RLV0035-EU, INTRACEPT ACCESS INSTRUMENTS- 2VB/RLV0045, INTRACEPT ACCESS INSTRUMENTS- 2VB EU/RLV0045-EU, INTRACEPT ACCESS INSTRUMENTS- 3VB/RLV0055, INTRACEPT ACCESS INSTRUMENTS- 3VB EU/RLV0055-EU, INTRACEPT ACCESS INSTRUMENTS 2 VB/RLV0065, INTRACEPT ACCESS INSTRMNTS ADDITIONAL VB/RLV0075, INTRACEPT ACCESS INSTRMNTS ADDTL VB EU/RLV0075-EU

Code information (models, lots, serial numbers)

Show the full code information (9k characters)

REF/UDI-DI/Lot/Expiration Date Range: RLV0035/852454006240/36231634, 36266333, 36266334, 36266335, 36442451, 36556998, 36577474, 36577475, 37355534, 37374614, 37642257, 37642258, 37667047, 37667049, 38711750, 38711751/04-Mar-2027 to 21-Jul-2027; RLV0035-EU/852454006448/36267790, 36909569, 38889437/04-Mar-2027 to 26-Jun-2027; RLV0045/852454006301/35218150, 35218151, 35218152, 35218153, 35219469, 35219470, 35229010, 35229011, 35229012, 35256597, 35318740, 35318741, 35318742, 35329853, 35329854, 35338734, 35360990, 35360991, 35360992, 35406849, 35406850, 35406851, 35406852, 35406853, 35406854, 35498172, 35508065, 35508066, 35523483, 35523484, 35525666, 35535444, 35535445, 35535446, 35535664, 35535665, 35536282, 35538112, 35538113, 35538114, 35538115, 35717797, 35717798, 35717861, 35729587, 35729588, 35729589, 35738091, 35738092, 35738093, 35755411, 35755412, 35755413, 35788378, 35788379, 35788400, 35789578, 35789579, 35789840, 35822293, 35822294, 35822295, 35831721, 35831722, 35831723, 35884013, 35884014, 35884015, 35884018, 35884019, 35884200, 35929731, 35929732, 35929733, 35939403, 35939404, 35939405, 36007510, 36051631, 36051632, 36051633, 36061072, 36061073, 36063527, 36115270, 36115271, 36162140, 36201849, 36211661, 36211662, 36247081, 36247082, 36247083, 36299784, 36299785, 36306155, 36306908, 36312074, 36312075, 36312076, 36327640, 36327641, 36334682, 36334683, 36334684, 36337568, 36337569, 36356333, 36368650, 36368651, 36390837, 36390839, 36390960, 36390962, 36390963, 36411346, 36440610, 36440611, 36451984, 36455377, 36455378, 36528101, 36538547, 36538548, 36554903, 36576169, 36576170, 36576171, 36578105, 36578106, 36578107, 36578108, 36709018, 36714451, 36719971, 36731035, 36731036, 36731037, 36733644, 36733645, 36743247, 36743248, 36765841, 36765842, 36765843, 36778439, 36779700, 36779701, 36789363, 36789364, 36793775, 36793776, 36802652, 36802653, 36802654, 36814178, 36819483, 36819484, 36825817, 36835476, 36839951, 36839952, 36892404, 36892405, 36907158, 36920578, 36931525, 36938443, 36938444, 36938445, 36948565, 36965128, 36977051, 36984738, 37006934, 37006935, 37021399, 37038790, 37060446, 37072745, 37072746, 37115658, 37126596, 37126597, 37143684, 37143685, 37182176, 37182177, 37191624, 37203924, 37210305, 37222391, 37222392, 37248543, 37248544, 37257912, 37278064, 37278065, 37291738, 37422836, 37422837, 37422838, 37434285, 37434286, 37443230,37467700, 37467701, 37467702, 37499290, 37499291, 37513740, 37513741, 37527332, 37558530, 37558531, 37586529, 37604798, 37604928, 37752893, 37752894, 37808050, 37808051, 37818544, 37818545, 37818546, 37864351, 37864352, 37876714, 37876715, 37876716, 37876717, 37909770, 37909771, 37923968, 37941727, 37941728, 38243913, 38243914, 38243915, 38253672, 38253673, 38253674, 38253675, 38257979, 38258120, 38258121, 38393517, 38393518, 38461277, 38461278, 38462080, 38462081, 38462082, 38462587, 38472779, 38473000, 38473001, 38473954, 38484605, 38484606, 38496711, 38517754, 38531680, 38600809, 38600810, 38600811, 38600812, 38632672, 38632673, 38632674, 38711753, 38712187, 38917000, 38917002, 39009366/03-Oct-2026 to 03-Mar-2028; RLV0045-EU/852454006431/36190488, 36254922, 36411351, 36909570, 37575379, 38654200/11-Feb-2027 to 04-Nov-2027; RLV0055/852454006226/35235794, 35235795, 35235796, 35235797, 35235798, 35235799, 35235940, 35244678, 35244679, 35245420, 35265922, 35309126, 35309128, 35309130, 35316016, 35328273, 35328274, 35367311, 35367312, 35367313, 35453674, 35461292, 35461293, 35461294, 35461295, 35461296, 35469325, 35469326, 35469327, 35498173, 35498174, 35508414, 35508415, 35516633, 35518863, 35518864, 35520116, 35522815, 35522816, 35522817, 35522818, 35530433, 35530434, 35582150, 35582151, 35582152, 35585353, 35585354, 35590649, 35590650, 35738094, 35738095, 35738096, 35738097, 35738098, 35738099, 35780448, 35780449, 35780450, 35789910, 35789911, 35789912, 35796422, 35796423, 35796424, 35831784, 35831785, 35831786, 35841864, 35841865, 35841866, 35849369, 35849370, 35849371.35871196, 35871197, 35871198, 35890764, 35890765, 35890766, 35989202, 35989203, 35989204, 36042033, 36042034, 36042035, 36051634, 36051635, 36051636, 36103927, 36103928, 36103930, 36168356, 36201850, 36254923, 36254924, 36258582, 36299786, 36299787, 36306157, 36306158, 36306159, 36315014, 36315015, 36327258, 36327259, 36334687, 36334688, 36334689, 36346737, 36346738, 36346739, 36346800, 36356510, 36356511, 36368652, 36368653, 36377745, 36421406, 36421407, 36421408, 36442449, 36442450, 36455379, 36466553, 36466554, 36474482, 36509579, 36509640, 36511507, 36518452, 36518454, 36519641, 36519642, 36538550, 36588268, 36588269, 36588270, 36588271, 36588273, 36627806, 36779702, 36779703, 36779704, 36789365, 36796277, 36796278, 36801923, 36879218, 36879219, 36879300, 36885626, 36885627, 36948834, 37018069, 37018070, 37021397, 37021398, 37030573, 37038794, 37060447, 37072747, 37072748, 37072749, 37085884, 37085885, 37097347, 37097348, 37097349, 37097352, 37105997, 37105998, 37114545, 37143682, 37143683, 37234124, 37234125, 37248545, 37266940, 37266941, 37278066, 37291818, 37291819, 37422839, 37423650, 37423651, 37434287, 37434288, 37476160, 37476161, 37486544, 37513375, 37513376, 37517300, 37527333, 37527334, 37527335, 37528240, 37528241, 37548528, 37548529, 37548530, 37548532, 37576081, 37586522, 37586530, 37597563, 37597564, 37606989, 37606990, 37606992, 37723478, 37752895, 37752896, 37752897, 37771990, 37771991, 37771993, 37772123, 37772124, 37772125, 37864353, 37864354, 37864355, 37885352, 37885353, 37930763, 37930764, 37930765, 37954165, 37954166, 37954167, 37954168, 38048470, 38196327, 38196328, 38243169, 38243170, 38252987, 38252988, 38252989, 38287216, 38287217, 38287218, 38404416, 38462594, 38474849, 38474880, 38474881, 38481950, 38481954, 38483325, 38496708, 38496709, 38529820, 38531658, 38531659, 38537775, 38540454, 38540782, 38540792, 38614681, 38711754, 38712189/10-Oct-2026 to 29-Oct-2027; RLV0055-EU/852454006318/36190489/30-Jan-2027 to 30-Jan-2027; RLV0065/852454006424/37166269, 37166270, 37205965, 37422790, 37422791, 37434847, 37434848, 37822997, 37822998, 37822999, 37837635, 37837636, 37837637, 37859920, 37903427, 37903428, 37909761, 37930774, 37930775, 37980230, 37980231, 37980232, 37980233, 38079204, 38079205, 38079206, 38079207, 38107765, 38107766, 38107767, 38107768, 38135922, 38135923, 38135924, 38154984, 38154985, 38159015, 38166823, 38166824, 38166825, 38166826, 38166827, 38166828, 38169819, 38170180, 38170693, 38170694, 38205657, 38205658, 38205659, 38222326, 38222327, 38222328, 38222329, 38222330, 38243916, 38339417, 38339418, 38339419, 38339880, 38339881, 38347588, 38355042, 38358582, 38358583, 38358584, 38360834, 38360835, 38366171, 38384137, 38384138, 38384139, 38392693, 38404412, 38404413, 38412000, 38412001, 38412003, 38412005, 38412006, 38412007, 38415242, 38415246, 38529822, 38531280, 38531281, 38559204, 38559205, 38559206, 38559207, 38568272, 38580272, 38580273, 38580274, 38580275, 38589983, 38589984, 38589985, 38590380, 38590381, 38590382, 38590383, 38590384, 38590385, 38590386, 38590387, 38590388, 38597090, 38623873, 38623874, 38623875, 38623876, 38632676, 38632678, 38643661, 38645060, 38645061, 38645062, 38645063, 38692073, 38692074, 38692075, 38692076, 38703597, 38708720, 38708721, 38708722, 38708723, 38712509, 38712511, 38712512, 38739451, 38739453, 38739454, 38739455, 38750004, 38750005, 38750007, 38774400, 38774401, 38802555, 38802556, 38802557, 38802558, 38809866, 38812686, 38812687, 38812688, 38821188, 38821189, 38821191, 38823209, 38823210, 38858563, 38858564, 38858565, 38858566, 38858570, 38869945, 38870375, 38870378, 38870783, 38870784, 38878535, 38878536, 38878538, 38878539, 38879225, 38879226, 38949331, 38949332, 38949333, 38949334, 39032543/30-Jul-2027 to 15-Mar-2028; RLV0075/852454006257/35490122, 35490123, 35490124, 35490125, 35490126, 35604788, 35615708, 35615709, 35705365, 35705366, 35901453, 35901454, 35901455, 35936488, 35936489, 35998826, 35998827, 35998828, 36159579, 36254921.36349203, 36349204, 36404502, 36404503, 36421409, 36835470, 36870653, 36870654, 36874432, 36874433, 36929050, 36929051, 36929052, 36929053, 36929054, 36983480, 36983481, 37060448, 37107838, 37107839, 37191554, 37191555, 37191556, 37204731, 37210306, 37335984, 37355535, 37364022, 37535449, 37535450, 37752898, 37943227, 37976478, 37976479, 37977720, 37977721, 38048471, 38179848, 38249218, 38249700, 38249991, 38253644, 38253645, 38253670, 38253671, 38347590, 38384136, 38426598, 38451722, 38451723, 38508092, 38529821, 38531016, 38539792, 38540765, 38540766, 38655662, 38730788, 38763797, 38763799/29-Oct-2026 to 21-Nov-2027; RLV0075-EU/852454006455/36267789, 36932250, 37575378/23-Jan-2027 to 14-Aug-2027

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates