Skip to content

Intracept RLV 0075Recalls & reported problems

Relievant Medsystems, Inc. · Intracept Access Instruments - Additional VB · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00852454006257), received between Nov 11, 2025 and Aug 14, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IIOpen, Classified

    Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection.

    Z-2694-2026 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Fragmentation3
  • Nonstandard Device1

Patient problems

  • Foreign Body In Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 14, 2026Material Fragmentation

    Patient problems: Foreign Body In Patient

    MAUDE report 50263444 ↗
  • MalfunctionDec 30, 2025Material Fragmentation; Nonstandard Device; Material Fragmentation; Nonstandard Device

    IT WAS REPORTED THAT A CURVED CANNULA WAS RETURNED FOR AN UNKNOWN REASON AND ANALYZED. IT WAS FOUND THAT THE CURVED CANNULA SHAFT TIPS WERE SHEARED. A LABELING REVIEW DID NOT REVEAL ANY ANOMALIES AS IT STATES THAT AS WITH ANY SURGICAL INSTRUMENT, CAREFUL ATTENTION MUST BE EXERCISED TO ENSURE THAT EXCESSIVE FORCE IS NOT PLACED ON THE INSTRUMENTS OR PROBE. IT IS UNKNOWN WHETHER ANY DEVICE FRAGMENTS WERE LEFT BEHIND INSIDE THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23931465 ↗
  • MalfunctionNov 11, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT A CURVED CANNULA WAS RETURNED FOR AN UNKNOWN REASON. A VISUAL INSPECTION REVEALED THAT THE CURVED CANNULA SHAFT TIP WAS SHEARED. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: Foreign Body In Patient

    MAUDE report 23519845 ↗

Other Probe, Radiofrequency Lesion devices

All Probe, Radiofrequency Lesion devices →

Sources and updates