Azurion recallZ-2286-2026 · Philips Medical Systems Nederland B.V. · initiated Dec 7, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
FDA root cause category: Software design
What the company is doing
Philips issued an IMPORTANT PRODUCT NOTICE to its consignees on 12/7/2023 via USPS mail. The notice explained the problem with the device and the circumstances under which if could occur and actions to be taken to minimize the effect of the problem. The notice further instructed the user to: - Keep this Important Product Notice with the documentation of the system until Philips corrects your system. - Circulate this notice to all users of the system so that they are aware of the issue. - Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Important Product Notice. For further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
Recall details
- Recall number
- Z-2286-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 7, 2023
- Posted by the FDA
- Jun 1, 2026
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 3 units
- Product code
- OWB
- FDA submissions
- K172822, K181830, K200917
- Distribution
- US Nationwide distribution.
Product description
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Code information (models, lots, serial numbers)
UDI (01)00884838099227(21)88, (01)00884838099227(21)135, (01)00884838099227(21)63.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls1 death197 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98977).
- Z-2284-2026 · Open, ClassifiedPhilips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Z-2285-2026 · Open, ClassifiedPhilips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Z-2287-2026 · Open, ClassifiedPhilips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Z-2288-2026 · Open, ClassifiedPhilips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Z-2289-2026 · Open, ClassifiedPhilips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Z-2290-2026 · Open, ClassifiedPhilips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Z-2291-2026 · Open, ClassifiedPhilips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page