AzurionRecalls & reported problems across all models

Philips Medical Systems Nederland B.V. · 22 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

69 open recalls name at least one Azurion model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
9,199
Deaths reported
13
Injuries reported
54
Recalls
71

We found 22 models of Azurion by Philips Medical Systems Nederland B.V. in the FDA's device registry (GUDID). Together they are named in 9,199 adverse event reports in the FDA's MAUDE database, received between Feb 20, 2019 and Sep 30, 2026 — classified as 13 deaths, 54 injuries, 9,132 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (3,557 reports).

71 FDA recall records name at least one of these models by its device identifier, 69 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 187 reportsDec 2023: 176 reportsJan 2024: 151 reportsFeb 2024: 230 reportsMar 2024: 139 reportsApr 2024: 136 reportsMay 2024: 170 reportsJun 2024: 123 reportsJul 2024: 130 reportsAug 2024: 185 reportsSep 2024: 260 reportsOct 2024: 257 reportsNov 2024: 233 reportsDec 2024: 180 reportsJan 2025: 246 reportsFeb 2025: 244 reportsMar 2025: 248 reportsApr 2025: 227 reportsMay 2025: 273 reportsJun 2025: 267 reportsJul 2025: 284 reportsAug 2025: 283 reportsSep 2025: 663 reportsOct 2025: 403 reportsNov 2025: 386 reportsDec 2025: 217 reportsJan 2026: 183 reportsFeb 2026: 160 reportsMar 2026: 145 reportsApr 2026: 97 reportsMay 2026: 103 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 20 reportsOct 2026: 0 reports202420252026663
Last 36 months. The most recent month may be incomplete.

Azurion recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 16, 2026Open, Classified

    Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications.

    Z-2932-2026 · Root cause: Software design · names 8 of these models · Recall details · FDA record ↗

  2. Jul 16, 2026Open, Classified

    Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications

    Z-2931-2026 · Root cause: Software design · names 14 of these models · Recall details · FDA record ↗

  3. Jun 17, 2026Open, Classified

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Z-2892-2026 · Root cause: Under Investigation by firm · names 14 of these models · Recall details · FDA record ↗

  4. Apr 30, 2026Open, Classified

    It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

    Z-2258-2026 · Root cause: Under Investigation by firm · names 14 of these models · Recall details · FDA record ↗

  5. Apr 28, 2026Open, Classified

    Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

    Z-2225-2026 · Root cause: Device Design · names 21 of these models · Recall details · FDA record ↗

  6. Apr 28, 2026Open, Classified

    The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

    Z-2227-2026 · Root cause: Component design/selection · names 10 of these models · Recall details · FDA record ↗

  7. Mar 31, 2026Open, Classified

    Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

    Z-2170-2026 · Root cause: Device Design · names 18 of these models · Recall details · FDA record ↗

  8. Mar 19, 2026Open, Classified

    Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

    Z-1805-2026 · Root cause: Labeling design · names 7 of these models · Recall details · FDA record ↗

  9. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1680-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  10. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1679-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  11. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1678-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  12. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1677-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  13. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1676-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  14. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1675-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  15. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1674-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  16. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1673-2026 · Root cause: Device Design · Recall details · FDA record ↗

  17. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1351-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  18. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1350-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  19. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1349-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  20. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1348-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  21. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1347-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  22. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1346-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  23. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1345-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  24. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1344-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  25. Dec 12, 2025Open, Classified

    Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

    Z-1113-2026 · Root cause: Software design · names 16 of these models · Recall details · FDA record ↗

  26. Dec 12, 2025Open, Classified

    Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

    Z-1091-2026 · Root cause: Software design · names 14 of these models · Recall details · FDA record ↗

  27. Dec 8, 2025Open, Classified

    Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

    Z-1079-2026 · Root cause: Software design · names 8 of these models · Recall details · FDA record ↗

  28. Aug 1, 2025Open, Classified

    Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

    Z-2502-2025 · Root cause: Under Investigation by firm · names 14 of these models · Recall details · FDA record ↗

  29. Jun 4, 2025Open, Classified

    Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.

    Z-2134-2025 · Root cause: Device Design · Recall details · FDA record ↗

  30. May 28, 2025Open, Classified

    A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).

    Z-1969-2025 · Root cause: Software design · names 14 of these models · Recall details · FDA record ↗

…and 41 older recalls, listed on each model's page.

Most reported Azurion problems

Device problems

  • Activation Failure3,557
  • Radiation Output Failure3,239
  • Positioning Failure1,378
  • No Display/Image556
  • Detachment of Device or Device Component116
  • Unintended System Motion87
  • Failure to Power Up65
  • Display or Visual Feedback Problem24

Patient problems

  • No Clinical Signs, Symptoms or Conditions9,062
  • Insufficient Information133
  • No Known Impact Or Consequence To Patient14
  • No Consequences Or Impact To Patient10
  • Crushing Injury7
  • Hair Loss7
  • Laceration(s)6
  • Bruise/Contusion5

Azurion models

22 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Azurion adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.